1
|
Guljakow J, Lang W. Analysis of the Lifetime of Neural Implants Using In Vitro Test Structures. SENSORS (BASEL, SWITZERLAND) 2023; 23:6263. [PMID: 37514558 PMCID: PMC10385412 DOI: 10.3390/s23146263] [Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/22/2023] [Revised: 06/28/2023] [Accepted: 07/06/2023] [Indexed: 07/30/2023]
Abstract
The aim of this work was to measure the lifetime of neural implant test samples at two different temperatures, using a method that allows the precise measurement of the sample lifetime, further analysis with the use of Weibull statistics, and examination of the applicability of the Van't Hoff rule. The correct estimation of the lifetime of neural implants is important to avoid preliminary failures, when used in humans. The novelty lies in the precise data due to the measurement approach, the application of the Weibull statistics to neural test samples, and the examination of the Van't Hoff rule's applicability to the longevity of polyimide-based neural implant samples. Several samples that consisted of interdigitated gold strands, encapsulated in polyimide were soaked in ringer solution. One batch was soaked at a temperature of 37 °C, and another was soaked at a temperature of 57 °C. Voltage was applied and measured to identify the occurrence of failures. The long-term experiment was stopped after 458 days for the samples at 37 °C and 423 days for the samples at 57 °C, with several samples still being intact at both temperature levels. The time to failure was measured and used to identify the Weibull parameters that would describe the behavior of the samples. The median lifetime of the samples changed from 363 days at 37 °C to 138 days at 57 °C. The scale and shape factor changed from 396 and 3.7 at 37 °C to 138 and 2 at 57 °C, respectively. The measured mean, median times, and Weibull scale factors were lower than expected from the Van't Hoff rule. The use of the Van't hoff rule with 2ΔT/10°C for accelerated lifetime tests would lead to an estimation of longer lifetimes than realistic. A reaction rate constant around 1.47 appears more appropriate. While a fourfold difference in lifetime would be expected, only a 2.65-fold difference in the median lifetime and a roughly 2.2-fold difference in the mean and Weibull scale factor were observed. The shift of the Weibull shape parameter from 3.7 at 37 °C to 2 at 57 °C with rising temperatures was observed, indicating differences in failure reasons and stronger aging at lower temperatures. The used method is simple to apply and interpret and allows for a precise anticipation of sample lifetimes.
Collapse
Affiliation(s)
- Jürgen Guljakow
- Institute for Microsensors, Actuators and Systems (IMSAS), University of Bremen, Otto-Hahn-Allee 1, 28359 Bremen, Germany
| | - Walter Lang
- Institute for Microsensors, Actuators and Systems (IMSAS), University of Bremen, Otto-Hahn-Allee 1, 28359 Bremen, Germany
| |
Collapse
|
2
|
Thin Film Encapsulation for LCP-Based Flexible Bioelectronic Implants: Comparison of Different Coating Materials Using Test Methodologies for Life-Time Estimation. MICROMACHINES 2022; 13:mi13040544. [PMID: 35457851 PMCID: PMC9028940 DOI: 10.3390/mi13040544] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 02/25/2022] [Revised: 03/22/2022] [Accepted: 03/28/2022] [Indexed: 02/01/2023]
Abstract
Liquid crystal polymer (LCP) has gained wide interest in the electronics industry largely due to its flexibility, stable insulation and dielectric properties and chip integration capabilities. Recently, LCP has also been investigated as a biocompatible substrate for the fabrication of multielectrode arrays. Realizing a fully implantable LCP-based bioelectronic device, however, still necessitates a low form factor packaging solution to protect the electronics in the body. In this work, we investigate two promising encapsulation coatings based on thin-film technology as the main packaging for LCP-based electronics. Specifically, a HfO2–based nanolaminate ceramic (TFE1) deposited via atomic layer deposition (ALD), and a hybrid Parylene C-ALD multilayer stack (TFE2), both with a silicone finish, were investigated and compared to a reference LCP coating. T-peel, water-vapour transmission rate (WVTR) and long-term electrochemical impedance spectrometry (EIS) tests were performed to evaluate adhesion, barrier properties and overall encapsulation performance of the coatings. Both TFE materials showed stable impedance characteristics while submerged in 60 °C saline, with TFE1-silicone lasting more than 16 months under a continuous 14V DC bias (experiment is ongoing). The results presented in this work show that WVTR is not the main factor in determining lifetime, but the adhesion of the coating to the substrate materials plays a key role in maintaining a stable interface and thus longer lifetimes.
Collapse
|
3
|
Oldroyd P, Malliaras GG. Achieving long-term stability of thin-film electrodes for neurostimulation. Acta Biomater 2022; 139:65-81. [PMID: 34020055 DOI: 10.1016/j.actbio.2021.05.004] [Citation(s) in RCA: 22] [Impact Index Per Article: 11.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/03/2021] [Revised: 05/06/2021] [Accepted: 05/06/2021] [Indexed: 12/17/2022]
Abstract
Implantable electrodes that can reliably measure brain activity and deliver an electrical stimulus to a target tissue are increasingly employed to treat various neurological diseases and neuropsychiatric disorders. Flexible thin-film electrodes have gained attention over the past few years to minimise invasiveness and damage upon implantation. Research has previously focused on optimising the electrode's electrical and mechanical properties; however, their chronic stability must be validated to translate electrodes from a research to a clinical application. Neurostimulation electrodes, which actively inject charge, have yet to reliably demonstrate continuous functionality for ten years or more in vivo, the accepted metric for clinical viability. Long-term stability can only be achieved if the focus switches to investigating how and why such devices fail. Unfortunately, there is a field-wide reluctance to investigate device stability and failures, which hinders device optimisation. This review surveys thin-film electrode designs with a focus on adhesion between electrode layers and the interactions with the surrounding environment. A comprehensive summary of the abiotic failure modes faced by such electrodes is presented, and to encourage investigation, systematic methods for analysing their origin are recommended. Finally, approaches to reducing the likelihood of device failure are offered. STATEMENT OF SIGNIFICANCE: Neural electrodes that can deliver an electrical stimulus to a target tissue are widely used to treat various neurological diseases. Essential to the function of these electrodes is the ability to safely stimulate the target tissue for extended periods (> 10 years); however, this has not yet been clinically achieved. The key to achieving long-term stability is an increased understanding of electrode interactions with the surrounding tissue and subsequent systematic analysis of their failure modes. This review highlights the need for a change in the approach to investigating electrode failure, and in doing so summarizes the common ways in which neural electrodes fail, methods for identifying them and approaches to preventing them.
Collapse
|
4
|
Massey TL, Gleick JR, Haque RUM. Automated Multiplexed Potentiostat System (AMPS) for High-Throughput Characterization of Neural Interfaces. IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE : HEALTHCARE TECHNOLOGY : [PROCEEDINGS]. IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE 2021; 2021:10.1109/biocas49922.2021.9644948. [PMID: 35211701 PMCID: PMC8862781 DOI: 10.1109/biocas49922.2021.9644948] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Abstract
Neural interfaces with increasing channel counts require a scalable means of testing. While multiplexed potentiostats have long been the solution to this problem, most have been dedicated to one specific probe design or potentiostat, limited in the electrochemical techniques available, inordinately expensive, or they support multiplexing of too few channels. We present the design of an automated multiplexed potentiostat system that addresses these limitations-it is easily generalizable to any probe and potentiostat, supports any electrochemical technique available with the potentiostat, is low-cost, and can readily be expanded to hundreds of channels with support for multiple simultaneous potentiostats. This paper discusses the design philosophy and architecture of our 512-channel, 4-potentiostat system before demonstrating functionality with electrochemical impedance spectroscopy data, cyclic voltammetry curves, and an example of electrochemical surface modification, all on functional implantable microelectrode arrays currently being used for in vivo electrophysiological studies. Finally, we discuss the limitations to some sensitive or high-frequency impedance measurements due to reactive parasitics.
Collapse
Affiliation(s)
- Travis L. Massey
- Materials Engineering Division, Lawrence Livermore National Laboratory, Livermore, CA 94550, USA
| | - Jeremy R. Gleick
- Materials Engineering Division, Lawrence Livermore National Laboratory, Livermore, CA 94550, USA
| | - Razi-ul M. Haque
- Materials Engineering Division, Lawrence Livermore National Laboratory, Livermore, CA 94550, USA
| |
Collapse
|
5
|
Lamont C, Grego T, Nanbakhsh K, Shah Idil A, Giagka V, Vanhoestenberghe A, Cogan S, Donaldson N. Silicone encapsulation of thin-film SiO x, SiO xN yand SiC for modern electronic medical implants: a comparative long-term ageing study. J Neural Eng 2021; 18:055003. [PMID: 33752188 PMCID: PMC8208634 DOI: 10.1088/1741-2552/abf0d6] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/30/2020] [Accepted: 03/22/2021] [Indexed: 02/03/2023]
Abstract
Objective.Ensuring the longevity of implantable devices is critical for their clinical usefulness. This is commonly achieved by hermetically sealing the sensitive electronics in a water impermeable housing, however, this method limits miniaturisation. Alternatively, silicone encapsulation has demonstrated long-term protection of implanted thick-film electronic devices. However, much of the current conformal packaging research is focused on more rigid coatings, such as parylene, liquid crystal polymers and novel inorganic layers. Here, we consider the potential of silicone to protect implants using thin-film technology with features 33 times smaller than thick-film counterparts.Approach.Aluminium interdigitated comb structures under plasma-enhanced chemical vapour deposited passivation (SiOx, SiOxNy, SiOxNy+ SiC) were encapsulated in medical grade silicones, with a total of six passivation/silicone combinations. Samples were aged in phosphate-buffered saline at 67 ∘C for up to 694 days under a continuous ±5 V biphasic waveform. Periodic electrochemical impedance spectroscopy measurements monitored for leakage currents and degradation of the metal traces. Fourier-transform infrared spectroscopy, x-ray photoelectron spectroscopy, focused-ion-beam and scanning-electron- microscopy were employed to determine any encapsulation material changes.Main results.No silicone delamination, passivation dissolution, or metal corrosion was observed during ageing. Impedances greater than 100 GΩ were maintained between the aluminium tracks for silicone encapsulation over SiOxNyand SiC passivations. For these samples the only observed failure mode was open-circuit wire bonds. In contrast, progressive hydration of the SiOxcaused its resistance to decrease by an order of magnitude.Significance.These results demonstrate silicone encapsulation offers excellent protection to thin-film conducting tracks when combined with appropriate inorganic thin films. This conclusion corresponds to previous reliability studies of silicone encapsulation in aqueous environments, but with a larger sample size. Therefore, we believe silicone encapsulation to be a realistic means of providing long-term protection for the circuits of implanted electronic medical devices.
Collapse
Affiliation(s)
- C Lamont
- Implanted Devices Group, Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, United Kingdom,Author to whom any correspondence should be addressed
| | - T Grego
- Implanted Devices Group, Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, United Kingdom
| | - K Nanbakhsh
- Department of Microelectronics, Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands
| | - A Shah Idil
- Implanted Devices Group, Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, United Kingdom
| | - V Giagka
- Department of Microelectronics, Faculty of Electrical Engineering, Mathematics and Computer Science, Delft University of Technology, Delft, The Netherlands,
Department of System Integration and Interconnection Technologies, Fraunhofer Institute for Reliability and Micro-integration IZM, Berlin, Germany
| | - A Vanhoestenberghe
- Implanted Devices Group, Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, United Kingdom,
Institute of Orthopaedics and Musculoskeletal Science, RNOH Trust, University College London, Stanmore, United Kingdom
| | - S Cogan
- Department of Bioengineering, University of Texas at Dallas, Richardson, TX, United States of America
| | - N Donaldson
- Implanted Devices Group, Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London, United Kingdom
| |
Collapse
|
6
|
Akgun OC, Nanbakhsh K, Giagka V, Serdijn WA. A Chip Integrity Monitor for Evaluating Moisture/Ion Ingress in mm-Sized Single-Chip Implants. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS 2020; 14:658-670. [PMID: 32746351 DOI: 10.1109/tbcas.2020.3007484] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/11/2023]
Abstract
For mm-sized implants incorporating silicon integrated circuits, ensuring lifetime operation of the chip within the corrosive environment of the body still remains a critical challenge. For the chip's packaging, various polymeric and thin ceramic coatings have been reported, demonstrating high biocompatibility and barrier properties. Yet, for the evaluation of the packaging and lifetime prediction, the conventional helium leak test method can no longer be applied due to the mm-size of such implants. Alternatively, accelerated soak studies are typically used instead. For such studies, early detection of moisture/ion ingress using an in-situ platform may result in a better prediction of lifetime functionality. In this work, we have developed such a platform on a CMOS chip. Ingress of moisture/ions would result in changes in the resistance of the interlayer dielectrics (ILD) used within the chip and can be tracked using the proposed system, which consists of a sensing array and an on-chip measurement engine. The measurement system uses a novel charge/discharge based time-mode resistance sensor that can be implemented using simple yet highly robust circuitry. The sensor array is implemented together with the measurement engine in a standard 0.18 μm 6-metal CMOS process. The platform was validated through a series of dry and wet measurements. The system can measure the ILD resistance with values of up to 0.504 peta-ohms, with controllable measurement steps that can be as low as 0.8 M Ω. The system works with a supply voltage of 1.8 V, and consumes 4.78 mA. Wet measurements in saline demonstrated the sensitivity of the platform in detecting moisture/ion ingress. Such a platform could be used both in accelerated soak studies and during the implant's life-time for monitoring the integrity of the chip's packaging.
Collapse
|
7
|
Nanbakhsh K, Ritasalo R, Serdijn WA, Giagka V. Towards CMOS Bulk Sensing for In-Situ Evaluation of ALD Coatings for Millimeter Sized Implants. ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL INTERNATIONAL CONFERENCE 2020; 2020:3436-3439. [PMID: 33018742 DOI: 10.1109/embc44109.2020.9176103] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/11/2023]
Abstract
To meet the dimensional requirements for bioelectronic medicine, new packaging solutions are needed that could enable small, light-weight and flexible implants. For protecting the implantable electronics against biofluids, recently various atomic layer deposited (ALD) coatings have been proposed with high barrier properties. Before implantation, however, the protective coating should be evaluated for any defects which could otherwise lead to leakage and device failure. In these cases, the conventional helium leak test method can no longer be used due to the millimeter size of the implant. Therefore, an in-situ sensing platform is needed that could evaluate the coating and justify the implantation of the final device. In this work, we explore the possibility of using the CMOS bulk for such a platform. Towards this aim, as a proof of concept, test chips were made in a standard 6-metal 0.18 µm CMOS process and for the connection to the bulk, a p+ diffusion was used. A group of samples was then coated with an ALD multilayer. For coating evaluation, off-chip DC current leakage and impedance measurements were carried out in saline between the CMOS bulk and a platinum reference electrode. Results were compared between non-coated and coated chips that clearly demonstrated the potential of using the bulk as a sensing platform for coating evaluations. This novel approach could pave the way towards an all integrated in-situ hermeticity test, currently missing in mm-size implants.
Collapse
|
8
|
Nanbakhsh K, Kluba M, Pahl B, Bourgeois F, Dekker R, Serdijn W, Giagka V. Effect of Signals on the Encapsulation Performance of Parylene Coated Platinum Tracks for Active Medical Implants. ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL INTERNATIONAL CONFERENCE 2020; 2019:3840-3844. [PMID: 31946711 DOI: 10.1109/embc.2019.8857702] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/10/2022]
Abstract
Platinum is widely used as the electrode material for implantable devices. Owing to its high biostability and corrosion resistivity, platinum could also be used as the main metallization for tracks in active implants. Towards this goal, in this work we investigate the stability of parylene-coated Pt tracks using passive and active tests. The test samples in this study are Pt-on-SiO2 interdigitated comb structures. During testing all samples were immersed in saline for 150 days; for passive testing, the samples were left unbiased, whilst for active testing, samples were exposed to two different stress signals: a 5 V DC and a 5 Vp 500 pulses per second biphasic signal. All samples were monitored over time using impedance spectroscopy combined with optical inspection. After the first two weeks of immersion, delamination spots were observed on the Pt tracks for both passive and actively tested samples. Despite the delamination spots, the unbiased samples maintained high impedances until the end of the study. For the actively stressed samples, two different failure mechanisms were observed which were signal related. DC stressed samples showed severe parylene cracking mainly due to the electrolysis of the condensed water. Biphasically stressed samples showed gradual Pt dissolution and migration. These results contribute to a better understanding of the failure mechanisms of Pt tracks in active implants and suggest that new testing paradigms may be necessary to fully assess the long-term reliability of these devices.
Collapse
|