Takahashi M, Takehara Y, Fujisaki K, Okuaki T, Fukuma Y, Tooyama N, Ichijo K, Amano T, Goshima S, Naganawa AS. Three Dimensional Ultra-short Echo Time MRI Can Depict Cholesterol Components of Gallstones Bright.
Magn Reson Med Sci 2021;
20:359-370. [PMID:
33390560 PMCID:
PMC8922345 DOI:
10.2463/mrms.mp.2020-0009]
[Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/16/2020] [Accepted: 09/10/2020] [Indexed: 11/21/2022] Open
Abstract
PURPOSE
Non-calcified cholesterol stones that are small in size are hard to be depicted on CT or magnetic resonance cholangiopancreatography. This institutional review board (IRB)-approved retrospective in vitro study aims to characterize contrast behaviors of 3 main components of the gallstones, i.e., cholesterol component (CC), bilirubin calcium component (BC) and CaCO3 (CO) on 3D radial scan with ultrashort TE (UTE) MRI, and to test the capability of depicting CC of gallstones as bright signals as compared to background saline.
METHODS
Fourteen representative gallstones from 14 patients, including 15 CC, 6 BC and 4 CO, were enrolled. The gallstones underwent MRI including fat-saturated T1-weighted image (fs-T1WI) and UTE MRI with dual echoes. The contrast-to-noise ratio (CNR) and the chemical analysis for the 25 portions of the stones were compared.
RESULTS
BC was bright on fs-T1WI, which did not change dramatically on UTE MRI and the signal did not remain on UTE subtraction image between dual echoes. Whereas the CC was negative or faintly positive signal on fs-T1WI, bright signal on UTE MRI and the contrast remained even higher on the UTE subtraction, which reflected their short T2 values. Median CNRs and standard errors of the segments on each imaging were as follows: on fs-T1WI, -10.2 ± 4.2 for CC, 149.7 ± 27.6 for BC and 37.9 ± 14.3 for CO; on UTE MRI first echo, 16.7 ± 3.3 for CC, 74.9 ± 21.3 for BC and 17.7 ± 8.4 for CO; on UTE subtraction image, 30.2 ±2.0 for CC, -11.2 ± 5.4 for BC and 17.8 ± 10.7 for CO. Linear correlations between CNRs and cholesterol concentrations were observed on fs-T1WI with r = -0.885, (P < 0.0001), UTE MRI first echo r = -0.524 (P = 0.0072) and UTE subtraction with r = 0.598 (P = 0.0016).
CONCLUSION
UTE MRI and UTE subtraction can depict CC bright.
Collapse