Gammie A, Kitney D, Drake M, Abrams P, Fry C. The calculation and comparison of the Detrusor Contractility Parameter and Watts Factor.
Neurourol Urodyn 2018;
37:2745-2752. [PMID:
29974503 DOI:
10.1002/nau.23745]
[Citation(s) in RCA: 6] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/15/2018] [Accepted: 06/12/2018] [Indexed: 01/22/2023]
Abstract
AIMS
To test the significance of association between a validated index of detrusor contractility, vCE , the Watts Factor, and the Detrusor Contractility Parameter, t20-80 ; and to test whether t20-80 depends on outflow tract resistance as indicated by Bladder Outlet Obstruction Index (BOOI).
METHODS
Thirty-seven pressure-flow traces from 20 male and 17 female patients were analyzed and forms of the Watts Factor, t20-80 and BOOI were compared with vCE .
RESULTS
The Detrusor Contractility Parameter, t20-80 , is significantly associated with vCE for both women and men without a high degree of bladder outlet obstruction. The Watts Factor only had a significant association with vCE at the point of maximum flow in women.
CONCLUSIONS
The Detrusor Contractility Parameter (DCP) (t20-80 ), can be measured easily from the pressure flow curves of a urodynamic test. The Watts Factor at maximum urine flow, WFQmax , can be readily calculated, but is only applicable to women. In both women and men without a high degree of bladder outlet obstruction, DCP is better associated with true detrusor contractility than any Watts Factor analysis.