1
|
Mailhiot S, Mankinen O, Li J, Kharbanda Y, Telkki VV, Urbańczyk M. CAT on MOUSE: Control and automation of temperature for single-sided NMR instruments such as NMR-MOUSE. MAGNETIC RESONANCE IN CHEMISTRY : MRC 2024; 62:252-258. [PMID: 37344254 DOI: 10.1002/mrc.5376] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/28/2023] [Revised: 06/06/2023] [Accepted: 06/08/2023] [Indexed: 06/23/2023]
Abstract
Temperature-dependent experiments are a rapidly growing area of interest for low-field NMR. In this work, we present a new device for wide-range temperature control for single-sided NMR instruments. The presented device, called CAT, is simple to build, inexpensive, and easy to modify to accommodate different samples. We present the capabilities of the device using a freezing temperature study of acetic acid/water mixtures. Additionally, we present the stability of the device over long measurement times. We believe that by introducing such a device with an open-source design, we allow researchers to use it in a wide range of applications and to fully incorporate variable-temperature studies in the world of single-sided instruments.
Collapse
Affiliation(s)
| | - Otto Mankinen
- NMR Research Unit, University of Oulu, Oulu, Finland
| | - Jing Li
- NMR Research Unit, University of Oulu, Oulu, Finland
- NIMBE, CEA, CNRS, Université de Paris Saclay, CEA Saclay, Gif-sur-Yvette, France
| | - Yashu Kharbanda
- NMR Research Unit, University of Oulu, Oulu, Finland
- Laboratoire Navier (Ecole des Ponts ParisTech-Université Gustave Eiffel), Champs-sur-Marne, France
| | | | - Mateusz Urbańczyk
- Institute of Physical Chemistry, Polish Academy of Sciences, Warsaw, Poland
| |
Collapse
|
2
|
Schmid E, Rondeau S, Rudszuck T, Nirschl H, Guthausen G. Inline NMR via a Dedicated V-Shaped Sensor. SENSORS (BASEL, SWITZERLAND) 2023; 23:2388. [PMID: 36904592 PMCID: PMC10007489 DOI: 10.3390/s23052388] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 01/23/2023] [Revised: 02/13/2023] [Accepted: 02/20/2023] [Indexed: 06/18/2023]
Abstract
Process monitoring and control require dedicated and reliable measures which reflect the status of the process under investigation. Although nuclear magnetic resonance is known to be a versatile analytical technique, it is only seldomly found in process monitoring. Single-sided nuclear magnetic resonance is one well known approach for being applied in process monitoring. The dedicated V-sensor is a recent approach that allows the inline investigation of materials in a pipe non-destructively and non-invasively. An open geometry of the radiofrequency unit is realized using a tailored coil, enabling the sensor to be applied for manifold mobile applications in in-line process monitoring. Stationary liquids were measured, and their properties were integrally quantified as the basis for successful process monitoring. The sensor, in its inline version, is presented along with its characteristics. An exemplary field of application is battery production in terms of anode slurries; thus, the first results on graphite slurries will demonstrate the added value of the sensor in process monitoring.
Collapse
Affiliation(s)
- Eric Schmid
- Institute of Mechanical Process Engineering and Mechanics, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
| | - Simon Rondeau
- Institute of Mechanical Process Engineering and Mechanics, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
| | - Thomas Rudszuck
- Institute of Mechanical Process Engineering and Mechanics, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
| | - Hermann Nirschl
- Institute of Mechanical Process Engineering and Mechanics, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
| | - Gisela Guthausen
- Institute of Mechanical Process Engineering and Mechanics, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
- Engler-Bunte-Institut, Chair of Water Chemistry and Water Technology, Karlsruhe Institute of Technology, 76131 Karlsruhe, Germany
| |
Collapse
|
3
|
Nie S, Ratzsch KF, Grage SL, Keller J, Ulrich AS, Lacayo-Pineda J, Wilhelm M. Correlation between Macroscopic Elasticity and Chain Dynamics of Natural Rubber during Vulcanization as Determined by a Unique Rheo-NMR Combination. Macromolecules 2021. [DOI: 10.1021/acs.macromol.1c00354] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Affiliation(s)
- Shouliang Nie
- Institute for Technical Chemistry and Polymer Chemistry, Karlsruhe Institute of Technology (KIT), 76131 Karlsruhe, Germany
| | | | - Stephan L. Grage
- Institute of Biological Interfaces (IBG-2) and Institute of Organic Chemistry (IOC), Karlsruhe Institute of Technology (KIT), 76131 Karlsruhe, Germany
| | - Jonas Keller
- Institute for Technical Chemistry and Polymer Chemistry, Karlsruhe Institute of Technology (KIT), 76131 Karlsruhe, Germany
| | - Anne S. Ulrich
- Institute of Biological Interfaces (IBG-2) and Institute of Organic Chemistry (IOC), Karlsruhe Institute of Technology (KIT), 76131 Karlsruhe, Germany
| | - Jorge Lacayo-Pineda
- Research and Development, Continental Reifen Deutschland GmbH, 30419 Hannover, Germany
| | - Manfred Wilhelm
- Institute for Technical Chemistry and Polymer Chemistry, Karlsruhe Institute of Technology (KIT), 76131 Karlsruhe, Germany
| |
Collapse
|
4
|
Gao Y, Blümich B. Analysis of three-site T 2-T 2 exchange NMR. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2020; 315:106740. [PMID: 32438312 DOI: 10.1016/j.jmr.2020.106740] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/08/2020] [Revised: 04/24/2020] [Accepted: 04/25/2020] [Indexed: 06/11/2023]
Abstract
T2-T2 exchange NMR is a unique method to investigate the pore space and fluid dynamics in porous media. While two-site relaxation exchange is well understood, three-site exchange is not. We analyze the solutions for three-site T2-T2 exchange NMR analytically and by computer simulation. Three main results are obtained. First, the exchange map can be asymmetric in the case of microscale vortex motion in violation of the principle of detailed balance. Second, the apparent longitudinal relaxation times and/or apparent transverse relaxation times can be complex valued. In the case of complex apparent transverse relaxation times, the three-site exchange map coalesces to a two-site exchange map with characteristic oscillations in the time domain. As a result of the oscillations, the shorter relaxation time is less than expected. Third, there can be negative cross-peaks in the exchange map for certain combinations of longitudinal and transverse relaxation times or if the mixing period is shorter than the evolution and detection periods. In view of these results experimental exchange maps may need to be reevaluated.
Collapse
Affiliation(s)
- Yang Gao
- College of Science, China University of Petroleum, Beijing 102249, China; Institut für Technische und Makromolekulare Chemie, RWTH Aachen University, Aachen 52056, Germany.
| | - Bernhard Blümich
- Institut für Technische und Makromolekulare Chemie, RWTH Aachen University, Aachen 52056, Germany
| |
Collapse
|
5
|
Blümich B, Singh K. Desktop NMR and Its Applications From Materials Science To Organic Chemistry. Angew Chem Int Ed Engl 2017; 57:6996-7010. [PMID: 29230908 DOI: 10.1002/anie.201707084] [Citation(s) in RCA: 57] [Impact Index Per Article: 8.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/12/2017] [Indexed: 12/19/2022]
Abstract
NMR spectroscopy is an indispensable method of analysis in chemistry, which until recently suffered from high demands for space, high costs for acquisition and maintenance, and operational complexity. This has changed with the introduction of compact NMR spectrometers suitable for small-molecule analysis on the chemical workbench. These spectrometers contain permanent magnets giving rise to proton NMR frequencies between 40 and 80 MHz. The enabling technology is to make small permanent magnets with homogeneous fields. Tabletop instruments with inhomogeneous fields have been in use for over 40 years for characterizing food and hydrogen-containing materials by relaxation and diffusion measurements. Related NMR instruments measure these parameters in the stray field outside the magnet. They are used to inspect the borehole walls of oil wells and to test objects nondestructively. The state-of-the-art of NMR spectroscopy, imaging and relaxometry with compact instruments is reviewed.
Collapse
Affiliation(s)
- Bernhard Blümich
- Institut für Technische und Makromolekulare Chemie, RWTH Aachen University, Aachen, Germany
| | - Kawarpal Singh
- Institut für Technische und Makromolekulare Chemie, RWTH Aachen University, Aachen, Germany
| |
Collapse
|
6
|
Blümich B, Singh K. NMR mit Tischgeräten und deren Anwendungen von der Materialwissenschaft bis zur organischen Chemie. Angew Chem Int Ed Engl 2017. [DOI: 10.1002/ange.201707084] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/31/2022]
Affiliation(s)
- Bernhard Blümich
- Institut für Technische und Makromolekulare Chemie; RWTH Aachen University; Aachen Deutschland
| | - Kawarpal Singh
- Institut für Technische und Makromolekulare Chemie; RWTH Aachen University; Aachen Deutschland
| |
Collapse
|
7
|
|
8
|
Gabrielle B, Gomez E, Korb JP. Probing Rubber Cross-Linking Generation of Industrial Polymer Networks at Nanometer Scale. J Phys Chem B 2016; 120:5581-9. [DOI: 10.1021/acs.jpcb.6b03601] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Affiliation(s)
| | | | - Jean-Pierre Korb
- Physique
de la Matière Condensée, Ecole Polytechnique-CNRS, 91128 Palaiseau, France
| |
Collapse
|
9
|
Song R, Song YQ, Vembusubramanian M, Paulsen JL. The robust identification of exchange from T2-T2 time-domain features. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2016; 265:164-171. [PMID: 26905815 DOI: 10.1016/j.jmr.2016.02.001] [Citation(s) in RCA: 20] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 12/16/2015] [Revised: 01/30/2016] [Accepted: 02/01/2016] [Indexed: 06/05/2023]
Abstract
Two-dimensional spin-spin relaxation (T2-T2) techniques have been developed to probe coupling between different environments such as diffusive coupling between small and large pores or chemical exchange with clays. In these studies, Numerical Laplace Inversion (NLI) is used to obtain two-dimensional T2-T2 relaxation distribution spectrum from the T2-T2 signal decays, and the off-diagonal peaks characterize coupling. Often, these coupling peaks are both weak and close to the diagonal and combined with the inherently ill-conditioned nature of the inversion, their presence is difficult to differentiate from inversion related artifacts and blurring. This manuscript presents a time domain based analysis to identify the presence of coupling that avoids the ambiguities present in T2-T2 spectra. The approach utilizes the symmetric nature of the two-dimensional time domain data, where the presence of curvature along one of these symmetries gives an unambiguous indicator of coupling. Measurements on porous glass beads are used to verify the technique.
Collapse
Affiliation(s)
- Ruobing Song
- Schlumberger Doll, 1 Hampshire St., Cambridge, MA 02139, United States.
| | - Yi-Qiao Song
- Schlumberger Doll, 1 Hampshire St., Cambridge, MA 02139, United States.
| | | | - Jeffrey L Paulsen
- Schlumberger Doll, 1 Hampshire St., Cambridge, MA 02139, United States.
| |
Collapse
|
10
|
Chen Z, Cai S, Huang Y, Lin Y. High-resolution NMR spectroscopy in inhomogeneous fields. PROGRESS IN NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY 2015; 90-91:1-31. [PMID: 26592943 DOI: 10.1016/j.pnmrs.2015.05.003] [Citation(s) in RCA: 13] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/03/2014] [Revised: 05/20/2015] [Accepted: 05/20/2015] [Indexed: 06/05/2023]
Abstract
High-resolution NMR spectroscopy, providing information on chemical shifts, J coupling constants, multiplet patterns, and relative peak areas, is a mainstream tool for analysis of molecular structures, conformations, compositions, and dynamics. Generally, a homogeneous magnetic field is a prerequisite for obtaining high-resolution NMR information. Magnetic field inhomogeneity, whether from non-ideal experimental conditions or from intrinsic magnetic susceptibility discontinuities in samples, represents a hurdle for applications of high-resolution NMR. Numerous techniques have been proposed for measuring high-resolution NMR spectra free from the influence of inhomogeneous magnetic fields. Besides developments and improvements in NMR instrumentation, various types of experimental approaches have been established for recovering NMR information in inhomogeneous magnetic fields. Three main types are systematically described in this review. In addition, other high-resolution NMR approaches or data processing methods are also briefly described. All high-resolution NMR approaches covered in this review have individual advantages and disadvantages in practical applications, and no one technique is applicable to all practical circumstances. Hence, they are complementary for high-resolution NMR applications in inhomogeneous fields. The underlying mechanisms of these approaches are presented, together with analyses of their applicability and efficiency.
Collapse
Affiliation(s)
- Zhong Chen
- Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, State Key Laboratory of Physical Chemistry of Solid Surfaces, Xiamen University, Xiamen, Fujian 361005, China.
| | - Shuhui Cai
- Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, State Key Laboratory of Physical Chemistry of Solid Surfaces, Xiamen University, Xiamen, Fujian 361005, China
| | - Yuqing Huang
- Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, State Key Laboratory of Physical Chemistry of Solid Surfaces, Xiamen University, Xiamen, Fujian 361005, China
| | - Yulan Lin
- Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, State Key Laboratory of Physical Chemistry of Solid Surfaces, Xiamen University, Xiamen, Fujian 361005, China
| |
Collapse
|
11
|
Dibbanti MK, Mauri M, Mauri L, Medaglia G, Simonutti R. Probing small network differences in sulfur-cured rubber compounds by combining nuclear magnetic resonance and swelling methods. J Appl Polym Sci 2015. [DOI: 10.1002/app.42700] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/10/2022]
Affiliation(s)
| | - Michele Mauri
- Department of Materials Science; University of Milano-Bicocca; 20125 Milan Italy
| | - Lucio Mauri
- Department of Materials Science; University of Milano-Bicocca; 20125 Milan Italy
| | - Gabriele Medaglia
- Pirelli Tyre S.p.A; Viale Piero e Alberto Pirelli, 25, 20126 Milan Italy
| | - Roberto Simonutti
- Department of Materials Science; University of Milano-Bicocca; 20125 Milan Italy
| |
Collapse
|
12
|
Zalesskiy SS, Danieli E, Blümich B, Ananikov VP. Miniaturization of NMR systems: desktop spectrometers, microcoil spectroscopy, and "NMR on a chip" for chemistry, biochemistry, and industry. Chem Rev 2014; 114:5641-94. [PMID: 24779750 DOI: 10.1021/cr400063g] [Citation(s) in RCA: 129] [Impact Index Per Article: 12.9] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/16/2022]
Affiliation(s)
- Sergey S Zalesskiy
- Zelinsky Institute of Organic Chemistry, Russian Academy of Sciences , Moscow, 119991, Russia
| | | | | | | |
Collapse
|
13
|
Washburn KE, Birdwell JE. Updated methodology for nuclear magnetic resonance characterization of shales. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2013; 233:17-28. [PMID: 23719372 DOI: 10.1016/j.jmr.2013.04.014] [Citation(s) in RCA: 24] [Impact Index Per Article: 2.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 02/06/2013] [Revised: 04/19/2013] [Accepted: 04/22/2013] [Indexed: 06/02/2023]
Abstract
Unconventional petroleum resources, particularly in shales, are expected to play an increasingly important role in the world's energy portfolio in the coming years. Nuclear magnetic resonance (NMR), particularly at low-field, provides important information in the evaluation of shale resources. Most of the low-field NMR analyses performed on shale samples rely heavily on standard T1 and T2 measurements. We present a new approach using solid echoes in the measurement of T1 and T1-T2 correlations that addresses some of the challenges encountered when making NMR measurements on shale samples compared to conventional reservoir rocks. Combining these techniques with standard T1 and T2 measurements provides a more complete assessment of the hydrogen-bearing constituents (e.g., bitumen, kerogen, clay-bound water) in shale samples. These methods are applied to immature and pyrolyzed oil shale samples to examine the solid and highly viscous organic phases present during the petroleum generation process. The solid echo measurements produce additional signal in the oil shale samples compared to the standard methodologies, indicating the presence of components undergoing homonuclear dipolar coupling. The results presented here include the first low-field NMR measurements performed on kerogen as well as detailed NMR analysis of highly viscous thermally generated bitumen present in pyrolyzed oil shale.
Collapse
|
14
|
Liberman A, Bergman E, Sarda Y, Nevo U. Faster imaging with a portable unilateral NMR device. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2013; 231:72-78. [PMID: 23597948 DOI: 10.1016/j.jmr.2013.03.009] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 12/24/2012] [Revised: 03/18/2013] [Accepted: 03/21/2013] [Indexed: 06/02/2023]
Abstract
Unilateral NMR devices are important tools in various applications such as non-destructive testing and well logging, but are not applied routinely for imaging, primarily because B0 inhomogeneity in these scanners leads to a relatively low signal and requires use of the slow single point imaging scan scheme. Enabling high quality, fast imaging could make this affordable and portable technology practical for various imaging applications as well as for new applications that are not yet feasible with MRI technology. The goal of this work was to improve imaging times in a portable unilateral NMR scanner. Both Compressed Sensing and Fast Spin Echo were modified and applied to fit the unique characteristics of a unilateral device. Two printed phantoms, allowing high resolution images, were scanned with both methods and compared to a standard scan and to a low pass scan to evaluate performance. Both methods were found to be feasible with a unilateral device, proving ways to accelerate single point imaging in such scanners. This outcome encourages us to explore how to further accelerate imaging times in unilateral NMR devices so that this technology might become clinically applicable in the future.
Collapse
Affiliation(s)
- Asaf Liberman
- Department of Biomedical Engineering, The Iby and Aladar Fleischman Faculty of Engineering, Tel Aviv University, Tel Aviv 69978, Israel
| | | | | | | |
Collapse
|
15
|
Affiliation(s)
- Alina Adams
- Institut für Technische und Makromolekulare Chemie; RWTH Aachen University; Templergraben 55 52056 Aachen Germany
| | - Bernhard Blümich
- Institut für Technische und Makromolekulare Chemie; RWTH Aachen University; Templergraben 55 52056 Aachen Germany
| |
Collapse
|
16
|
Goga NO, Demco DE, Kolz J, Ferencz R, Haber A, Casanova F, Blümich B. Surface UV aging of elastomers investigated with microscopic resolution by single-sided NMR. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2008; 192:1-7. [PMID: 18294890 DOI: 10.1016/j.jmr.2007.10.017] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/25/2007] [Revised: 10/27/2007] [Accepted: 10/30/2007] [Indexed: 05/25/2023]
Abstract
Depth profiles taken from the surface of UV irradiated natural rubber sheets have been measured with microscopic resolution using a Profile NMR-MOUSE. An NMR observable related to the sum of the spin echoes in the Carr-Purcell-Meiboom-Gill pulse sequence was used to characterize the cross-link density changes produced by the action of UV radiation in each sheet. The aging process was investigated as function of irradiation time and penetration depth. An exponential attenuation law with a space dependent absorption coefficient describes the change in the NMR observable with penetration depth. An Avrami model is used to describe the dependence of the absorption coefficient on the aging time. The method can be applied to investigate the effect of various aging agents on the surfaces of elastomers.
Collapse
Affiliation(s)
- N O Goga
- Institute for Technical and Macromolecular Chemistry, RWTH Aachen University, Worringerweg 1, D-52072 Aachen, Germany
| | | | | | | | | | | | | |
Collapse
|
17
|
Demas V, Herberg JL, Malba V, Bernhardt A, Evans L, Harvey C, Chinn SC, Maxwell RS, Reimer J. Portable, low-cost NMR with laser-lathe lithography produced microcoils. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2007; 189:121-9. [PMID: 17897853 DOI: 10.1016/j.jmr.2007.08.011] [Citation(s) in RCA: 18] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 02/13/2007] [Revised: 07/31/2007] [Accepted: 08/03/2007] [Indexed: 05/17/2023]
Abstract
Nuclear Magnetic Resonance (NMR) is unsurpassed in its ability to non-destructively probe chemical identity. Portable, low-cost NMR sensors would enable on-site identification of potentially hazardous substances, as well as the study of samples in a variety of industrial applications. Recent developments in RF microcoil construction (i.e. coils much smaller than the standard 5mm NMR RF coils), have dramatically increased NMR sensitivity and decreased the limits-of-detection (LOD). We are using advances in laser pantographic microfabrication techniques, unique to LLNL, to produce RF microcoils for field deployable, high sensitivity NMR-based detectors. This same fabrication technique can be used to produce imaging coils for MRI as well as for standard hardware shimming or "ex-situ" shimming of field inhomogeneities typically associated with inexpensive magnets. This paper describes a portable NMR system based on the use of a 2 kg hand-held permanent magnet, laser-fabricated microcoils, and a compact spectrometer. The main limitations for such a system are the low resolution and sensitivity associated with the low field values and quality of small permanent magnets, as well as the lack of large amounts of sample of interest in most cases. The focus of the paper is on the setting up of this system, initial results, sensitivity measurements, discussion of the limitations and future plans. The results, even though preliminary, are promising and provide the foundation for developing a portable, inexpensive NMR system for chemical analysis. Such a system will be ideal for chemical identification of trace substances on site.
Collapse
Affiliation(s)
- Vasiliki Demas
- Lawrence Livermore National Laboratory, Chemistry, Material, and Life Sciences Directorate, 7000 East Avenue, Livermore, CA 94550, USA
| | | | | | | | | | | | | | | | | |
Collapse
|
18
|
Herberg JL, Chinn SC, Sawvel AM, Gjersing E, Maxwell RS. Characterization of local deformation in filled-silicone elastomers subject to high strain – NMR MOUSE and Magnetic Resonance Imaging as a diagnostic tool for detection of inhomogeneities. Polym Degrad Stab 2006. [DOI: 10.1016/j.polymdegradstab.2005.12.006] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
|
19
|
Mowery DM, Assink RA, Celina M. Sensitivity of proton NMR relaxation times in a HTPB based polyurethane elastomer to thermo-oxidative aging. POLYMER 2005. [DOI: 10.1016/j.polymer.2005.08.093] [Citation(s) in RCA: 21] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/25/2022]
|
20
|
Wang M, Bertmer M, Demco DE, Blümich B. Segmental and Local Chain Mobilities in Elastomers by13C−1H Residual Heteronuclear Dipolar Couplings. J Phys Chem B 2004. [DOI: 10.1021/jp048392+] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/28/2022]
|
21
|
Saalwächter K. Detection of Heterogeneities in Dry and Swollen Polymer Networks by Proton Low-Field NMR Spectroscopy. J Am Chem Soc 2003; 125:14684-5. [PMID: 14640625 DOI: 10.1021/ja038278p] [Citation(s) in RCA: 56] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/28/2022]
Abstract
We report on the implementation of modern proton multiple-quantum NMR methods for the characterization of molecular order and dynamics in polymer networks and melts on cost-efficient low-resolution low-field NMR instrumentation. The method permits the extraction of chain order parameter distributions, and is therefore sensitive to a heterogeneity of cross-links and other topological constraints. Data from samples with bimodal network chain length distributions acquired at 20 MHz (0.47 T) are in quantitative agreement with results obtained at high field (500 MHz). It is shown for the first time that the chain order distribution is broadened upon swelling, providing evidence for the nontrivial nature of the swelling process.
Collapse
Affiliation(s)
- Kay Saalwächter
- Institut für Makromolekulare Chemie, Universität Freiburg, Stefan-Meier-Strasse 31, D-79104 Freiburg, Germany.
| |
Collapse
|
22
|
Pedersen HT, Ablett S, Martin DR, Mallett MJD, Engelsen SB. Application of the NMR-MOUSE to food emulsions. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2003; 165:49-58. [PMID: 14568516 DOI: 10.1016/s1090-7807(03)00243-x] [Citation(s) in RCA: 22] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/24/2023]
Abstract
The application of the NMR-MObile Universal Surface Explorer (NMR-MOUSE) to study food systems is evaluated using oil-in-water emulsions, and the results are compared to those obtained using a conventional low-field NMR (LF-NMR) instrument. The NMR-MOUSE is a small and portable LF-NMR system with a one-sided magnet layout that is used to replace the conventional magnet and probe on a LF-NMR instrument. The high magnetic field gradients associated with the one-sided MOUSE magnet result in NMR signal decays being dominated by molecular diffusion effects, which makes it possible to discriminate between the NMR signals from oil and water. Different data acquisition parameters as well as different approaches to the analysis of the NMR data from a range of oil-in-water emulsions are evaluated, and it is demonstrated how the concentration of oil and water can be determined from the NMR-MOUSE signals. From these model systems it is concluded that the NMR-MOUSE has good potential for the quantitative analysis of intact food products.
Collapse
Affiliation(s)
- H T Pedersen
- Centre for Advanced Food Studies, Department of Dairy and Food Science, Food, Technology, The Royal Veterinary and Agricultural University, 1958, Frederiksberg C, Denmark.
| | | | | | | | | |
Collapse
|
23
|
Casanova F, Blümich B. Two-dimensional imaging with a single-sided NMR probe. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2003; 163:38-45. [PMID: 12852905 DOI: 10.1016/s1090-7807(03)00123-x] [Citation(s) in RCA: 40] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/24/2023]
Abstract
A new low field unilateral NMR sensor equipped with a two-dimensional gradient coil system was built. A new NMR-MOUSE concept using a simple bar magnet instead of the classical U-shaped geometry was used to produce magnetic field profiles comparatively homogeneous in extended lateral planes defining a suitable field of view for 2D spatial localization. Slice selection along the depth direction is obtained by means of the highly constant static magnetic field gradient produced by this magnet geometry. Implementing a two-dimensional phase-encoding imaging method 2D cross sections of objects were obtained with high spatial resolution. By retuning the probe it was possible to change the depth of the selected slice obtaining a 3D imaging method. The details of the construction of the new device are presented together with imaging tests to show the quality of space encoding.
Collapse
Affiliation(s)
- F Casanova
- Institute for Technical Chemistry and Macromolecular Chemistry, Aachen University of Technology, RWTH, D-52056 Aachen, Germany
| | | |
Collapse
|
24
|
Variable rotation composite pulses for high resolution nuclear magnetic resonance using inhomogeneous magnetic and radiofrequency fields. Chem Phys Lett 2002. [DOI: 10.1016/s0009-2614(02)01116-8] [Citation(s) in RCA: 16] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/22/2022]
|
25
|
Heise H, Sakellariou D, Meriles CA, Moulé A, Pines A. Two-dimensional high-resolution NMR spectra in matched B(0) and B(1) field gradients. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2002; 156:146-151. [PMID: 12081452 DOI: 10.1006/jmre.2002.2545] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/23/2023]
Abstract
In a recent publication we presented a method to obtain highly resolved NMR spectra in the presence of an inhomogeneous B(0) field with the help of a matched RF gradient. If RF gradient pulses are combined with "ideal" 90 degrees pulses to form inhomogeneous z rotation pulses, the line broadening caused by the B(0) gradient can be refocused, while the full chemical shift information is maintained. This approach is of potential use for NMR spectroscopy in an inhomogeneous magnetic field produced by an "ex-situ" surface spectrometer. In this contribution, we extend this method toward two-dimensional spectroscopy with high resolution in one or both dimensions. Line narrowing in the indirect dimension can be achieved by two types of nutation echoes, thus leading to depth-sensitive NMR spectra with full chemical shift information. If the nutation echo in the indirect dimension is combined with a stroboscopic acquisition using inhomogeneous z-rotation pulses, highly resolved two-dimensional correlation spectra can be obtained in matched field gradients. Finally, we demonstrate that an INEPT coherence transfer from proton to carbon spins is possible in inhomogeneous B(0) fields. Thus, it is possible to obtain one-dimensional (13)C NMR spectra with increased sensitivity and two-dimensional HETCOR spectra in the presence of B(0) gradients of 0.4 mT/cm. These schemes may be of some value for ex-situ NMR analysis of materials and biological systems.
Collapse
Affiliation(s)
- Henrike Heise
- Materials Sciences Division, Lawrence Berkeley National Laboratory and Department of Chemistry, University of California-Berkeley, Berkeley, California 94720, USA
| | | | | | | | | |
Collapse
|
26
|
|
27
|
Balibanu F, Hailu K, Eymael R, Demco DE, Blumich B. Nuclear magnetic resonance in inhomogeneous magnetic fields. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2000; 145:246-258. [PMID: 10910693 DOI: 10.1006/jmre.2000.2089] [Citation(s) in RCA: 25] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/23/2023]
Abstract
The response of the spin system has been investigated by numerical simulations in the case of a nuclear magnetic resonance (NMR) experiment performed in inhomogeneous static and radiofrequency fields. The particular case of the NMR-MOUSE was considered. The static field and the component of the radiofrequency field perpendicular to the static field were evaluated as well as the spatial distribution of the maximum NMR signal detected by the surface coil. The NMR response to various pulse sequences was evaluated numerically for the case of an ensemble of isolated spins (1/2). The behavior of the echo train in Carr-Purcell-like pulse sequences used for measurements of transverse relaxation and self-diffusion was simulated and compared with the experiment. The echo train is shown to behave qualitatively differently depending on the particular phase schemes used in these pulse sequences. Different echo trains are obtained, because of the different superposition of Hahn and stimulated echoes forming mixed echoes as a result of the spatial distribution of pulse flip angles. The superposition of Hahn and stimulated echoes originating from different spatial regions leads to distortions of the mixed echoes in intensity, shape, and phase. The volume selection produced by Carr-Purcell-like pulse sequences is also investigated for the NMR-MOUSE. The developed numerical simulation procedure is useful for understanding a variety of experiments performed with the NMR-MOUSE and for improving its performance. Copyright 2000 Academic Press.
Collapse
Affiliation(s)
- F Balibanu
- Institut fur Technische Chemie und Makromolekulare Chemie, Rheinisch-Westfalische Technische Hochschule, Worringerweg 1, Aachen, D-52056, Germany
| | | | | | | | | |
Collapse
|
28
|
Prado PJ, Blümich B, Schmitz U. One-dimensional imaging with a palm-size probe. JOURNAL OF MAGNETIC RESONANCE (SAN DIEGO, CALIF. : 1997) 2000; 144:200-206. [PMID: 10828188 DOI: 10.1006/jmre.2000.2038] [Citation(s) in RCA: 15] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/23/2023]
Abstract
A new portable magnetic resonance imaging device was built. Spatially resolved NMR was achieved by placing a gradient coil pair and a Helmholtz pair type radio-frequency probe in the gap between two antiparallel polarized permanent magnets. The flat face of the low-field (nu(proton) = 20 MHz) apparatus allowed for the study of arbitrarily large objects in situ. Relaxation time weighted 1D images were achieved over a 15-mm field of view by a single-point spin-echo sequence. A phase encoding time on the order of 200 micros permited the scanning of a wide range of heterogeneous materials.
Collapse
Affiliation(s)
- P J Prado
- Quantum Magnetics, Inc., 7740 Kenamar Court, San Diego, California, 92121, USA
| | | | | |
Collapse
|
29
|
Grinberg F, Garbarczyk M, Kuhn W. Influence of the cross-link density and the filler content on segment dynamics in dry and swollen natural rubber studied by the NMR dipolar-correlation effect. J Chem Phys 1999. [DOI: 10.1063/1.480478] [Citation(s) in RCA: 33] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022] Open
|