1
|
Tong Q, Cai J, Wang Z, Sun Y, Liang X, Xu Q, Mahamoud OA, Qian Y, Qian Z. Recent Advances in the Modification and Improvement of Bioprosthetic Heart Valves. SMALL (WEINHEIM AN DER BERGSTRASSE, GERMANY) 2024; 20:e2309844. [PMID: 38279610 DOI: 10.1002/smll.202309844] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 10/30/2023] [Revised: 12/10/2023] [Indexed: 01/28/2024]
Abstract
Valvular heart disease (VHD) has become a burden and a growing public health problem in humans, causing significant morbidity and mortality worldwide. An increasing number of patients with severe VHD need to undergo heart valve replacement surgery, and artificial heart valves are in high demand. However, allogeneic valves from donors are lacking and cannot meet clinical practice needs. A mechanical heart valve can activate the coagulation pathway after contact with blood after implantation in the cardiovascular system, leading to thrombosis. Therefore, bioprosthetic heart valves (BHVs) are still a promising way to solve this problem. However, there are still challenges in the use of BHVs. For example, their longevity is still unsatisfactory due to the defects, such as thrombosis, structural valve degeneration, calcification, insufficient re-endothelialization, and the inflammatory response. Therefore, strategies and methods are needed to effectively improve the biocompatibility and longevity of BHVs. This review describes the recent research advances in BHVs and strategies to improve their biocompatibility and longevity.
Collapse
Affiliation(s)
- Qi Tong
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Jie Cai
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Zhengjie Wang
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Yiren Sun
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Xuyue Liang
- Department of Biotherapy, Cancer Center and State Key Laboratory of Biotherapy, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Qiyue Xu
- School of Basic Medicine, Mudanjiang Medical University, Mudanjiang, Heilongjiang, 157011, P. R. China
| | - Oumar Abdel Mahamoud
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Yongjun Qian
- Department of Cardiovascular Surgery, National Clinical Research Center for Geriatrics, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| | - Zhiyong Qian
- Department of Biotherapy, Cancer Center and State Key Laboratory of Biotherapy, West China Hospital, Sichuan University, #37 Guoxue Alley, Chengdu, Sichuan, 610041, P. R. China
| |
Collapse
|
2
|
Balmforth D, Dimagli A, Benedetto U, Uppal R. Fifty years of the pericardial valve: Long-term results in the aortic position. J Card Surg 2021; 36:2865-2875. [PMID: 33982282 DOI: 10.1111/jocs.15604] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/11/2021] [Revised: 03/23/2021] [Accepted: 04/18/2021] [Indexed: 11/26/2022]
Abstract
It is now 50 years since the development of the first pericardial valve in 1971. In this time significant progress has been made in refining valve design aimed at improving the longevity of the prostheses. This article reviews the current literature regarding the longevity of pericardial heart valves in the aortic position. Side by side comparisons of freedom from structural valve degeneration are made for the valves most commonly used in clinical practice today, including stented, stentless, and sutureless valves. Strategies to reduce structural valve degeneration are also discussed including methods of tissue fixation and anti-calcification, ways to minimise mechanical stress on the valve, and the role of patient prosthesis mismatch.
Collapse
Affiliation(s)
| | | | | | - Rakesh Uppal
- Barts Heart Centre, St Bartholomew's Hospital, London, UK
| |
Collapse
|
3
|
Claramunt R, Alvarez-Ayuso L, García-Páez JM, Ros A, Casado MC. Changes in the mechanical properties of chemically treated bovine pericardium after a short uniaxial cyclic test. Artif Organs 2012; 37:183-8. [PMID: 23043423 DOI: 10.1111/j.1525-1594.2012.01538.x] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/27/2022]
Abstract
The mechanical behavior of calf pericardium, a biomaterial utilized in the manufacture of cardiac bioprostheses, in response to a short tensile cyclic test has been evaluated. The trial involved 120 samples cut longitudinally or transversely, subjected to 10 cycles until a stress of between 1 and 3 MPa was reached. Tests of hardness and tear propagation were performed, and the results were compared with a control series. The energy loss was also computed, and it was approximately 10-fold greater in the first cycle than the loss in the subsequent nine cycles. Despite this singularity, they correlated very precisely. The effect of the direction in which the tissue is cut on energy loss was not significant nor the difference between hardness prior to and after testing. The results of the tear propagation tests gave no statistical differences prior to and after testing. From the obtained results, it seems that the test carried out does not affect significantly the mechanical properties of calf pericardium.
Collapse
Affiliation(s)
- Rafael Claramunt
- Department of Structural Mechanics, Escuela Técnica Superior de Ingenieros Industriales, Universidad Politécnica de Madrid, Madrid, Spain.
| | | | | | | | | |
Collapse
|
4
|
Claramunt R, García Páez J, Alvarez L, Spottorno J, Ros A, Casado M. Short-term fatigue testing can predict medium-term pericardium behaviour. J Mech Behav Biomed Mater 2011; 4:1929-35. [DOI: 10.1016/j.jmbbm.2011.06.009] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/29/2010] [Revised: 06/13/2011] [Accepted: 06/13/2011] [Indexed: 11/15/2022]
|
5
|
The Mosaic porcine bioprosthesis: Role of age on clinical performance in aortic position. J Thorac Cardiovasc Surg 2011; 141:1440-8.e1. [DOI: 10.1016/j.jtcvs.2010.05.052] [Citation(s) in RCA: 10] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/09/2009] [Revised: 04/22/2010] [Accepted: 05/20/2010] [Indexed: 11/20/2022]
|
6
|
Riess FC, Bader R, Cramer E, Hansen L, Kleijnen B, Wahl G, Wallrath J, Winkel S, Bleese N. Hemodynamic performance of the Medtronic Mosaic porcine bioprosthesis up to ten years. Ann Thorac Surg 2007; 83:1310-8. [PMID: 17383332 DOI: 10.1016/j.athoracsur.2006.07.034] [Citation(s) in RCA: 40] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/03/2006] [Revised: 07/14/2006] [Accepted: 07/18/2006] [Indexed: 11/19/2022]
Abstract
BACKGROUND The Mosaic bioprosthesis (Medtronic, Minneapolis, MN) is a third-generation stented porcine bioprosthesis combining physiologic fixation and amino oleic acid antimineralization treatment to improve hemodynamic performance and durability. The findings of this single-center experience with this valve were evaluated to determine the clinical and hemodynamic performance. METHODS Between February 1994 and October 1999, we enrolled 255 patients with aortic valve replacement (AVR) with a mean age of 67 years (range, 23 to 82 years) and 47 patients with mitral valve replacement (MVR) with a mean age of 67 years (range, 41 to 84 years) in this post-United States Food and Drug Administration approval prospective and nonrandomized clinical trial. Patients were followed-up, including serial echocardiographic assessment, within 30 days, at 6 months, and annually thereafter. The cumulative follow-up was 1540 patient-years for AVR (mean, 6.1 years; maximum, 10 years) and 250 patient-years for MVR (mean, 5.4 years, maximum; 10 years). RESULTS Early mortality after AVR (<30 days) was 0.8%; late mortality per patient-year was 3.5%, including a valve-related/unexplained mortality of 1.1%. Early mortality after MVR (<30 days) was 0.0%; late mortality per patient-year was 2.8%, including a valve-related/unexplained mortality of 1.2%. Median postoperative gradient and effective orifice area for all valves after AVR were (early, n = 252; 5 years, n = 161; 9 years, n = 43) 13.7, 12.3, and 11.7 mm Hg and 1.9, 1.8, and 1.8 cm2 at early, 5 years, and 9 years, respectively. With MVR respective data were (early, n = 46; 5 years, n = 25; 7 years, n = 13) 4.6, 4.1, and 3.9 mm Hg and 1.8, 2.2, and 2.3 cm2. At 10 years, freedom from adverse events in the AVR group and MVR group was, respectively, thromboembolism, 86.6% +/- 6.6% and 86.3% +/- 9.8%; permanent neurologic event, 91.2% +/- 6.8% and 90.9% +/- 8.7%; valve thrombosis, 98.2% +/- 0.8% and 100%; structural valve deterioration, 87.1% +/- 6.7% and 100%. CONCLUSIONS Our midterm results demonstrate clinical safety and good performance of the Mosaic bioprosthesis. Continued follow-up will determine if this new design will provide increased durability.
Collapse
|
7
|
Bottio T, Valente M, Rizzoli G, Tarzia V, Bisleri G, Pettenazzo E, Gerosa G, Thiene G. Commissural dehiscence: A rare and peculiar cause of porcine valve structural deterioration. J Thorac Cardiovasc Surg 2006; 132:1017-22. [PMID: 17059917 DOI: 10.1016/j.jtcvs.2006.06.012] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/02/2006] [Revised: 05/23/2006] [Accepted: 06/07/2006] [Indexed: 11/19/2022]
Abstract
OBJECTIVE Calcification is the main cause of structural valve deterioration; however, other causes of failure have been identified, and among them, dehiscence of a commissure from the stent has been reported in several models of porcine valves. The aim of this study was to analyze the rate and mode of occurrence of this complication in first- and second-generation porcine bioprosthetic explants. METHODS Among 586 porcine xenografts explanted and analyzed at the Institute of Pathological Anatomy of the University of Padua, 17 (2.9%) have been replaced for incompetence because of commissural dehiscence. All these explants were in the mitral position, with the exception of a Carpentier-Edwards supra-annular aortic valve prosthesis. RESULTS Dehiscence was observed in 9 (1.9%) of 455 Hancock standard explants, in 1 (3.2%) of 31 Hancock II, in 3 (8.6%, 2 standard and 1 supra-annular) of 35 Carpentier-Edwards, in 1 (2.4%) of 42 Bioimplants, and in 3 (50%) of 6 Xenotech after a mean time function of 157 +/- 50, 156, 96 +/- 29, 143, and 130 +/- 8 months, respectively. Dehiscence was the sole cause of incompetence in 6 cases. An impending commissural dehiscence caused by blood creeping was observed in one case. This might be an explanation for the dehiscence other than excessive trimming of the aortic wall. CONCLUSIONS Commissural dehiscence is an uncommon and peculiar mode of failure of porcine valves implanted in the mitral position and was observed earlier and more frequently with Carpentier-Edwards porcine explants (P < .05). We speculate that pericardial strip protection of the suture between the Dacron fabric and porcine aortic wall, as used in the Biocor porcine valve, might prevent this complication.
Collapse
Affiliation(s)
- Tomaso Bottio
- Department of Cardiovascular Surgery, University of Padua Medical School, Padua, Italy.
| | | | | | | | | | | | | | | |
Collapse
|
8
|
García Páez JM, Carrera A, Jorge E, Millán I, Cordón A, Maestro MA, Rocha A, Morales S, Castillo-Olivares JL. Determination of the force necessary for the propagation of tears in ostrich and calf pericardium. J Biomed Mater Res B Appl Biomater 2006; 79:229-35. [PMID: 16637030 DOI: 10.1002/jbm.b.30533] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
The durability of prosthetic heart valve leaflets made of biological materials is limited. A tear in the biomaterial accelerates their early failure, but microtearing of the collagen fibers may be responsible for their medium-term failure. We studied the force necessary to propagate tearing in two biomaterials: ostrich and calf pericardium. One hundred twenty samples of each tissue were tested in an Elmendorf pendulum capable of measuring the force required to tear a tissue in which a predefined slit had been made. The forces required to produce tears, ranging between 2.5 and 0.25 cm in length, were determined. For ostrich pericardium, this force ranged between 67.67 and 4.80 newton, while that required to tear the same lengths of calf pericardium ranged between 70.67 and 4.70 newton. The function that relates the tearing force to the length of the tear was expressed as follows: y = 20.62x + 1.77x(2) (R(2) = 0.923) for ostrich pericardium and y = 45.57x - 7.21x(2) (R(2) = 0.936) for calf pericardium, where y is the force in newton and x is the length in centimeter. Calf pericardium was found to have a greater resistance to tearing. However, these results should be interpreted with caution owing to the fact that the thickness of the majority of the samples of ostrich pericardium was significantly less than that of calf pericardium. A more careful selection and utilization of adult ostrich pericardium would probably improve these results.
Collapse
|
9
|
Páez JMG, Carrera A, Jorge E, Millán I, Cordon A, Maestro MA, Rocha A, Castillo-Olivares JL. Resistance to tearing of calf and ostrich pericardium: Influence of the type of suture material and the direction of the suture line. J Biomed Mater Res B Appl Biomater 2004; 69:125-34. [PMID: 15116401 DOI: 10.1002/jbm.b.20014] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
The tearing of the valve leaflet of a cardiac bioprosthesis can cause early failure of this device, which is employed to replace a diseased native valve. This report involves the study of the behavior of 312 tissue samples (152 of calf pericardium and 160 of ostrich pericardium) treated with glutaraldehyde and subsequently subjected to tear testing. The samples were cut in the two principal directions: longitudinally, or root to apex, and transversely. They included a series of control samples that were left unsutured, and the remaining samples were repaired with the use of two different suture techniques: a running suture in the direction of the load and a telescoping suture perpendicular to the load. Four commercially available suture materials were employed: Pronova, nylon, Gore-Tex, or silk. The unsutured control samples of both types of pericardium exhibited a similar anisotropic behavior in the tear test. The mean resistance to tearing of the calf pericardium was 24.29 kN m in samples cut longitudinally and 34.78 kN m in those cut transversely (p =.03); the values were 28.08 kN m and 37.12 kN m (p =.002), respectively, in ostrich pericardium. The series repaired with the telescoping suture always exhibited greater resistance to tearing, with values that ranged between 44.34 and 64.27 kN for the samples of calf pericardium and from 41.65 to 47.65 kN for those obtained from ostrich. These assays confirm the anisotropic behavior of calf and ostrich pericardium treated with glutaraldehyde when subjected to tear testing, as well as the loss of this behavior in ostrich pericardium after suturing. Suturing techniques, such as the telescoping model, that provide a greater resistance to tearing should be studied for use in the design of the valve leaflets of cardiac bioprostheses made of biological materials.
Collapse
|
10
|
Hartz RS, Deleon SY, Lane J, Dorotan J, Joyce J, Urbina E, Ross-Ascuitto N, Ascuitto R. Medtronic freestyle valves in right ventricular outflow tract reconstruction. Ann Thorac Surg 2003; 76:1896-900. [PMID: 14667607 DOI: 10.1016/s0003-4975(03)01301-8] [Citation(s) in RCA: 22] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 10/26/2022]
Abstract
BACKGROUND Various pulmonary valve substitutes, with their inherent limitations, have been used in children and young adults. We chose the Medtronic Freestyle (Medtronics, Minneapolis, MN) valve because of its excellent hemodynamics, known durability in adults, and design features that allow modifications during implantation. METHODS Over a 3 1/2 year period the Freestyle valve was implanted in 47 patients age 2 to 58 years (mean 14.2, median 12.0) in the pulmonary position. All patients had pulmonic stenosis and(or) insufficiency from previous operations for tetralogy (27), pulmonary atresia (6), truncus (4), or other diagnosis (10). The indication for surgery was pure pulmonary insufficiency in 11 patients, pulmonic stenosis in 3, and mixed stenosis and insufficiency in 33. Root replacement technique was used with additional enlargement of the pulmonary artery branches in 10 patients. RESULTS Intraoperatively, one patient sustained a right ventricle tear and one a circumflex coronary artery injury during the dissection. There was one postoperative death. Two patients developed late subvalvular pannus formation, one of whom required reoperation. One patient was found to have an echo gradient of 95 mm Hg due to decreased leaflet motion and underwent cardiac catheterization at which the peak systolic gradient was determined to be 50 mm Hg. He has not required reintervention during his 3 1/2 years of follow-up. The remaining 43 patients have minimal gradients or insufficiency. All surviving patients are in New York Heart Association (NYHA) Class I. CONCLUSIONS The Medtronic Freestyle valve is an attractive alternative for RVOT (right ventricular outflow tract) reconstruction in children. It is readily available, versatile, and has excellent hemodynamic characteristics. Although long term follow-up is not yet available, longevity of this prosthesis, and freedom from complications, will hopefully be superior to valves with stents.
Collapse
Affiliation(s)
- Renee S Hartz
- Department of Surgery and Pediatrics, Tulane University Medical Center, New Orleans, Louisiana, USA
| | | | | | | | | | | | | | | |
Collapse
|
11
|
Butany J, Fayet C, Ahluwalia MS, Blit P, Ahn C, Munroe C, Israel N, Cusimano RJ, Leask RL. Biological replacement heart valves. Identification and evaluation. Cardiovasc Pathol 2003; 12:119-39. [PMID: 12763552 DOI: 10.1016/s1054-8807(03)00002-4] [Citation(s) in RCA: 25] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/16/2022] Open
Affiliation(s)
- Jagdish Butany
- Department of Pathology, Toronto General Hospital, Toronto, ON, Canada.
| | | | | | | | | | | | | | | | | |
Collapse
|
12
|
de la Fuente A, Sánchez JR, Fernández JL, Romero J, Berjon J, Moriones I. The Medtronic Intact Bioprosthesis: Clinical and Hemodynamic Performance Over 13 Years. Artif Organs 2002; 26:851-5. [PMID: 12296924 DOI: 10.1046/j.1525-1594.2002.06984.x] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
We evaluated our results over 13 years with the aortic-position Medtronic Intact bioprosthesis. Our study involved 91 consecutive patients with isolated aortic valve replacement with the Medtronic Intact bioprosthesis. The follow-up was complete for 95%. Mean follow-up was 6.61 years (range 16 days-13 years), 590 patient years. Early mortality rate was 3.3%. Late mortality was 23 patients. Survival at 13 years was 53.52% (SD = 7.63%). The linearized rate of major thromboembolism was 0.34% per patient year; rate of major bleeding events was 0.33% per patient year. The rate of nonstructural dysfunction was 0.16% per patient year. Rate of reoperation was 0.53% and rate of structural valve deterioration was 0.16% per patient year. New York Heart Association (NYHA) postoperative classes were I to II in 92.21%. Gradients were as follows: 21 to 23.87 mm Hg, 23 to 18 mm Hg, 25 to 15.5 mm Hg, and 27 to 16.50 mm Hg. Structural valve deterioration was low during the 13 years of follow-up. Valve gradients and areas remained the same over the follow-up period. The Medtronic Intact bioprosthesis shows excellent clinical and hemodynamic performance at 13 years of follow up.
Collapse
|
13
|
Jamieson WR, Lemieux MD, Sullivan JA, Munro IA, Métras J, Cartier PC. Medtronic Intact porcine bioprosthesis experience to twelve years. Ann Thorac Surg 2001; 71:S278-81. [PMID: 11388204 DOI: 10.1016/s0003-4975(01)02548-6] [Citation(s) in RCA: 16] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/22/2022]
Abstract
BACKGROUND The Medtronic Intact porcine bioprosthesis was further evaluated to determine the influence of zero-pressure glutaraldehyde fixation on structural valve deterioration (SVD). METHODS From 1986 through 1996, at three Canadian centers, 1,272 patients had 1,296 procedures: 836 aortic valve replacement (AVR), 332 mitral valve replacement (MVR), 14 tricuspid valve replacement, 3 pulmonary valve replacement, and 111 multiple valve replacements. The mean age of the patient population was 67 years (range 9 to 91 years). The total follow-up was 8,011 patient-years (mean 6.2 years). RESULTS The late mortality (overall) was 4.8% and 6.7% per patient-year for AVR with or without concomitant procedures, respectively; and 4.7% and 10.4% per patient-year for MVR, respectively. There were 51 cases of SVD (AVR 22 of 836; MVR 23 of 332; pulmonary valve replacement 1 of 3; and multiple valve replacement 5 of 111). The actuarial freedom from SVD at 12 years for AVR was 94.3%+/-3.3% for patients aged 61 to 70 years and 97.7%+/-1.1% for those more than 70 years; for MVR actuarial freedom from SVD at 12 years was 93.7%+/-3.9% for patients more than 70 years. The actual freedom at 12 years from SVD for AVR was 92.4%+/-3.1% for patients aged 51 to 60 years, 96.1%+/-2.1% for those 61 to 70 years, and 98.4%+/-0.7% for those older than 70 years; for MVR actual freedom from SVD at 12 years was 89.6%+/-3.2% for patients 61 to 70 years and 96.6%+/-3.4% for those more than 70 years. CONCLUSIONS The Medtronic Intact porcine bioprosthesis, formulated with tissue preservation at zero-pressure fixation, has encouraging freedom from structural failure.
Collapse
Affiliation(s)
- W R Jamieson
- Department of Surgery, University of British Columbia, Vancouver, Canada.
| | | | | | | | | | | |
Collapse
|
14
|
O'Brien MF, Gardner MA, Garlick B, Jalali H, Harrocks S, McLoughlin L. The stentless xenograft aortic valve: The wheel turns around. Heart Lung Circ 2000; 9:61-73. [PMID: 16351997 DOI: 10.1046/j.1444-2892.2000.00032.x] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/20/2022]
Abstract
BACKGROUND A brief overview of the historical pathways of both stented and stentless porcine xenografts is presented in order to understand the return to and continuing clinical use of stentless devices. In addition, 7-11 years of durability with various models of stentless porcine valves has now accumulated and is beginning to be of relevance in determining the future place of this xenograft. Stentlessness and anticalcium agents, coupled with the poor results of stented xenografts in certain patient groups, have led to a resurgence of the clinical use of stentless xenograft valves for aortic valve replacement. An overview of the present state and future of stentless valves is given. METHODS At both The Prince Charles Hospital and St Andrew's War Memorial Hospital, Queensland, Australia, 307 patients have received the Model 300 CryoLife-O'Brien stentless composite aortic xenograft from December 1992 to February 2000. Associated procedures were required in 56% of patients (mostly coronary artery bypass, mean 2.4 grafts, in 144 patients (47%) and left ventricular myomectomy in 34 patients (11%)). RESULTS The hospital mortality (four early deaths) has been 1.3 +/- 1% (CL 95%) and the follow-up 100% for this analysis. The mean patient age was 73 years (range 57-89 years with 16% being 80 years and over). Morbid events have included six perivalvar leaks: four trivial and identified only on echo Doppler (no clinical murmurs) and two patients requiring reoperation at 10 days and 12 weeks with simple successful repair verified on subsequent echocardiograms. Of the 307 patients over the 7 year period, three valves only have been explanted, two for endocarditis at 1.5 and 3.5 years and one for possible technically induced structural failure at 15 months (probable needle damage). With this exception, there has been as yet no other intrinsic leaflet failure. Four early thromboembolic events (4 days-5 weeks) in patients with atrial fibrillation (no anticoagulants used postoperatively with the first 80 patients) constituted the important early morbid events. Late mortality of this elderly patient cohort has occurred in 27 patients over 7 years of maximum follow-up. One death (endocarditis) has been valve related at 5 years. Serial echocardiography (some 700 echoes in the study of this valve) has demonstrated a mean gradient of 7-9 mmHg with a very low incidence of trivial incompetence (96%) on Doppler examination with implant valve sizes ranging from 21 to 29 mm. One patient had significant regurgitation requiring reoperation. There has been no progression of either incompetence or stenosis of the remaining patients in this follow-up, now into the eighth postoperative year. CONCLUSION The early and intermediate results appear excellent in this elderly patient cohort. Nevertheless, important surveillance is obviously required to determine the durability at 10-12 years, a crucial time when stented porcine xenografts began to show an obvious failure rate from structural deterioration, in the middle and elderly aged patient cohort. An attempt is made to outline the future of this type of stentless xenograft and to justify that its cautious use should probably be extended down to the over 50 year age patient cohort.
Collapse
Affiliation(s)
- M F O'Brien
- The Prince Charles Hospital and St Andrew's War Memorial Hospital, Brisbane, Queensland, Australia.
| | | | | | | | | | | |
Collapse
|
15
|
Wong SP, Legget ME, Greaves SC, Barratt-Boyes BG, Milsom FP, Raudkivi PJ. Early experience with the mosaic bioprosthesis: a new generation porcine valve. Ann Thorac Surg 2000; 69:1846-50. [PMID: 10892935 DOI: 10.1016/s0003-4975(00)01167-x] [Citation(s) in RCA: 22] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
BACKGROUND The Mosaic bioprosthesis is a new generation stented porcine valve. METHODS Between May 1995 and April 1998, this valve was implanted in the aortic position in 98 patients (70 men; mean age, 69.2 years [34.2 to 83.6 years]). Preoperatively 35 patients were in New York Heart Association functional class 3 or 4. Fifty-nine patients underwent concomitant procedures. The mean duration at follow-up in January 1999 was 23.7 +/- 10.2 months (0.3 to 39.4 months) and totaled 193 patient-years. All but one survivor was in New York Heart Association class 1 or 2. RESULTS Early complications included 1 death, 3 reoperations for bleeding, greater than mild regurgitation (paravalvar) in 1 patient and thromboembolism in 4 patients. Late complications included four deaths, study-valve endocarditis in 3 patients, more than mild regurgitation or hemolysis in 2, and thromboembolism in 2 patients. Late follow-up echocardiography in all survivors showed a mean transaortic gradient of 13.6 +/- 6.7 mm Hg, and an aortic valve area of 1.80 +/- 0.61 cm2. Valve replacement was followed by a significant and sustained decrease in left ventricular mass for all valve sizes. There has been no primary structural valve failure. CONCLUSIONS The early experience with the Mosaic valve in the aortic position has been promising.
Collapse
Affiliation(s)
- S P Wong
- Department of Cardiology, Green Lane Hospital, Auckland, New Zealand
| | | | | | | | | | | |
Collapse
|