Gülcher OJ, Vis A, Peirlinck M, Kluin J. Balancing the ventricular outputs of pulsatile total artificial hearts. Artif Organs 2023; 47:1809-1817. [PMID: 37702086 DOI: 10.1111/aor.14641] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/19/2023] [Revised: 08/15/2023] [Accepted: 08/28/2023] [Indexed: 09/14/2023]
Abstract
BACKGROUND Maintaining balanced left and right cardiac outputs in a total artificial heart (TAH) is challenging due to the need for continuous adaptation to changing hemodynamic conditions. Proper balance in ventricular outputs of the left and right ventricles requires a preload-sensitive response and mechanisms to address the higher volumetric efficiency of the right ventricle. METHODS This review provides a comprehensive overview of various methods used to balance left and right ventricular outputs in pulsatile total artificial hearts, categorized based on their actuation mechanism. RESULTS Reported strategies include incorporating compliant materials and/or air cushions inside the ventricles, employing active control mechanisms to regulate ventricular filling state, and utilizing various shunts (such as hydraulic or intra-atrial shunts). Furthermore, reducing right ventricular stroke volume compared to the left often serves to balance the ventricular outputs. Individually controlled actuation of both ventricles in a pulsatile TAH seems to be the simplest and most effective way to achieve proper preload sensitivity and left-right output balance. Pneumatically actuated TAHs have the advantage to respond passively to preload changes. CONCLUSION Therefore, a pneumatic TAH that comprises two individually actuated ventricles appears to be a more desirable option-both in terms of simplicity and efficacy-to respond to changing hemodynamic conditions.
Collapse
Affiliation(s)
- Oskar J Gülcher
- Department of Cardiothoracic Surgery, Amsterdam University Medical Centers, Location University of Amsterdam, Amsterdam, The Netherlands
- Department of Biomechanical Engineering, Delft University of Technology, Delft, The Netherlands
| | - Annemijn Vis
- Department of Cardiothoracic Surgery, Amsterdam University Medical Centers, Location University of Amsterdam, Amsterdam, The Netherlands
| | - Mathias Peirlinck
- Department of Biomechanical Engineering, Delft University of Technology, Delft, The Netherlands
| | - Jolanda Kluin
- Department of Cardiothoracic Surgery, Amsterdam University Medical Centers, Location University of Amsterdam, Amsterdam, The Netherlands
- Department of Cardiothoracic Surgery, Thorax Center, Erasmus MC, Rotterdam, The Netherlands
| |
Collapse
|
2
|
Vis A, Arfaee M, Khambati H, Slaughter MS, Gummert JF, Overvelde JTB, Kluin J. The ongoing quest for the first total artificial heart as destination therapy. Nat Rev Cardiol 2022; 19:813-828. [PMID: 35668176 DOI: 10.1038/s41569-022-00723-8] [Citation(s) in RCA: 9] [Impact Index Per Article: 4.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Accepted: 05/09/2022] [Indexed: 12/18/2022]
Abstract
Many patients with end-stage heart disease die because of the scarcity of donor hearts. A total artificial heart (TAH), an implantable machine that replaces the heart, has so far been successfully used in over 1,700 patients as a temporary life-saving technology for bridging to heart transplantation. However, after more than six decades of research on TAHs, a TAH that is suitable for destination therapy is not yet available. High complication rates, bulky devices, poor durability, poor biocompatibility and low patient quality of life are some of the major drawbacks of current TAH devices that must be addressed before TAHs can be used as a destination therapy. Quickly emerging innovations in battery technology, wireless energy transmission, biocompatible materials and soft robotics are providing a promising opportunity for TAH development and might help to solve the drawbacks of current TAHs. In this Review, we describe the milestones in the history of TAH research and reflect on lessons learned during TAH development. We summarize the differences in the working mechanisms of these devices, discuss the next generation of TAHs and highlight emerging technologies that will promote TAH development in the coming decade. Finally, we present current challenges and future perspectives for the field.
Collapse
Affiliation(s)
- Annemijn Vis
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Maziar Arfaee
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Husain Khambati
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands.,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands
| | - Mark S Slaughter
- Department of Cardiovascular and Thoracic Surgery, University of Louisville, Louisville, KY, USA
| | - Jan F Gummert
- Department of Thoracic and Cardiovascular Surgery, Heart and Diabetes Center NRW, Ruhr-University Bochum, Bad Oeynhausen, Germany
| | - Johannes T B Overvelde
- Autonomous Matter Department, AMOLF, Amsterdam, The Netherlands.,Institute for Complex Molecular Systems and Department of Mechanical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands
| | - Jolanda Kluin
- Cardiothoracic Surgery, Amsterdam UMC location University of Amsterdam, Amsterdam, The Netherlands. .,Heart Failure and Arrhythmias, Amsterdam Cardiovascular Sciences, Amsterdam, The Netherlands.
| |
Collapse
|
3
|
Pelletier BA, Blaszczyk YM, Carstens P, Alvarez G, Lamping F, Laumen M, Finocchiaro T, Steinseifer U. Novel Optical Position Sensing for Miniaturized Applications and Validation in a Total Artificial Heart. IEEE Trans Biomed Eng 2016; 63:478-84. [DOI: 10.1109/tbme.2015.2463729] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/10/2022]
|
4
|
Casas F, Weber S, Klatte R, Goel V, Smith WA. Cycle testing of the MagScrew total artificial heart external battery pack: update I. Artif Organs 2011; 35:188-91. [PMID: 21323686 DOI: 10.1111/j.1525-1594.2010.01057.x] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Abstract
MagScrew total artificial heart (TAH) external battery pack (EBP) cycle bench testing continued over a period of 18 months using two fresh Wilson Greatbatch lithium ion EBPs during continuous charge and discharge cycles under a simulated TAH system current requirement. The same electronic load developed for our initial testing was used to simulate the MagScrew current waveforms typically observed during nominal operation. The current load profiles for this test were modified from the ones previously described and applied to the EBP under test during a voltage-defined discharge cycle. The test ended when EBP#2 reached end of life at 1450 cycles. At that point, EBP#1 remained healthy with a capacity of 175 min until full discharge. Performance of EBP#2 was still within expected ranges. Performance of EBP#1 exceeded expectations. These differences are probably caused by slight manufacturing changes. More tests will provide additional data to define a statistical distribution to better characterize EBP performance. In conclusion, endurance performance of the EBP remained satisfactory.
Collapse
Affiliation(s)
- Fernando Casas
- Lerner Research Institute, Department of Biomedical Engineering, Cleveland Clinic, Cleveland, OH 44195, USA.
| | | | | | | | | |
Collapse
|
5
|
Fukamachi K, Horvath DJ, Massiello AL, Fumoto H, Horai T, Rao S, Golding LAR. An innovative, sensorless, pulsatile, continuous-flow total artificial heart: device design and initial in vitro study. J Heart Lung Transplant 2009; 29:13-20. [PMID: 19782599 DOI: 10.1016/j.healun.2009.05.034] [Citation(s) in RCA: 69] [Impact Index Per Article: 4.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/29/2009] [Revised: 05/27/2009] [Accepted: 05/27/2009] [Indexed: 11/17/2022] Open
Abstract
BACKGROUND We are developing a very small, innovative, continuous-flow total artificial heart (CFTAH) that passively self-balances left and right pump flows and atrial pressures without sensors. This report details the CFTAH design concept and our initial in vitro data. METHODS System performance of the CFTAH was evaluated using a mock circulatory loop to determine the range of systemic and pulmonary vascular resistance (SVR and PVR) levels over which the design goal of a maximum absolute atrial pressure difference of 10 mm Hg is achieved for a steady-state flow condition. Pump speed was then modulated at 2,600 +/- 900 rpm to induce flow and arterial pressure pulsation to evaluate the effects of speed pulsations on the system performance. An automatic control mode was also evaluated. RESULTS Using only passive self-regulation, pump flows were balanced and absolute atrial pressure differences were maintained at <10 mm Hg over a range of SVR (750 to 2,750 dyne.sec.cm(-5)) and PVR (135 to 600 dyne.sec.cm(-5)) values far exceeding normal levels. The magnitude of induced speed pulsatility affected relative left/right performance, allowing for an additional active control to improve balanced flow and pressure. The automatic control mode adjusted pump speed to achieve targeted pump flows based on sensorless calculations of SVR and CFTAH flow. CONCLUSIONS The initial in vitro testing of the CFTAH with a single, valveless, continuous-flow pump demonstrated its passive self-regulation of flows and atrial pressures and a new automatic control mode.
Collapse
Affiliation(s)
- Kiyotaka Fukamachi
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA
| | | | | | | | | | | | | |
Collapse
|
6
|
Ootaki C, Yamashita M, Ootaki Y, Kamohara K, Weber S, Klatte RS, Smith WA, Massiello AL, Emancipator SN, Golding LAR, Fukamachi K. Reduced pulsatility induces periarteritis in kidney: role of the local renin-angiotensin system. J Thorac Cardiovasc Surg 2008; 136:150-8. [PMID: 18603068 DOI: 10.1016/j.jtcvs.2007.12.023] [Citation(s) in RCA: 68] [Impact Index Per Article: 4.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/09/2007] [Revised: 11/11/2007] [Accepted: 12/06/2007] [Indexed: 11/28/2022]
Abstract
OBJECTIVE The need for pulsatility in the circulation during long-term mechanical support has been a subject of debate. We compared histologic changes in calf renal arteries subjected to various degrees of pulsatile circulation in vivo. We addressed the hypothesis that the local renin-angiotensin system may be implicated in these histologic changes. METHODS AND RESULTS Sixteen calves were implanted with devices giving differing degrees of pulsatile circulation: 6 had a continuous flow left ventricular assist device (LVAD); 6 had a continuous flow right ventricular assist device (RVAD); and 4 had a pulsatile total artificial heart (TAH). Six other calves were histologic and immunohistochemical controls. In the LVAD group, the pulsatility index was significantly lower (0.28 +/- 0.07 LVAD vs 0.56 +/- 0.08 RVAD, vs 0.53 +/- 0.10 TAH; P < 0.01), and we observed severe periarteritis in all cases in the LVAD group. The number of angiotensin II type 1 receptor-positive cells and angiotensin converting enzyme-positive cells in periarterial areas was significantly higher in the LVAD group (angiotensin II type 1 receptor: 350 +/- 139 LVAD vs 8 +/- 6 RVAD, vs 3 +/- 2 TAH, vs 3 +/- 2 control; P < .001; angiotensin-converting enzyme: 325 +/- 59 LVAD vs 6 +/- 4 RVAD, vs 6 +/- 5 TAH, vs 3 +/- 1 control; P < .001). CONCLUSIONS The reduced pulsatility produced by a continuous flow LVAD implantation induced severe periarteritis in the kidneys. The local renin-angiotensin system was up-regulated in the inflammatory cells only in the continuous flow LVAD group.
Collapse
Affiliation(s)
- Chiyo Ootaki
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio 44195, USA
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
7
|
Abstract
Incompetent inflow valves have been reported with clinical pulsatile left ventricular assist devices that use bioprosthetic valves. Suspected as the cause of premature valve failure within these devices, absolute pressures and instantaneous pressure changes were evaluated in the MagScrew total artificial heart (TAH). The MagScrew TAH is a passively filling pulsatile pump which uses a reciprocating magnetic actuating mechanism under various control modes to propel blood into circulation. Both right and left ejection speeds were modulated and optimized at the onset of hydraulic eject. These various speed profiles were evaluated in vitro at 220 beats per minute (bpm), 100% pump fill, mean aortic pressure of 100 mm Hg and mean pulmonary artery pressure of 20 mm Hg. The pressure inside the left and right pump chambers was measured with Millar Mikro-Tip catheter and captured using Power Lab at a rate of 40 kHz. The pump chamber peak pressure, operating with unmodified eject speeds, measured on average 183 mm Hg for the left and 133 mm Hg for the right. Eject speed profiling for both pumps reduced the peak pressure by 10% and 28% for the left and right pump, respectively. Future studies will assess software controlled optimization of the eject speed profiles under any operating condition and how effective it is in vivo.
Collapse
|
8
|
Kamohara K, Weber S, Klatte RS, Luangphakdy V, Flick CR, Ootaki Y, Akiyama M, Cingoz F, Ootaki C, Kopcak MW, Liu J, Chen JF, Navia JL, Smith WA, Fukamachi K. Hemodynamic and metabolic changes during exercise in calves with total artificial hearts of different sizes yet similar output. Artif Organs 2007; 31:667-76. [PMID: 17725694 DOI: 10.1111/j.1525-1594.2007.00445.x] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Abstract
To evaluate the effects of downsizing of the total artificial heart (TAH), we compared the anaerobic threshold (AT) values in calves with two different types of TAH (Cleveland Clinic-Nimbus TAH and the downsized MagScrew TAH). Exercise studies were performed using a treadmill in 12 calves. During the exercise, parameters to obtain the AT were measured. To evaluate the determinants of the AT, a linear regression analysis was performed between AT and potential variables. AT values from 29 studies revealed no significant differences between the two different TAHs, with no significant differences in hemodynamic or oxygen metabolic parameters. AT values correlated well with pump flow/body weight (Q) multiplied by the hemoglobin level, regardless of the TAH used. In conclusion, downsizing of the original TAH design did not reduce AT without any significant differences in hemodynamic or oxygen metabolic parameters during exercise in calves.
Collapse
Affiliation(s)
- Keiji Kamohara
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, OH, USA
| | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
9
|
Casas F, Weber S, Klatte R, Luangphakdy V, Smith WA. Cycle testing of the MagScrew total artificial heart external battery pack. Artif Organs 2007; 31:698-702. [PMID: 17725697 DOI: 10.1111/j.1525-1594.2007.00444.x] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
Abstract
MagScrew total artificial heart (TAH) external battery pack (EBP) cycle bench testing was conducted on two Wilson Greatbatch (Clarence, NY, USA) lithium ion EBPs over a period of 22 months during continuous charge and discharge cycles under a simulated TAH system current requirement. A custom electronic load was developed to simulate the MagScrew current waveforms typically observed during nominal operation. These current load profiles were applied to the EBP under test during a voltage-defined discharge cycle. EBP endurance indicated a 240-min discharge cycle on a new battery diminishing linearly to 175 min after 800 cycles. A second linear trend started at this knee with 150 min of discharge time at 850 cycles until 10 min at 1600 cycles. Even at 1300 cycles, the EBP could still provide enough power for 60 min of nominal operation. In conclusion, the endurance performance of this EBP was satisfactory while exhibiting a predictable wear-out trend.
Collapse
Affiliation(s)
- Fernando Casas
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic Foundation, Cleveland, OH 44195, USA.
| | | | | | | | | |
Collapse
|
10
|
Kamohara K, Weber S, Klatte RS, Ootaki Y, Akiyama M, Kopcak MW, Luangphakdy V, Flick CR, Chen JF, Navia JL, Smith WA, Fukamachi K. Replacement of the left-side valves of an implanted total artificial heart. ASAIO J 2006; 52:368-72. [PMID: 16883114 DOI: 10.1097/01.mat.0000227731.46835.1e] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022] Open
Abstract
The MagScrew total artificial heart (TAH) is under development. Despite its anticipated durability and reliability, the possibility of a bioprosthetic valve malfunction exists. As a result, the potential for valve replacement surgery, instead of device replacement, would be desirable after a TAH implant. In two of our 90-day animal experiments, we successfully replaced the left-side valves through a left thoracotomy opposite to the right-sided incision site for the initial TAH implant. The results of these cases suggest that the left-side valves could also be replaced through a left thoracotomy approach in humans. To confirm the ability to access the left-side valves in humans, four human cadaver studies were performed with the use of a mock pump designed for human application. This report describes the operative techniques for left-side valve replacement in animals and discusses the advantages of a left thoracotomy in clinical situations, based on results from the human cadaver studies.
Collapse
Affiliation(s)
- Keiji Kamohara
- Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio, USA
| | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|