Stulak JM, Cowger J, Haft JW, Romano MA, Aaronson KD, Pagani FD. Device exchange after primary left ventricular assist device implantation: indications and outcomes.
Ann Thorac Surg 2012;
95:1262-7; discussion 1267-8. [PMID:
23063192 DOI:
10.1016/j.athoracsur.2012.08.031]
[Citation(s) in RCA: 59] [Impact Index Per Article: 4.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/01/2012] [Revised: 08/03/2012] [Accepted: 08/06/2012] [Indexed: 10/27/2022]
Abstract
BACKGROUND
Patients are being supported for longer periods with implantable left ventricular assist devices (LVADs) owing to longer transplantation wait times and approval of LVADs for destination therapy. This comes with an increased potential need for device exchange when complications arise. There are few data examining this patient population.
METHODS
Between August 1998 and January 2012, 45 patients (34 men) underwent 57 device exchanges after primary pulsatile or continuous-flow LVAD implantation. The median age at the initial LVAD implantation was 58 years (range, 28-78 years) and the median time to first device exchange was 15 months (range, immediate-56 months). Indications for primary LVAD included bridge to transplantation in all but 10 patients, and devices included the HeartMate I (Thoratec, Pleasanton, CA) in 16 patients, the HeartMate II (Thoratec) in 21 patients, the HeartWare HVAD (HeartWare, Framingham, MA) in 2 patients, the DuraHeart I (Terumo Heart, Ann Arbor, MI) in 1 patient, and other devices in 5 patients. Indications for reoperation included device/component failure (n=24), major driveline infection (n=15), pump thrombus (n=15), and other indications (n=2).
RESULTS
Pumps implanted in 57 reoperations included the HeartMate I in 15 patients, the HeartMate II in 35 patients, the HeartWare HVAD in 2 patients, the DuraHeart I in 2 patients, and other devices in 3 patients. Early mortality occurred in 2/57 (3.5%) patients. Median follow-up was 18 months (range, 1-113 months); median length of LVAD therapy after the first device exchange was 13 months (range, 1-59 months). Actuarial 1-year survival and freedom from repeated device exchange after the first exchange was 89% and 79%, respectively.
CONCLUSIONS
Device exchange may be required after LVAD implantation. This can be performed with low early mortality and no adverse effect on late survival. Multiple reoperations may be required in some patients.