Morisaka H, Matsuura K, Yamaguchi H, Ichikawa T, Onishi H. Effect of decreased contrast injection flow rate on aortic enhancement in 80-KV peak CT with contrast dose reduction.
Acta Radiol 2021;
64:353-359. [PMID:
34923851 DOI:
10.1177/02841851211067144]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
BACKGROUND
Effect of decreased injection flow rate of contrast agent at the same iodine dose and delivery rate on aortic enhancement has not been clearly elucidated.
PURPOSE
To evaluate the effect of decreased injection flow rate of contrast agent on aortic peak enhancement in a dynamic flow phantom and on aortic enhancement in clinical dynamic 80-kVp computed tomography (CT) with contrast dose reduction.
MATERIAL AND METHODS
In the dynamic flow phantom experiment, the effect of a decreased injection flow rate at the same total iodine dose and delivery rate on simulated aortic peak enhancement was evaluated. In the clinical retrospective study, we searched 312 patients with renal dysfunction who underwent an 80-kVp abdominal dynamic CT with 40% reduction of contrast agent from a standard 120-kVp protocol and measured the aortic enhancement at the level of the hepatic hilum. Independent predictors for aortic enhancement were determined by multiple linear regression analysis, and after adjustment of significant predictors, independent variables for acquiring optimal aortic enhancement, ≥300 HU, were determined by multiple logistic regression analysis.
RESULTS
In the phantom experiment, decreased flow rate showed a significant but small descent effect (6%-9%) on simulated aortic peak enhancement. In the multiple linear regression analysis, only age was an independent predictor of aortic enhancement; there was no independent predictor for optimal age-adjusted aortic enhancement of ≥300 HU.
CONCLUSIONS
Decreased injection flow rate had a small influence on aortic enhancement in vitro but had no significant effect on the aortic enhancement in clinical dynamic 80-kVp CT.
Collapse