Padeletti L, Pieragnoli P, Di Biase L, Colella A, Landolina M, Moro E, Orazi S, Vicentini A, Maglia G, Pensabene O, Raciti G, Barold SS. Is a Dual-Sensor Pacemaker Appropriate in Patients with Sino-Atrial Disease? Results from the DUSISLOG Study.
PACING AND CLINICAL ELECTROPHYSIOLOGY: PACE 2006;
29:34-40. [PMID:
16441715 DOI:
10.1111/j.1540-8159.2006.00301.x]
[Citation(s) in RCA: 30] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 12/01/2022]
Abstract
BACKGROUND
Rate-responsive pacemakers (PMs) are often supplied with accelerometer (XL) and minute ventilation (MV) sensors to provide a physiologic rate response according to patient needs. No information is available about the real benefit of dual-sensor rate-responsive pacing on the daily life of patients.
METHODS
DUSISLOG (Dual Sensor vs Single Sensor comparison using patient activity LOGbook) is a two-arm prospective, randomized, multicenter study that enrolled 105 patients who received a rate-responsive PM (Insignia), Guidant Corp.). After 1 month of DDD pacing at 60 ppm lower rate, a single sensor (XL or MV, randomized) was activated for 3 months at the manufacturer's suggested nominal settings, followed by a 3-month period with dual sensors optimized with automatic response. During the last month of each period, the following data concerning patient physical activity were retrieved from PM diagnostics (Activity Log): mean percentage of physical activity, mean intensity of activity. Quality of life (QoL) scores and 6-minute walk test (WT) were also recorded.
RESULTS
Single-sensor rate-responsive pacing resulted in symptomatic benefit equally with XL and MV sensors while no additional benefit was found using dual sensor. In a subgroup analysis, patients (17%) with marked chronotropic incompetence and with 0% atrial sensing received benefits from single sensor with an additional advantage from sensor (QoL: +21 +/- 14% P < 0.05; WT: +17 +/- 7% P < 0.02).
CONCLUSION
In most patients with rate-responsive devices, a single sensor is sufficient to achieve a satisfactory rate response. A dual sensor combination and optimization provides an additional benefit only in a selected population with an advanced atrial chronotropic disease.