1
|
Kunkyab T, Lakrad K, Jirasek A, Oldham M, Quinn B, Hyde D, Adamson J. Clinical applicability of Linac-integrated CBCT based NIPAM 3D dosimetry: a dual-institutional investigation. Phys Med Biol 2024; 69:155002. [PMID: 38959910 DOI: 10.1088/1361-6560/ad5eef] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/18/2023] [Accepted: 07/03/2024] [Indexed: 07/05/2024]
Abstract
Objective.To develop and benchmark a novel 3D dose verification technique consisting of polymer gel dosimetry (PGD) with cone-beam-CT (CBCT) readout through a two-institution study. The technique has potential for wide and robust applicability through reliance on CBCT readout.Approach. Three treatment plans (3-field, TG119-C-shape spine, 4-target SRS) were created by two independent institutions (Institutions A and B). A Varian Truebeam linear accelerator was used to deliver the plans to NIPAM polymer gel dosimeters produced at both institutions using an identical approach. For readout, a slow CBCT scan mode was used to acquire pre- and post-irradiation images of the gel (1 mm slice thickness). Independent gel analysis tools were used to process the PGD images (A: VistaAce software, B: in-house MATLAB code). Comparing planned and measured doses, the analysis involved a combination of 1D line profiles, 2D contour plots, and 3D global gamma maps (criteria ranging between 2%1 mm and 5%2 mm, with a 10% dose threshold).Main results. For all gamma criteria tested, the 3D gamma pass rates were all above 90% for 3-field and 88% for the SRS plan. For the C-shape spine plan, we benchmarked our 2% 2 mm result against previously published work using film analysis (93.4%). For 2%2 mm, 99.4% (Institution A data), and 89.7% (Institution B data) were obtained based on VistaAce software analysis, 83.7% (Institution A data), and 82.9% (Institution B data) based on MATLAB.Significance. The benchmark data demonstrate that when two institutions follow the same rigorous procedures gamma passing rates up to 99%, for 2%2 mm criteria can be achieved for substantively different treatment plans. The use of different software and calibration techniques may have contributed to the variation in the 3D gamma results. By sharing the data across institutions, we observe the gamma passing rate is more consistent within each pipeline, indicating the need for standardized analysis methods.
Collapse
Affiliation(s)
- Tenzin Kunkyab
- Department of Physics, University of British Columbia-Okanagan Campus, Kelowna, British Columbia, Canada
- BC Cancer Center, Kelowna, British Columbia
| | - Kawtar Lakrad
- Department of Physics, Hassan II University, Casablanca, Morocco
- Duke University Medical Center, Durham, NC
| | - Andrew Jirasek
- Department of Physics, University of British Columbia-Okanagan Campus, Kelowna, British Columbia, Canada
| | | | - Benjamin Quinn
- Modus Medical Devices Inc./IBA Dosimetry, London, Ontario, Canada
| | - Derek Hyde
- Department of Physics, University of British Columbia-Okanagan Campus, Kelowna, British Columbia, Canada
- BC Cancer Center, Kelowna, British Columbia
| | | |
Collapse
|
2
|
De Deene Y. Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D. Gels 2022; 8:599. [PMID: 36135311 PMCID: PMC9498652 DOI: 10.3390/gels8090599] [Citation(s) in RCA: 16] [Impact Index Per Article: 8.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/19/2022] [Revised: 09/07/2022] [Accepted: 09/10/2022] [Indexed: 12/22/2022] Open
Abstract
Gel dosimetry was developed in the 1990s in response to a growing need for methods to validate the radiation dose distribution delivered to cancer patients receiving high-precision radiotherapy. Three different classes of gel dosimeters were developed and extensively studied. The first class of gel dosimeters is the Fricke gel dosimeters, which consist of a hydrogel with dissolved ferrous ions that oxidize upon exposure to ionizing radiation. The oxidation results in a change in the nuclear magnetic resonance (NMR) relaxation, which makes it possible to read out Fricke gel dosimeters by use of quantitative magnetic resonance imaging (MRI). The radiation-induced oxidation in Fricke gel dosimeters can also be visualized by adding an indicator such as xylenol orange. The second class of gel dosimeters is the radiochromic gel dosimeters, which also exhibit a color change upon irradiation but do not use a metal ion. These radiochromic gel dosimeters do not demonstrate a significant radiation-induced change in NMR properties. The third class is the polymer gel dosimeters, which contain vinyl monomers that polymerize upon irradiation. Polymer gel dosimeters are predominantly read out by quantitative MRI or X-ray CT. The accuracy of the dosimeters depends on both the physico-chemical properties of the gel dosimeters and on the readout technique. Many different gel formulations have been proposed and discussed in the scientific literature in the last three decades, and scanning methods have been optimized to achieve an acceptable accuracy for clinical dosimetry. More recently, with the introduction of the MR-Linac, which combines an MRI-scanner and a clinical linear accelerator in one, it was shown possible to acquire dose maps during radiation, but new challenges arise.
Collapse
Affiliation(s)
- Yves De Deene
- Liverpool & Macarthur Cancer Therapy Centres, Liverpool, NSW 1871, Australia; or
- Ingham Institute, Liverpool, NSW 2170, Australia
- School of Science, Western Sydney University, Penrith, NSW 2751, Australia
| |
Collapse
|
3
|
Maras P, Jaszczak M, Kozicki M. Basic features of VIC-T dosimeter with spiral CT readout, CT scanning conditions and data processing with a new polyGeVero-CT software package. Radiat Phys Chem Oxf Engl 1993 2021. [DOI: 10.1016/j.radphyschem.2021.109730] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
|
4
|
Natanasabapathi G, Warmington L, Watanabe Y. Evaluation of two calibration methods for MRI-based polymer gel dosimetry. Appl Radiat Isot 2021; 174:109754. [PMID: 34030113 DOI: 10.1016/j.apradiso.2021.109754] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/06/2020] [Revised: 04/22/2021] [Accepted: 04/23/2021] [Indexed: 10/21/2022]
Abstract
Polymer gel dosimetry (PGD) can provide three-dimensional (3D) dose data for evaluation of the dose calculation algorithms used by treatment planning systems (TPS). Although the PGD technique, particularly with MRI, is now ready for clinical applications, an accurate calibration method is vital for treatment validation in 3D. This study evaluated the single-phantom electron beam (SPE) method that used the depth-dose data of a 9 MeV electron beam. This technique was compared with the multi-vial x-ray (MVX) method that used nine small vials irradiated with various doses. We tested two regression equations, i.e., third-order polynomial and tangent functions, and two dose-normalization methods, i.e., one-point and two-point methods. These methods were evaluated using a dose distribution generated by a 3 cm × 3 cm open arc beam. We used MAGAT polymer gel manufactured in-house. We found that the SPE method required a smaller dose scaling for the dose comparison. The tangent function showed better data fitting than the polynomial function with smaller uncertainty of the estimated coefficients. We did not observe a distinct advantage of the SPE method over the MVX method for the 3D dose comparison with the test case. From this study, we infer that the SPE method with the tangent function as the regression equation and one-point dose normalization is a good calibration option for the MRI-based polymer gel dosimetry.
Collapse
Affiliation(s)
- Gopishankar Natanasabapathi
- Department of Radiation Oncology, Dr. B. R. A. IRCH, All India Institute of Medical Sciences, New Delhi, India.
| | - Leighton Warmington
- Department of Radiation Oncology, University of Minnesota, Minneapolis, MN, USA
| | - Yoichi Watanabe
- Department of Radiation Oncology, University of Minnesota, Minneapolis, MN, USA
| |
Collapse
|
5
|
Maynard E, Heath E, Hilts M, Jirasek A. Evaluation of an x-ray CT polymer gel dosimetry system in the measurement of deformed dose. Biomed Phys Eng Express 2020; 6:035031. [PMID: 33438676 DOI: 10.1088/2057-1976/ab895a] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Abstract
This study is an evaluation of the use of a N-isopropylacrylamide (NIPAM)-based x-ray CT polymer gel dosimetry (PGD) system in the measurement of deformed dose. This work also compares dose that is measured by the gel dosimetry system to dose calculated by a novel deformable dose accumulation algorithm, defDOSXYZnrc, that uses direct voxel tracking. Deformable gels were first irradiated using a single 3.5 × 5 cm2 open field and the static dose was compared to defDOSXYZnrc as a control measurement. Gel measurement was found to be in excellent agreement with defDOSXYZnrc in the static case with gamma passing rates of 94.5% using a 3%/3 mm criterion and 93.3% using a 3%/2 mm criterion. Following the static measurements, a deformable gel was irradiated with the same single field under an external compression of 25 mm and then released from this compression for dosimetric read out. The measured deformed dose was then compared to deformed dose calculated by defDOSXYZnrc based on deformation vectors produced by the Velocity AI deformable image registration (DIR) algorithm. In the deformed dose distribution there were differences in the measured and calculated field position of up to 0.8 mm and differences in the measured in calculated field size of up to 11.9 mm. Gamma pass rates were 60.0% using a 3%/3 mm criterion and 56.8% using a 3%/2 mm criterion for the deforming measurements representing a decrease in agreement compared to the control measurements. Further analysis showed that passing rates increased to 86.5% using a 3%/3 mm criterion and 70.5% using a 3%/2 mm criterion in voxels within 5 mm of fiducial markers used to guide the deformable image registration. This work represents the first measurement of deformed dose using x-ray CT polymer gel dosimetry. Overall these results highlight some of the challenges in the calculation and measurement of deforming dose and provide insight into possible strategies for improvement.
Collapse
Affiliation(s)
- E Maynard
- Department of Physics and Astronomy, University of Victoria, Victoria, BC V8W 2Y2, Canada
| | | | | | | |
Collapse
|
6
|
Kozicki M, Sąsiadek E, Kadlubowski S, Dudek M, Karbownik I. Radiation sensitive polyacrylonitrile microfibres doped with PDA nano-particles. Radiat Phys Chem Oxf Engl 1993 2020. [DOI: 10.1016/j.radphyschem.2018.01.021] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/18/2022]
|
7
|
Clinical radiotherapy application of N-vinylpyrrolidone-containing 3D polymer gel dosimeters with remote external MR-reading. Phys Med 2020; 69:134-146. [PMID: 31901838 DOI: 10.1016/j.ejmp.2019.11.014] [Citation(s) in RCA: 20] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/05/2019] [Revised: 10/25/2019] [Accepted: 11/15/2019] [Indexed: 11/20/2022] Open
Abstract
PURPOSE Advanced 3D dosimetry is required for verifications of complex dose distributions in modern radiotherapy. Two 3D polymer gel dosimeters, coupled with magnetic resonance (MR) imaging (3 T MRI) readout and data processing with polyGeVero® software, were tested for the verification of calculated 3D dose distributions by a treatment planning system (TPS) and ArcCHECK®-3DVH®, related to eradication of a lung tumour. METHODS N-vinylpyrrolidone-containing 3D polymer gel dosimeters were used: VIC (containing ascorbic acid and copper sulfate pentahydrate) and VIC-T (containing tetrakis(hydroxymethyl)phosphonium chloride). Three remote centers were involved in the dosimeters preparation and irradiation (Poland), and MRI (Austria). Cross beam calibration of the dosimeters and verification of a 3D dose distribution calculated with an Eclipse External Beam TPS and ArcCHECK®-3DVH® were performed. The 3D-to-3D comparisons of the VIC and VIC-T with TPS and ArcCHECK®-3DVH® along with ArcCHECK®-3DVH® versus TPS dose matrixes were performed with the aid of the polyGeVero® by analyzing dose profiles, isodoses lines, gamma index, gamma angle, dose difference, and related histograms. RESULTS The measured MR-relaxation rate (R2 = 1/T2) for the dosimeters relates to the dose, as follows: R2 = 0.0928 ± 0.0008 [Gy-1 s-1] × D [Gy] + 2.985 ± 0.012 [s-1] (VIC) and 0.1839 ± 0.0044 [Gy-1 s-1] × D [Gy] + 2.519 ± 0.053 [s-1] (VIC-T). The 3D-to-3D comparisons revealed a good agreement between the measured and calculated 3D dose distributions. CONCLUSIONS VIC and VIC-T with 3T MRI readout and polyGeVero® showed potential for verifications of calculated irradiation plans. The results obtained suggest the implementation of the irradiation plan for eradication of the lung tumour.
Collapse
|
8
|
BrahimiMoussa S, Benamar M, LounisMokrani Z. Characterization of the chemical and structural modifications induced by gamma rays on the MAGIC polymer gel. Radiat Phys Chem Oxf Engl 1993 2020. [DOI: 10.1016/j.radphyschem.2019.108451] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/26/2022]
|
9
|
Kouvati K, Jaszczak M, Papagiannis P, Kadlubowski S, Wach R, Maras P, Dudek M, Kozicki M. Leuco crystal violet-Pluronic F-127 3D radiochromic gel dosimeter. ACTA ACUST UNITED AC 2019; 64:175017. [DOI: 10.1088/1361-6560/ab2f5d] [Citation(s) in RCA: 17] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
10
|
Maynard E, Heath E, Hilts M, Jirasek A. Introduction of a deformable x-ray CT polymer gel dosimetry system. Phys Med Biol 2018. [PMID: 29515019 DOI: 10.1088/1361-6560/aab501] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/16/2023]
Abstract
This study introduces the first 3D deformable dosimetry system based on x-ray computed tomography (CT) polymer gel dosimetry and establishes the setup reproducibility, deformation characteristics and dose response of the system. A N-isopropylacrylamide (NIPAM)-based gel formulation optimized for x-ray CT gel dosimetry was used, with a latex balloon serving as the deformable container and low-density polyethylene and polyvinyl alcohol providing additional oxygen barrier. Deformable gels were irradiated with a 6 MV calibration pattern to determine dosimetric response and a dosimetrically uniform plan to determine the spatial uniformity of the response. Wax beads were added to each gel as fiducial markers to track the deformation and setup of the gel dosimeters. From positions of the beads on CT images the setup reproducibility and the limits and reproducibility of gel deformation were determined. Comparison of gel measurements with Monte Carlo dose calculations found excellent dosimetric accuracy, comparable to that of an established non-deformable dosimetry system, with a mean dose discrepancy of 1.5% in the low-dose gradient region and a gamma pass rate of 97.9% using a 3%/3 mm criterion. The deformable dosimeter also showed good overall spatial dose uniformity throughout the dosimeter with some discrepancies within 20 mm of the edge of the container. Tracking of the beads within the dosimeter found that sub-millimetre setup accuracy is achievable with this system. The dosimeter was able to deform and relax when externally compressed by up to 30 mm without sustaining any permanent damage. Internal deformations in 3D produced average marker movements of up to 12 mm along the direction of compression. These deformations were also shown to be reproducible over 100 consecutive deformations. This work has established several important characteristics of a new deformable dosimetry system which shows promise for future clinical applications, including the validation of deformable dose accumulation algorithms.
Collapse
Affiliation(s)
- E Maynard
- Department of Physics and Astronomy, University of Victoria, Victoria, BC V8W 2Y2, Canada
| | | | | | | |
Collapse
|
11
|
Abdel-Fattah A, Beshir W, Hassan H, Soliman Y. Radiation-induced coloration of nitro blue tetrazolium gel dosimeter for low dose applications. RADIAT MEAS 2017. [DOI: 10.1016/j.radmeas.2017.04.001] [Citation(s) in RCA: 17] [Impact Index Per Article: 2.4] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
|
12
|
Maynard E, Hilts M, Heath E, Jirasek A. Evaluation of accuracy and precision in polymer gel dosimetry. Med Phys 2017; 44:736-746. [PMID: 28035662 DOI: 10.1002/mp.12080] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/26/2016] [Revised: 11/22/2016] [Accepted: 12/17/2016] [Indexed: 01/01/2023] Open
Abstract
PURPOSE To assess the overall reproducibility and accuracy of an X-ray computed tomography (CT) polymer gel dosimetry (PGD) system and investigate what effects the use of generic, interbatch, and intrabatch gel calibration have on dosimetric and spatial accuracy. METHODS A N-isopropylacrylamide (NIPAM)-based gel formulation optimized for X-ray CT gel dosimetry was used, and the results over four different batches of gels were analyzed. All gels were irradiated with three 6 MV beams in a calibration pattern at both the bottom and top of the dosimeter. Postirradiation CT images of the gels were processed using background subtraction, image averaging, adaptive mean filtering, and remnant artifact removal. The gel dose distributions were calibrated using a Monte Carlo (Vancouver Island Monte Carlo system) calculated dose distribution of the calibration pattern. Using the calibration results from all gels, an average or "generic" calibration curve was calculated and this generic calibration curve was used to calibrate each of the gels within the sample. For each of the gels, the irradiation pattern at the bottom of the dosimeter was also calibrated using the irradiation pattern at the top of the dosimeter to evaluate intragel calibration. RESULTS Comparison of gel measurements with Monte Carlo dose calculations found excellent dosimetric accuracy when using an average (or generic) calibration with a mean dose discrepancy of 1.8% in the low-dose gradient region which compared to a "best-case scenario" self-calibration method with a mean dose discrepancy of 1.6%. The intragel calibration method investigated produced large dose discrepancies due to differences in dose response at the top and bottom of the dosimeter, but the use of a dose-dependent correction reduced these dose errors. Spatial accuracy was found to be excellent for the average calibration method with a mean distance-to-agreement (DTA) of 0.63 mm and 99.6% of points with a DTA < 2 mm in high-dose gradient regions. This compares favorably to the self-calibration method which produced a mean DTA of 0.61 mm and 99.8% of points with a DTA < 2 mm. Gamma analysis using a 3%/3 mm criterion also found good agreement between the gel measurement and Monte Carlo dose calculation when using either the average calibration or self-calibration methods (96.8% and 98.2%, respectively). CONCLUSIONS An X-ray CT PGD system was evaluated and found to have excellent dosimeteric and spatial accuracy when compared to Monte Carlo dose calculations and the use of generic and interbatch calibration methods were found to be effective. The establishment of the accuracy and reproducibility of this system provides important information for clinical implementation.
Collapse
Affiliation(s)
- Evan Maynard
- Department of Physics and Astronomy, University of Victoria, Victoria, BC, V8W 2Y2, Canada
| | - Michelle Hilts
- Department of Physics, University of British Columbia-Okanagan Campus, Kelowna, BC, V1V 1V7, Canada.,Medical Physics, BC Cancer Agency, Centre for the Southern Interior, Kelowna, BC, V1Y 5L3, Canada
| | - Emily Heath
- Carleton Laboratory for Radiotherapy Physics, Department of Physics, Carleton University, Ottawa, ON, K1S 5B6, Canada
| | - Andrew Jirasek
- Department of Physics, I.K. Barber School of Arts and Sciences, University of British Columbia, Kelowna, BC V1V 1V7, Canada
| |
Collapse
|
13
|
Kakakhel MB, Jirasek A, Johnston H, Kairn T, Trapp JV. Improving the quality of reconstructed X-ray CT images of polymer gel dosimeters: zero-scan coupled with adaptive mean filtering. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2017; 40:159-165. [PMID: 28168587 DOI: 10.1007/s13246-017-0523-9] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 12/11/2015] [Accepted: 01/12/2017] [Indexed: 11/25/2022]
Abstract
This study evaluated the feasibility of combining the 'zero-scan' (ZS) X-ray computed tomography (CT) based polymer gel dosimeter (PGD) readout with adaptive mean (AM) filtering for improving the signal to noise ratio (SNR), and to compare these results with available average scan (AS) X-ray CT readout techniques. NIPAM PGD were manufactured, irradiated with 6 MV photons, CT imaged and processed in Matlab. AM filter for two iterations, with 3 × 3 and 5 × 5 pixels (kernel size), was used in two scenarios (a) the CT images were subjected to AM filtering (pre-processing) and these were further employed to generate AS and ZS gel images, and (b) the AS and ZS images were first reconstructed from the CT images and then AM filtering was carried out (post-processing). SNR was computed in an ROI of 30 × 30 for different pre and post processing cases. Results showed that the ZS technique combined with AM filtering resulted in improved SNR. Using the previously-recommended 25 images for reconstruction the ZS pre-processed protocol can give an increase of 44% and 80% in SNR for 3 × 3 and 5 × 5 kernel sizes respectively. However, post processing using both techniques and filter sizes introduced blur and a reduction in the spatial resolution. Based on this work, it is possible to recommend that the ZS method may be combined with pre-processed AM filtering using appropriate kernel size, to produce a large increase in the SNR of the reconstructed PGD images.
Collapse
Affiliation(s)
- M B Kakakhel
- Department of Physics and Applied Mathematics (DPAM), Pakistan Institute of Engineering and Applied Sciences, Nilore, 45450, Islamabad, Pakistan
| | - A Jirasek
- I.K. Barber School of Arts and Science, University of British Columbia, Okanagan Campus, Vancouver, BC, Canada
| | - H Johnston
- Department of Physics and Astronomy, University of Victoria, Victoria, BC, V8W 2Y2, Canada
| | - T Kairn
- Genesis Cancer Care Queensland, Brisbane, QLD, Australia.
- Science and Engineering Faculty, Queensland University of Technology, Brisbane, QLD, Australia.
| | - J V Trapp
- Science and Engineering Faculty, Queensland University of Technology, Brisbane, QLD, Australia
| |
Collapse
|
14
|
Kozicki M, Jaszczak M, Maras P, Dudek M, Cłapa M. On the development of a VIPARndradiotherapy 3D polymer gel dosimeter. Phys Med Biol 2017; 62:986-1008. [DOI: 10.1088/1361-6560/aa5089] [Citation(s) in RCA: 42] [Impact Index Per Article: 6.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
|
15
|
Cheng KY, Hsieh LL, Shih CT. A Comprehensive Evaluation of NIPAM Polymer Gel Dosimeters on Three Orthogonal Planes and Temporal Stability Analysis. PLoS One 2016; 11:e0155797. [PMID: 27192217 PMCID: PMC4871428 DOI: 10.1371/journal.pone.0155797] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/04/2015] [Accepted: 05/04/2016] [Indexed: 01/01/2023] Open
Abstract
Polymer gel dosimeters have been proven useful for dose evaluation in radiotherapy treatments. Previous studies have demonstrated that using a polymer gel dosimeter requires a 24 h reaction time to stabilize and further evaluate the measured dose distribution in two-dimensional dosimetry. In this study, the short-term stability within 24 h and feasibility of N-isopropylacrylamide (NIPAM) polymer gel dosimeters for use in three-dimensional dosimetry were evaluated using magnetic resonance imaging (MRI). NIPAM gels were used to measure the dose volume in a clinical case of intensity-modulated radiation therapy (IMRT). For dose readouts, MR images of irradiated NIPAM gel phantoms were acquired at 2, 5, 12, and 24 h after dose delivery. The mean standard errors of dose conversion from using dose calibration curves (DRC) were calculated. The measured dose volumes at the four time points were compared with those calculated using a treatment planning system (TPS). The mean standard errors of the dose conversion from using the DRCs were lower than 1 Gy. Mean pass rates of 2, 5, 12, and 24 h axial dose maps calculated using gamma evaluation with 3% dose difference and 3 mm distance-to-agreement criteria were 83.5% ± 0.9%, 85.9% ± 0.6%, 98.7% ± 0.3%, and 98.5% ± 0.9%, respectively. Compared with the dose volume histogram of the TPS, the absolute mean relative volume differences of the 2, 5, 12, and 24 h measured dose volumes were lower than 1% for the irradiated region with an absorbed dose higher than 2.8 Gy. It was concluded that a 12 h reaction time was sufficient to acquire accurate dose volume using the NIPAM gels with MR readouts.
Collapse
Affiliation(s)
- Kai-Yuan Cheng
- Department of Medical Imaging and Radiological Sciences, Central Taiwan University of Science and Technology, Taichung, Taiwan
- * E-mail:
| | - Ling-Ling Hsieh
- Department of Medical Imaging and Radiological Sciences, Central Taiwan University of Science and Technology, Taichung, Taiwan
- Graduate Institute of Pharmaceutical Science and Technology, Central Taiwan University of Science and Technology, Taichung, Taiwan
| | - Cheng-Ting Shih
- 3D Printing Medical Research Center, China Medical University Hospital, China Medical University, Taichung, Taiwan
| |
Collapse
|
16
|
Johnston H, Hilts M, Jirasek A. Incorporating multislice imaging into x-ray CT polymer gel dosimetry. Med Phys 2015; 42:1666-77. [PMID: 25832056 DOI: 10.1118/1.4914419] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE To evaluate multislice computed tomography (CT) scanning for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD) and to establish a baseline assessment of image noise and uniformity in an unirradiated gel dosimeter. METHODS A 16-slice CT scanner was used to acquire images through a 1 L cylinder filled with water. Additional images were collected using a single slice machine. The variability in CT number (NCT) associated with the anode heel effect was evaluated and used to define a new slice-by-slice background subtraction artifact removal technique for CT PGD. Image quality was assessed for the multislice system by evaluating image noise and uniformity. The agreement in NCT for slices acquired simultaneously using the multislice detector array was also examined. Further study was performed to assess the effects of increasing x-ray tube load on the constancy of measured NCT and overall scan time. In all cases, results were compared to the single slice machine. Finally, images were collected throughout the volume of an unirradiated gel dosimeter to quantify image noise and uniformity before radiation is delivered. RESULTS Slice-by-slice background subtraction effectively removes the variability in NCT observed across images acquired simultaneously using the multislice scanner and is the recommended background subtraction method when using a multislice CT system. Image noise was higher for the multislice system compared to the single slice scanner, but overall image quality was comparable between the two systems. Further study showed NCT was consistent across image slices acquired simultaneously using the multislice detector array for each detector configuration of the slice thicknesses examined. In addition, the multislice system was found to eliminate variations in NCT due to increasing x-ray tube load and reduce scanning time by a factor of 4 when compared to imaging a large volume using a single slice scanner. Images acquired through an unirradiated, active gel revealed NCT varies between the top and bottom of the 1 L cylinder as well as across the diameter of the cylinder by up to 7 HU. CONCLUSIONS Multislice CT imaging has been evaluated for CT PGD and found to be the superior technique compared to single slice imaging in terms of the time required to complete a scan and the tube load characteristics associated with each scanning method. The implementation of multislice scanning is straightforward and expected to facilitate routine gel dosimetry measurements for complex dose distributions in modern RT centers.
Collapse
Affiliation(s)
- H Johnston
- Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2, Canada
| | - M Hilts
- Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2, Canada and Medical Physics, BC Cancer Agency, Vancouver Island Centre, Victoria, British Columbia V8R 6V5, Canada
| | - A Jirasek
- Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2, Canada and Department of Physics, University of British Columbia-Okanagan Campus, Kelowna, British Columbia V1V 1V7, Canada
| |
Collapse
|
17
|
|