Knies M, Teske E, Kooistra H. Comparison of Doppler ultrasonic sphygmomanometry, oscillometry and high-definition oscillometry for non-invasive blood pressure measurement in conscious cats. J Feline Med Surg 2024; 26:1098612X241231471. [PMID: 38546192 PMCID: PMC10983607 DOI: 10.1177/1098612x241231471] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 04/19/2024]
Abstract
OBJECTIVES Systemic arterial hypertension is a common occurrence and can have serious adverse consequences in cats. Therefore, measuring blood pressure is very important. There are many indirect blood pressure measurement devices available. This study compared Doppler, oscillometric (petMAP Graphic II, SunTech Vet20, the Cardell Insight-X0000) and high-definition oscillometry devices for the non-invasive measurement of blood pressure in conscious cats. METHODS In this prospective study, blood pressure was measured in 32 cats using the different devices according to the recommendations of the American College of Veterinary Internal Medicine Consensus Statement. Blood pressures (systolic blood pressure [SBP], diastolic blood pressure [DBP], mean arterial pressure [MAP]), time to complete measurements, number of attempts needed, coefficient of variation (CV) between the blood pressure measurements of the different devices, ease of completing measurements and apparent stress level of the cat were assessed. RESULTS There was a significant difference between devices in the time taken to obtain blood pressure readings and the number of attempts necessary to obtain six reliable measurements. The CV of the Doppler device was significantly smaller than that of the rest of the devices, but there were no other differences between the devices. The mean SBP, DBP and MAP measured by the petMAP device were significantly higher than the measurements from the other devices. The perceived ease of measurement was not significantly different between the various machines. The perceived level of stress of measurement with the Doppler device was significantly higher compared with the other devices but did not lead to an increased SBP. CONCLUSIONS AND RELEVANCE Using a Doppler device to measure blood pressure in conscious cats is fast, relatively easy and gives reliable results. A disadvantage is that the Doppler device can only measure SBP, while oscillometric devices also provide DBP and MAP. However, in veterinary medicine, systolic hypertension is considered the most relevant.
Collapse
Affiliation(s)
- Marieke Knies
- AniCura Veterinary Referral Centre Haaglanden, Dordrecht, The Netherlands
- Department of Clinical Sciences, Faculty of Veterinary Medicine, Utrecht University, Utrecht, The Netherlands
| | - Erik Teske
- Department of Clinical Sciences, Faculty of Veterinary Medicine, Utrecht University, Utrecht, The Netherlands
| | - Hans Kooistra
- Department of Clinical Sciences, Faculty of Veterinary Medicine, Utrecht University, Utrecht, The Netherlands
| |
Collapse
|
2
|
Weidinger AK, Bergmann M, König M, Zablotski Y, Hartmann K. Anti-rabies humoral immune response in cats after concurrent vs separate vaccination against rabies and feline leukaemia virus using canarypox-vectored vaccines. J Feline Med Surg 2024; 26:1098612X231218643. [PMID: 38358295 PMCID: PMC10911302 DOI: 10.1177/1098612x231218643] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/16/2024]
Abstract
OBJECTIVES Some expert groups recommend that cats should be vaccinated with non-adjuvanted feline leukaemia virus (FeLV) and rabies vector vaccines, which, in the European Union, are currently not licensed for concurrent use and have to be administered at least 14 days apart (different from the USA) and thus at separate visits, which is associated with more stress for cats and owners. The aim of this study was to assess the anti-rabies antibody response in cats after vaccination against rabies and FeLV at concurrent vs separate (4 weeks apart) visits using two canarypox-vectored vaccines (Purevax Rabies and Purevax FeLV; Boehringer Ingelheim) and to evaluate the occurrence of vaccine-associated adverse events (VAAEs). METHODS Healthy FeLV antigen-negative client-owned kittens (n = 106) were prospectively included in this randomised study. All kittens received primary vaccinations against rabies (week 0) and FeLV (weeks 4 and 8). After 1 year, the study group (n = 52) received booster vaccinations against rabies and FeLV concurrently at the same visit (weeks 50-52). The control group (n = 54) received booster vaccinations against rabies (weeks 50-52) and FeLV (weeks 54-56) separately. Anti-rabies virus antibodies (anti-RAV Ab) were determined by fluorescent antibody virus neutralisation assay at weeks 4, 50-52 and 54-56, and compared between both groups using a Mann-Whitney U-test. RESULTS Four weeks after the first rabies vaccination, 87/106 (82.1%) kittens had a titre ⩾0.5 IU/ml and 19/106 (17.9%) had a titre <0.5 IU/ml. Four weeks after the 1-year rabies booster, all cats had adequate anti-RAV Ab according to the World Organisation for Animal Health (⩾0.5 IU/ml), and the titres of the study group (median = 14.30 IU/ml) and the control group (median = 21.39 IU/ml) did not differ significantly (P = 0.141). VAAEs were observed in 7/106 (6.6%) cats. CONCLUSIONS AND RELEVANCE Concurrent administration of Purevax FeLV and Purevax Rabies vector vaccines at the 1-year booster does not interfere with the development of anti-RAV Ab or cause more adverse effects and thus represents a better option than separate vaccination visits for cats and owners.
Collapse
Affiliation(s)
- Anna-Karina Weidinger
- LMU Small Animal Clinic, Centre for Clinical Veterinary Medicine, LMU Munich, Munich, Germany
| | - Michèle Bergmann
- LMU Small Animal Clinic, Centre for Clinical Veterinary Medicine, LMU Munich, Munich, Germany
| | - Matthias König
- Institute of Virology, Faculty of Veterinary Medicine, Justus Liebig University, Giessen, Hessen, Germany
| | - Yury Zablotski
- LMU Small Animal Clinic, Centre for Clinical Veterinary Medicine, LMU Munich, Munich, Germany
| | - Katrin Hartmann
- LMU Small Animal Clinic, Centre for Clinical Veterinary Medicine, LMU Munich, Munich, Germany
| |
Collapse
|
3
|
Casas C, Dye C. Comparison of Thames Medical CAT+ Doppler and SunTech Vet 20 oscillometric devices for non-invasive blood pressure measurement in conscious cats. J Feline Med Surg 2024; 26:1098612X231216350. [PMID: 38381465 PMCID: PMC10911307 DOI: 10.1177/1098612x231216350] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/22/2024]
Abstract
OBJECTIVES A comparative assessment of systolic blood pressure (BP) measurement agreement and precision in two commonly used non-invasive BP devices was carried out in conscious cats. METHODS Systolic BP measurements were obtained from 50 conscious cats as part of their clinical investigations. All measurements were taken by the same operator and were performed according to the American College of Veterinary Internal Medicine (ACVIM) consensus guidelines. The same cuff location and cuff size were used for paired measurements. The order of device use was randomised, and an arousal score was assigned during each procedure. Precision was assessed using standard deviation and coefficient of variance comparisons. Agreement was assessed using graphical and statistical comparisons of derived 'delta mean' and 'mean of means' data. RESULTS A total of 50 cats aged between 3 months and 15 years were enrolled (29 domestic shorthair, seven domestic longhair, three British Shorthair, two Ragdoll, two Sphynx, two Persian, one Siamese, one Burmese, one Russian Blue, one Maine Coon and one Oriental) with a body weight in the range of 1.2-6.2 kg. BP measurements were in the range of 95-179 mmHg. Oscillometric measurements were associated with lower arousal scores. The Doppler device demonstrated superior repeatability precision. Body weight, sex, cuff size, cuff location, arousal score and the order in which the devices were used had no significant effect on precision or agreement. Correlation between the two devices was 0.0837 and agreement was considered clinically acceptable (<10 mmHg) in 32/50 (64%) cats. Of the 18 cats with suboptimal agreement, the oscillometric mean BP was higher in 14 (78%) cats. CONCLUSIONS AND RELEVANCE Based on this study sample, suboptimal agreement between paired CAT+ Doppler and SunTech Vet20 oscillometric BP measurements in 36% of conscious cats suggests that these devices should not be used interchangeably. Compared with Doppler, oscillometric measurements were consistently higher, particularly at higher blood pressures. Further studies are required to assess which device has superior accuracy in conscious cats.
Collapse
Affiliation(s)
- Clara Casas
- Department of Small Animal Medicine, Pride Veterinary Centre, IVC Evidensia, Derby, DE24 8HX, UK
| | - Charlotte Dye
- Department of Small Animal Medicine, Pride Veterinary Centre, IVC Evidensia, Derby, DE24 8HX, UK
| |
Collapse
|
4
|
Uematsu S, Summers S, Keys D, Quimby J. Effect of audible static on blood pressure measurement by Doppler ultrasonic sphygmomanometry in cats. J Feline Med Surg 2023; 25:1098612X231197084. [PMID: 37751181 PMCID: PMC10812029 DOI: 10.1177/1098612x231197084] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 08/06/2023] [Indexed: 09/27/2023]
Abstract
OBJECTIVES The aim of the present study was to determine whether the use of headphones to eliminate audible static during Doppler ultrasonic sphygmomanometry affects blood pressure (BP) measurement in conscious young adult (aged 1-6 years) and mature adult/senior (aged ⩾7 years) cats. METHODS A randomized crossover study was conducted. Healthy client-owned cats (>1 year) were enrolled. Blood pressure measurements were obtained twice, 14 days apart, with or without the use of headphones worn by a veterinarian. A fear, anxiety and stress (FAS) score (0 = relaxed; 4 = severe signs) was recorded. A linear mixed-effects model was used to compare the effect of wearing headphones on BP measurement. RESULTS In total, 18 young adult and 14 mature adult and senior cats with a median age of 5 years (range 1-14 years) were enrolled. Of the cats, 47% (15/32) had an average BP measurement that was at least 10 mmHg higher when using headphones compared with when not using headphones, of which a majority (11/15, 73%) were young adult cats. The average BP measurement was not different when using headphones compared with when not using headphones (mean difference -7 mmHg; 95% confidence interval -14 to 0.6; P = 0.07). When compared within age groups, the average BP measurement taken when using headphones (125 ± 15 mmHg) was lower compared with the measurement taken when not using headphones (137 ± 17 mmHg) in young adult cats (P = 0.02). CONCLUSIONS AND RELEVANCE The reduction in the average BP measurement with the use of headphones suggests this method may be helpful in reducing situational hypertension, particularly in young adult cats. Our findings also highlight the importance of consistent use of headphones when comparing serial measurements in a cat.
Collapse
Affiliation(s)
- Sayaka Uematsu
- Department of Clinical Sciences, Oregon State University, Corvallis, OR, USA
| | - Stacie Summers
- Department of Clinical Sciences, Oregon State University, Corvallis, OR, USA
| | | | - Jessica Quimby
- Department of Veterinary Clinical Sciences, The Ohio State University, Columbus, OH, USA
| |
Collapse
|
5
|
Characterization of sinoatrial automaticity in Microcebus murinus to study the effect of aging on cardiac activity and the correlation with longevity. Sci Rep 2023; 13:3054. [PMID: 36810863 PMCID: PMC9944915 DOI: 10.1038/s41598-023-29723-5] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/26/2022] [Accepted: 02/09/2023] [Indexed: 02/24/2023] Open
Abstract
Microcebus murinus, or gray mouse lemur (GML), is one of the smallest primates known, with a size in between mice and rats. The small size, genetic proximity to humans and prolonged senescence, make this lemur an emerging model for neurodegenerative diseases. For the same reasons, it could help understand how aging affects cardiac activity. Here, we provide the first characterization of sinoatrial (SAN) pacemaker activity and of the effect of aging on GML heart rate (HR). According to GML size, its heartbeat and intrinsic pacemaker frequencies lie in between those of mice and rats. To sustain this fast automaticity the GML SAN expresses funny and Ca2+ currents (If, ICa,L and ICa,T) at densities similar to that of small rodents. SAN automaticity was also responsive to β-adrenergic and cholinergic pharmacological stimulation, showing a consequent shift in the localization of the origin of pacemaker activity. We found that aging causes decrease of basal HR and atrial remodeling in GML. We also estimated that, over 12 years of a lifetime, GML generates about 3 billion heartbeats, thus, as many as humans and three times more than rodents of equivalent size. In addition, we estimated that the high number of heartbeats per lifetime is a characteristic that distinguishes primates from rodents or other eutherian mammals, independently from body size. Thus, cardiac endurance could contribute to the exceptional longevity of GML and other primates, suggesting that GML's heart sustains a workload comparable to that of humans in a lifetime. In conclusion, despite the fast HR, GML replicates some of the cardiac deficiencies reported in old people, providing a suitable model to study heart rhythm impairment in aging. Moreover, we estimated that, along with humans and other primates, GML presents a remarkable cardiac longevity, enabling longer life span than other mammals of equivalent size.
Collapse
|
6
|
Navarro I, Summers S, Rishniw M, Quimby J. Cross-sectional survey of non-invasive indirect blood pressure measurement practices in cats by veterinarians. J Feline Med Surg 2022; 24:1195-1202. [PMID: 35109690 PMCID: PMC10812353 DOI: 10.1177/1098612x211067015] [Citation(s) in RCA: 4] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Accepted: 11/28/2021] [Indexed: 12/31/2022]
Abstract
OBJECTIVES This survey of small animal veterinarians endeavored to: (1) determine current methods of indirect blood pressure measurement; (2) detail techniques used to reduce situational hypertension; and (3) better understand the obstacles to performing blood pressure measurement in cats. METHODS An online survey was produced and circulated to members of the Veterinary Information Network. A total of 733 veterinarians who saw cats in their practice and had access to at least one indirect blood pressure device completed the entirety of the survey. RESULTS Ninety-six percent (703/733) of veterinarians who completed the survey reported measuring indirect blood pressure in cats in their practice, with veterinary technicians conducting most (600/703; 85.3%) of these measurements. Few veterinarians (30/733; 4.1%) did not measure blood pressure, with these veterinarians citing several obstacles including: difficulty interpreting results with the occurrence of fear, anxiety and stress in cats (20/30; 66.7%); difficulty performing measurements in cats (17/30; 56.7%); and technical staff being uncomfortable performing measurements (12/30; 40.0%). Most veterinarians (300/435; 69.0%) in this survey preferred an ultrasonic Doppler flow detector with sphygmomanometry, with many (272/300; 90.7%) perceiving that the results obtained with this device were more trustworthy compared with results obtained with oscillometry. Ninety percent (633/703) of veterinarians employed techniques to reduce situational hypertension in cats. Techniques perceived to be most helpful among veterinarians included: using a quiet location (454/633; 71.7%); minimizing restraint (316/633; 49.9%); performing blood pressure prior to other procedures (eg, phlebotomy, cystocentesis) (302/633; 47.7%); avoiding other animals (219/633; 34.6%); and allowing time for acclimation (167/633; 26.4%). CONCLUSIONS AND RELEVANCE This survey study of veterinarians helps clarify obstacles to routine blood pressure measurement in conscious cats. Veterinarians reported several strategies that they felt reduced situational hypertension in cats. The data inform modifications of techniques to increase the frequency and perceived reliability of blood pressure measurement in at-risk cats.
Collapse
|
7
|
Hanås S, Larsson A, Rydén J, Lilliehöök I, Häggström J, Tidholm A, Höglund K, Ljungvall I, Holst BS. Cardiac troponin I in healthy Norwegian Forest Cat, Birman and domestic shorthair cats, and in cats with hypertrophic cardiomyopathy. J Feline Med Surg 2022; 24:e370-e379. [PMID: 36073987 PMCID: PMC9511503 DOI: 10.1177/1098612x221117115] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/21/2022]
Abstract
Objectives The aims of this study were to assess the potential associations between the serum cardiac troponin I (cTnI) concentration in healthy cats and feline characteristics, systolic blood pressure, heart rate (HR), echocardiographic measurements and storage time; and to compare cTnI concentrations in healthy cats with concentrations in cats with hypertrophic cardiomyopathy (HCM), with or without left atrial enlargement (LAE) and in cats with HCM, to assess potential associations between cTnI concentration and echocardiographic variables. Methods Cardiac TnI was analysed using an Abbott ARCHITECT ci16200 analyser in serum from prospectively included healthy Norwegian Forest Cat (NF; n = 33), Birman (n = 33) and domestic shorthair (DSH; n = 30) cats, and from 39 cats with HCM, with or without LAE. Results In healthy cats, higher cTnI concentrations were found in Birman cats than in NF cats (P = 0.014) and in neutered male cats than in intact females (P = 0.032). Cardiac TnI was positively associated with HR (P <0.0001). In cats with HCM, cTnI concentration was positively associated with left ventricular wall thickness and with left atrial-to-aortic root ratio (all P ⩽0.010). Cats with HCM had higher cTnI concentrations than healthy cats, and cTnI concentrations were higher in cats with HCM and LAE than in those with HCM without LAE (all P = 0.0003). Conclusions and relevance Breed and sex may affect serum cTnI concentrations in healthy cats. The cTnI concentration increased with increasing severity of HCM.
Collapse
Affiliation(s)
- Sofia Hanås
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden.,Evidensia Specialist Animal Hospital Strömsholm, Strömsholm, Sweden
| | - Anders Larsson
- Department of Medical Sciences, Clinical Chemistry, Uppsala University, Uppsala, Sweden
| | - Jesper Rydén
- Department of Energy and Technology, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Inger Lilliehöök
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Jens Häggström
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Anna Tidholm
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden.,Anicura Albano Animal Hospital, Stockholm, Sweden
| | - Katja Höglund
- Department of Anatomy, Physiology and Biochemistry, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Ingrid Ljungvall
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Bodil S Holst
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| |
Collapse
|
8
|
Sparkes A, Garelli-Paar C, Blondel T, Guillot E. 'The Mercury Challenge': feline systolic blood pressure in primary care practice - a European survey. J Feline Med Surg 2022; 24:e310-e323. [PMID: 35757930 PMCID: PMC9511504 DOI: 10.1177/1098612x221105844] [Citation(s) in RCA: 4] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/15/2022]
Abstract
OBJECTIVES The aim of this study was to collect data from a substantial number of older cats having their systolic blood pressure (SBP) measured in a variety of clinical practices, to describe the findings and assess variables that affected the duration of assessment and the values obtained. METHODS An international (European-based) multicentre convenience sample survey of cats ⩾7 years of age attending veterinary clinics and having SBP measured as part of their clinical assessment. Information gathered included details of the cat, concomitant disease(s) or therapies, SBP results, device used, time taken to assess SBP and the demeanor of the cat. RESULTS Useable data were available from 8884 cats aged 7-26 years, from 811 clinics across 16 countries. The device used to measure SBP was Doppler in 47.4% and oscillometry in 48.5%. The demeanor of the cat was reported to be calm in 45.7%, anxious in 41.9% and nervous in 8.9%; and the duration of assessment was reported to be <5 mins in 50.4%, 5-10 minutes in 41.7% and >10 mins in 7.9%. Concomitant chronic kidney disease (CKD) was reported in 21.8%, hyperthyroidism in 12.0% or both in 3.1%. The median SBP was 150 mmHg (range 80-310), with 18.6% classified as hypertensive (SBP 160-179 mmHg) and 21.1% as severely hypertensive (SBP ⩾180 mmHg). The measured SBP was significantly affected by the cat's demeanor, duration of SBP assessment, presence of CKD and/or hyperthyroidism, the cat's sex and age, and the presence of concomitant therapy. The duration of SBP assessment was significantly affected by the cat's demeanor. CONCLUSIONS AND RELEVANCE In veterinary clinics, SBP can be measured in most cats within a short period of time using either Doppler or oscillometric equipment. The presence of CKD or hyperthyroidism was associated with significantly higher SBP values, and anxious or nervous cats had higher SBP values and took longer to obtain SBP assessments.
Collapse
Affiliation(s)
- Andrew Sparkes
- Simply Feline Veterinary Consultancy, Shaftesbury, Dorset, UK
| | | | - Thomas Blondel
- Innovation and Development, Ceva Santé Animale, Libourne, France
| | - Emilie Guillot
- Companion Animal Franchise, Ceva Santé Animale, Libourne, France
| |
Collapse
|
9
|
Hanås S, Holst BS, Ljungvall I, Tidholm A, Olsson U, Häggström J, Höglund K. Influence of clinical setting and cat characteristics on indirectly measured blood pressure and pulse rate in healthy Birman, Norwegian Forest, and Domestic Shorthair cats. J Vet Intern Med 2021; 35:801-811. [PMID: 33675246 PMCID: PMC7995443 DOI: 10.1111/jvim.16096] [Citation(s) in RCA: 9] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/14/2020] [Revised: 02/17/2021] [Accepted: 02/19/2021] [Indexed: 12/13/2022] Open
Abstract
BACKGROUND Measured indirect blood pressure (BP) results in cats in a clinical environment might be affected by stress and characteristics of the cats. HYPOTHESIS To investigate the influence of clinical setting, cat characteristics, and life situation on BP and pulse rate (PR) in healthy cats. ANIMALS Ninety-four healthy Domestic Shorthair, Birman and Norwegian Forest cats. METHODS Blood pressure measured by high-definition oscillometry in 3 settings: cat placed in its own carrier with veterinarian present; cat placed in carrier with owner alone present; and cat placed on table with veterinarian present. Statistical analyses were performed using mixed linear models. RESULTS Systolic BP (SBP) did not differ among settings. Higher mean arterial pressure (MAP), diastolic BP (DBP), and PR were found when measurements were performed with cat placed on table, rather than in carrier. Coefficients of variation (CVs) higher for SBP, MAP, DBP, and PR when measured with cat placed on table than in carrier. Birman cats had lower BP than other breeds. Systolic BP, MAP, DBP, and PR increased with age. Cats allowed outdoors had lower PR than cats living strictly indoors. CONCLUSION AND CLINICAL IMPORTANCE No difference in SBP was found among settings, but measuring BP with the cat placed on the examination table gave higher MAP, DBP, PR, and CV than measuring BP with the cat in its carrier. Breed affected BP, with lower BP in Birman cats than other breeds. Blood pressure increased with age. Pulse rate was lower in cats allowed outdoors than cats living strictly indoors.
Collapse
Affiliation(s)
- Sofia Hanås
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden.,Evidensia Specialist Animal Hospital Strömsholm, Strömsholm, Sweden
| | - Bodil S Holst
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Ingrid Ljungvall
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Anna Tidholm
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden.,Anicura Albano Animal Hospital, Stockholm, Sweden
| | - Ulf Olsson
- Unit of Applied Statistics and Mathematics, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Jens Häggström
- Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden
| | - Katja Höglund
- Department of Anatomy, Physiology and Biochemistry, Swedish University of Agricultural Sciences, Uppsala, Sweden
| |
Collapse
|