1
|
Shen L, Antonuk LE, El-Mohri Y, Liang AK, Zhao Q, Jiang H. Theoretical investigation of the signal performance of HgI 2x-ray converters incorporating a Frisch grid structure at mammographic energies. Phys Med Biol 2021; 66. [PMID: 34252890 DOI: 10.1088/1361-6560/ac1365] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/05/2021] [Accepted: 07/12/2021] [Indexed: 11/12/2022]
Abstract
Active matrix, flat-panel imagers (AMFPIs) suffer from decreased detective quantum efficiency under conditions of low dose per image frame (such as for digital breast tomosynthesis, fluoroscopy and cone-beam CT) due to low signal compared to the additive electronic noise. One way to address this challenge is to introduce a high-gain x-ray converter called particle-in-binder mercuric iodide (PIB HgI2) which exhibits 3-10 times higher x-ray sensitivity compared to that of a-Se and CsI:Tl converters employed in commercial AMFPI systems. However, a remaining challenge for practical implementation of PIB HgI2is the high level of image lag, which is believed to largely originate from the trapping of holes. Towards addressing this challenge, this paper reports a theoretical investigation of the use of a Frisch grid structure embedded in the converter to suppress hole signal-which would be expected to reduce image lag. The grid acts as a third electrode sandwiched between a continuous top electrode and pixelated bottom electrodes having a 100μm pitch. Signal properties of such a detector are investigated as a function of VDR (the ratio of the voltage difference between the electrodes in the region below the grid to that above the grid), grid pitch (the center-to-center distance between two neighboring grid wires) andRGRID(the ratio of grid wire width to grid pitch) for mammographic x-ray energies. The results show that smaller grid pitch suppresses hole signal to a higher degree (up to ∼96%) while a larger gap between grid wires and higher VDR provide minimally impeded electron transport. Examination of the tradeoff between maximizing electron signal and minimizing hole signal indicates that a grid design having a grid pitch of 20μm withRGRIDof 50% and 65% provides hole signal suppression of ∼93% and ∼95% for VDR of 1 and 3, respectively.
Collapse
Affiliation(s)
- Liuxing Shen
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| | - Larry E Antonuk
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| | - Youcef El-Mohri
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| | - Albert K Liang
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| | - Qihua Zhao
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| | - Hao Jiang
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI 48109, United States of America
| |
Collapse
|
2
|
Tu SJ, Chen WY, Wu CT. Uncertainty measurement of radiomics features against inherent quantum noise in computed tomography imaging. Eur Radiol 2021; 31:7865-7875. [PMID: 33852047 DOI: 10.1007/s00330-021-07943-5] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/17/2020] [Revised: 03/18/2021] [Accepted: 03/25/2021] [Indexed: 12/24/2022]
Abstract
OBJECTIVES Quantum noise is a random process in X-ray-based imaging systems. We addressed and measured the uncertainty of radiomics features against this quantum noise in computed tomography (CT) images. METHODS A clinical multi-detector CT scanner, two homogeneous phantom sets, and four heterogeneous samples were used. A solid tumor tissue removed from a male BALB/c mouse was included. We the placed phantom sets on the CT scanning table and repeated 20 acquisitions with identical imaging settings. Regions of interest were delineated for feature extraction. Statistical quantities-average, standard deviation, and percentage uncertainty-were calculated from these 20 repeated scans. Percentage uncertainty was used to measure and quantify feature stability against quantum noise. Twelve radiomics features were measured. Random noise was added to study the robustness of machine learning classifiers against feature uncertainty. RESULTS We found the ranges of percentage uncertainties from homogeneous soft tissue phantoms, homogeneous bone phantoms, and solid tumor tissue to be 0.01-2138%, 0.02-15%, and 0.18-16%, respectively. Overall, it was found that the CT features ShortRunHighGrayLevelEmpha (SRHGE) (0.01-0.18%), ShortRunLowGrayLevelEmpha (SRLGE) (0.01-0.41%), LowGrayLevelRunEmpha (LGRE) (0.01-0.39%), and LongRunLowGrayLevelEmpha (LRLGE) (0.02-0.66%) were the most stable features against the inherent quantum noise. The most unstable features were cluster shade (1-2138%) and max probability (1-16%). The impact of random noise to the prediction accuracy by different machine learning classifiers was found to be between 0 and 12%. CONCLUSIONS Twelve features were used for uncertainty measurements. The upper and lower bounds of percentage uncertainties were determined. The quantum noise effect on machine learning classifiers is model dependent. KEY POINTS • Quantum noise is a random process and is intrinsic to X-ray-based imaging systems. This inherent quantum noise creates unpredictable fluctuations in the gray-level intensities of image pixels. Extra cautions and further validations are strongly recommended when unstable radiomics features are selected by a predictive model for disease classification or treatment outcome prognosis. • We addressed and used the statistical quantity of percentage uncertainty to measure the uncertainty of radiomics features against the inherent quantum noise in computed tomography (CT) images. • A clinical multi-detector CT scanner, two homogeneous phantom sets, and four heterogeneous samples were used in the stability measurement. A solid tumor tissue removed from a male BALB/c mouse was included in the heterogeneous sample.
Collapse
Affiliation(s)
- Shu-Ju Tu
- Department of Medical Imaging and Radiological Sciences, College of Medicine, Chang Gung University, 259 Wen-Hua First Road, Kwei-Shan, Tao-Yuan, 333, Taiwan. .,Department of Medical Imaging and Intervention, Linkou Chang Gung Memorial Hospital, Tao-Yuan, Taiwan.
| | - Wei-Yuan Chen
- Department of Medical Imaging and Radiological Sciences, College of Medicine, Chang Gung University, 259 Wen-Hua First Road, Kwei-Shan, Tao-Yuan, 333, Taiwan.,Department of Medical Imaging and Intervention, Linkou Chang Gung Memorial Hospital, Tao-Yuan, Taiwan
| | - Chen-Te Wu
- Department of Medical Imaging and Radiological Sciences, College of Medicine, Chang Gung University, 259 Wen-Hua First Road, Kwei-Shan, Tao-Yuan, 333, Taiwan.,Department of Medical Imaging and Intervention, Linkou Chang Gung Memorial Hospital, Tao-Yuan, Taiwan
| |
Collapse
|
3
|
Altunbas C, Alexeev T, Miften M, Kavanagh B. Effect of grid geometry on the transmission properties of 2D grids for flat detectors in CBCT. Phys Med Biol 2019; 64:225006. [PMID: 31585444 DOI: 10.1088/1361-6560/ab4af4] [Citation(s) in RCA: 10] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
To suppress scatter in cone beam computed tomography (CBCT), two-dimensional antiscatter grids (2D grid) have been recently proposed. In this work, we developed several grid prototypes with higher grid ratios and smaller grid pitches than previous designs, and quantified their primary and scatter transmission properties in the context of CBCT for radiation therapy. Three focused 2D grid prototypes were developed with grid ratios at 12 and 16, and grid pitches at 2 and 3 mm. Their scatter transmission properties were measured between 80-140 kVp, and benchmarked against a high performance radiographic grid (1D grid) using a Varian TrueBeam CBCT system. The effect of source-grid misalignment on the primary transmission and the improvement in contrast-to-noise ratio (CNR) were also evaluated. Changing the grid pitch from two to three mm increased the average primary transmission from 84% to 89%. Maximum scatter-to-primary ratio (SPR) with grid ratio of 12 was 0.3, and increasing the grid ratio to 16 reduced SPR by 30%. A 10 mm misalignment in 2D grid position led to a 6%-8% reduction in average primary transmission, and reduction was more pronounced for the higher grid ratio. 2D grids provided up to factor of seven lower SPR and 21% better primary transmission than the 1D grid, and their scatter transmission exhibited lower energy dependence. While 2D grids provided up to factor of 2.3 higher CNR improvement, a significant variation in CNR improvement was not observed among different grid pitch and ratios. In summary, grid ratio of 16 and grid pitch of 2 mm can keep SPRs below 0.2 even in high scatter conditions, while keeping primary transmission fractions above 80%, key benefits of the investigated 2D grids in improving image quality of CBCT. However, such grids require precise alignment in source-grid geometry during CBCT acquisitions. This study also implies that 2D grids can provide substantially better scatter suppression and primary transmission than high-performance 1D grids currently available.
Collapse
Affiliation(s)
- Cem Altunbas
- Author to whom correspondence should be addressed
| | | | | | | |
Collapse
|
4
|
Wu P, Stayman JW, Sisniega A, Zbijewski W, Foos D, Wang X, Aygun N, Stevens R, Siewerdsen JH. Statistical weights for model-based reconstruction in cone-beam CT with electronic noise and dual-gain detector readout. ACTA ACUST UNITED AC 2018; 63:245018. [DOI: 10.1088/1361-6560/aaf0b4] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/11/2022]
|
5
|
Koniczek M, Antonuk LE, El-Mohri Y, Liang AK, Zhao Q. Theoretical investigation of the noise performance of active pixel imaging arrays based on polycrystalline silicon thin film transistors. Med Phys 2017; 44:3491-3503. [PMID: 28376261 DOI: 10.1002/mp.12257] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/02/2017] [Revised: 02/02/2017] [Accepted: 03/21/2017] [Indexed: 02/01/2023] Open
Abstract
PURPOSE Active matrix flat-panel imagers, which typically incorporate a pixelated array with one a-Si:H thin-film transistor (TFT) per pixel, have become ubiquitous by virtue of many advantages, including large monolithic construction, radiation tolerance, and high DQE. However, at low exposures such as those encountered in fluoroscopy, digital breast tomosynthesis and breast computed tomography, DQE is degraded due to the modest average signal generated per interacting x-ray relative to electronic additive noise levels of ~1000 e, or greater. A promising strategy for overcoming this limitation is to introduce an amplifier into each pixel, referred to as the active pixel (AP) concept. Such circuits provide in-pixel amplification prior to readout as well as facilitate correlated multiple sampling, enhancing signal-to-noise and restoring DQE at low exposures. In this study, a methodology for theoretically investigating the signal and noise performance of imaging array designs is introduced and applied to the case of AP circuits based on low-temperature polycrystalline silicon (poly-Si), a semiconductor suited to manufacture of large area, radiation tolerant arrays. METHODS Computer simulations employing an analog circuit simulator and performed in the temporal domain were used to investigate signal characteristics and major sources of electronic additive noise for various pixel amplifier designs. The noise sources include photodiode shot noise and resistor thermal noise, as well as TFT thermal and flicker noise. TFT signal behavior and flicker noise were parameterized from fits to measurements performed on individual poly-Si test TFTs. The performance of three single-stage and three two-stage pixel amplifier designs were investigated under conditions relevant to fluoroscopy. The study assumes a 20 × 20 cm2 , 150 μm pitch array operated at 30 fps and coupled to a CsI:Tl x-ray converter. Noise simulations were performed as a function of operating conditions, including sampling mode, of the designs. The total electronic additive noise included noise contributions from each circuit component. RESULTS The total noise results were found to exhibit a strong dependence on circuit design and operating conditions, with TFT flicker noise generally found to be the dominant noise contributor. For the single-stage designs, significantly increasing the size of the source-follower TFT substantially reduced flicker noise - with the lowest total noise found to be ~574 e [rms]. For the two-stage designs, in addition to tuning TFT sizes and introducing a low-pass filter, replacing a p-type TFT with a resistor (under the assumption in the study that resistors make no flicker noise contribution) resulted in significant noise reduction - with the lowest total noise found to be ~336 e [rms]. CONCLUSIONS A methodology based on circuit simulations which facilitates comprehensive explorations of signal and noise characteristics has been developed and applied to the case of poly-Si AP arrays. The encouraging results suggest that the electronic additive noise of such devices can be substantially reduced through judicious circuit design, signal amplification, and multiple sampling. This methodology could be extended to explore the noise performance of arrays employing other pixel circuitry such as that for photon counting as well as other semiconductor materials such as a-Si:H and a-IGZO.
Collapse
Affiliation(s)
- Martin Koniczek
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI, 48109, USA
| | - Larry E Antonuk
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI, 48109, USA
| | - Youcef El-Mohri
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI, 48109, USA
| | - Albert K Liang
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI, 48109, USA
| | - Qihua Zhao
- Department of Radiation Oncology, University of Michigan, Ann Arbor, MI, 48109, USA
| |
Collapse
|
6
|
Mackenzie A, Marshall NW, Hadjipanteli A, Dance DR, Bosmans H, Young KC. Characterisation of noise and sharpness of images from four digital breast tomosynthesis systems for simulation of images for virtual clinical trials. Phys Med Biol 2017; 62:2376-2397. [PMID: 28151431 DOI: 10.1088/1361-6560/aa5dd9] [Citation(s) in RCA: 22] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/29/2022]
Abstract
In-depth evaluation of the noise and sharpness characteristics of FujiFilm Innovality, GE SenoClaire, Hologic Selenia Dimensions and Siemens Inspiration digital breast tomosynthesis (DBT) systems was performed with the intention of improving image simulation for virtual clinical trials. Noise power spectra (NPS) and modulation transfer function curves (MTF) were measured for planar modes and for the first and central projections for DBT modes. In DBT mode, the x-ray beam was blocked for the projections before the central projection in order to remove the influence of lag and ghosting from the previous images. A quadratic fit between the NPS and linearised pixel value gave the noise coefficients for planar and DBT imaging modes. The spatial frequencies corresponding to an MTF of 0.5 (MTF0.5) were calculated from the MTF measurements made on the breast support and at 40 mm above the breast support. This was done for the first and the central projections. The percentage of signal carried over from the first projection to subsequent images (lag) was measured using a slit. The noise associated with lag was also evaluated. The DBT modes typically had lower electronic noise coefficients but higher structural noise coefficients compared to the respective planar mode MTF0.5 measured 40 mm above the table was between 6% and 47% lower for continuous scanning systems compared to 1% lower for step and shoot systems. For wide angle DBT, the MTF0.5 of the first projection was 18% (FujiFilm) and 28% (Siemens) lower than for the central projection. Lag in the second projection was 2.2%, 0.3%, 0.8% for the FujiFilm, GE and Hologic systems respectively. In all cases, the noise associated with lag was negligible. Current modelling frameworks for virtual clinical trials of breast DBT systems need to be adapted to account for signals from lag and variations in the MTF at wide angles.
Collapse
Affiliation(s)
- Alistair Mackenzie
- National Coordinating Centre for the Physics of Mammography, Royal Surrey County Hospital, Guildford, Surrey GU2 7XX, United Kingdom
| | | | | | | | | | | |
Collapse
|
7
|
Jee KW, Zhang R, Bentefour EH, Doolan PJ, Cascio E, Sharp G, Flanz J, Lu HM. Investigation of time-resolved proton radiography using x-ray flat-panel imaging system. Phys Med Biol 2017; 62:1905-1919. [DOI: 10.1088/1361-6560/aa5a43] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
|
8
|
Xu J, Sisniega A, Zbijewski W, Dang H, Stayman JW, Wang X, Foos DH, Aygun N, Koliatsos VE, Siewerdsen JH. Evaluation of detector readout gain mode and bowtie filters for cone-beam CT imaging of the head. Phys Med Biol 2016; 61:5973-92. [PMID: 27435162 DOI: 10.1088/0031-9155/61/16/5973] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The effects of detector readout gain mode and bowtie filters on cone-beam CT (CBCT) image quality and dose were characterized for a new CBCT system developed for point-of-care imaging of the head, with potential application to diagnosis of traumatic brain injury, intracranial hemorrhage (ICH), and stroke. A detector performance model was extended to include the effects of detector readout gain on electronic digitization noise. The noise performance for high-gain (HG), low-gain (LG), and dual-gain (DG) detector readout was evaluated, and the benefit associated with HG mode in regions free from detector saturation was quantified. Such benefit could be realized (without detector saturation) either via DG mode or by incorporation of a bowtie filter. Therefore, three bowtie filters were investigated that varied in thickness and curvature. A polyenergetic gain correction method was developed to equalize the detector response between the flood-field and projection data in the presence of a bowtie. The effect of bowtie filters on dose, scatter-to-primary ratio, contrast, and noise was quantified in phantom studies, and results were compared to a high-speed Monte Carlo (MC) simulation to characterize x-ray scatter and dose distributions in the head. Imaging in DG mode improved the contrast-to-noise ratio (CNR) by ~15% compared to LG mode at a dose (D 0, measured at the center of a 16 cm CTDI phantom) of 19 mGy. MC dose calculations agreed with CTDI measurements and showed that bowtie filters reduce peripheral dose by as much as 50% at the same central dose. Bowtie filters were found to increase the CNR per unit square-root dose near the center of the image by ~5-20% depending on bowtie thickness, but reduced CNR in the periphery by ~10-40%. Images acquired at equal CTDIw with and without a bowtie demonstrated a 24% increase in CNR at the center of an anthropomorphic head phantom. Combining a thick bowtie filter with a short arc (180° + fan angle) scan centered on the posterior of the head reduced dose to the eye lens by up to 90%. Acquisition in DG mode (without a bowtie filter) was beneficial to the detection of small, low contrast lesions (e.g. subtle ICH) in CBCT. While bowtie filters were found to reduce dose, mitigate sensor saturation at the periphery in HG mode, and improve CNR at the center of the image, the image quality at the periphery was slightly reduced compared to DG mode, and the use of a bowtie required careful implementation of the polyenergetic flood-field correction to avoid artifacts.
Collapse
Affiliation(s)
- Jennifer Xu
- Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205, USA
| | | | | | | | | | | | | | | | | | | |
Collapse
|
9
|
Liu L, Antonuk LE, El-Mohri Y, Zhao Q, Jiang H. Theoretical investigation of the design and performance of a dual energy (kV and MV) radiotherapy imager. Med Phys 2015; 42:2072-84. [DOI: 10.1118/1.4915120] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/20/2022] Open
|
10
|
Comparison of Digital Flat-Panel Detector and Conventional Angiography Machines: Evaluation of Stent Detection Rates, Visibility Scores, and Dose-Area Products. AJR Am J Roentgenol 2012; 198:946-54. [DOI: 10.2214/ajr.11.7245] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/03/2023]
|
11
|
Abstract
PURPOSE The authors examined the effect of detector lag on the noise power spectrum (NPS) of CT images reconstructed with filtered backprojection (FBP). METHODS The authors derived an analytical expression of the NPS with detector lag, and then verified it using computer simulations with parallel beam and fan beam geometries. The dependence of the NPS on the amount of lag, location within the scanned field of view (FOV), and the number of views used in the reconstruction (samples per rotation) was investigated using constant and view dependent noise in the raw data. RESULTS Detector lag introduces noise correlation in the azimuthal direction. The effect on the NPS is a frequency dependent reduction in amplitude. In small regions of the image, the effect is primarily in the frequencies corresponding to the azimuthal direction. The noise blurring and NPS filtering increases with increasing radial distance, and therefore regions at larger radial distances have lower noise power. With the same detector lag response function, the amount of noise correlation and NPS filtering decreases with increasing number of views. CONCLUSIONS The shape of the NPS depends on the detector lag coefficients, location of the region, and the number of views used in the reconstruction. In general, the noise correlation caused by detector lag decreased the amplitude of the NPS.
Collapse
Affiliation(s)
- Jongduk Baek
- Department of Radiology, Stanford University, Stanford, California 94305, USA.
| | | |
Collapse
|
12
|
Bogaert E, Bacher K, Lapere R, Thierens H. Does digital flat detector technology tip the scale towards better image quality or reduced patient dose in interventional cardiology? Eur J Radiol 2009; 72:348-53. [DOI: 10.1016/j.ejrad.2008.07.028] [Citation(s) in RCA: 18] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/20/2008] [Accepted: 07/28/2008] [Indexed: 11/28/2022]
|
13
|
Antonuk LE, Zhao Q, El-Mohri Y, Du H, Wang Y, Street RA, Ho J, Weisfield R, Yao W. An investigation of signal performance enhancements achieved through innovative pixel design across several generations of indirect detection, active matrix, flat-panel arrays. Med Phys 2009; 36:3322-39. [PMID: 19673228 DOI: 10.1118/1.3049602] [Citation(s) in RCA: 19] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/05/2023] Open
Abstract
Active matrix flat-panel imager (AMFPI) technology is being employed for an increasing variety of imaging applications. An important element in the adoption of this technology has been significant ongoing improvements in optical signal collection achieved through innovations in indirect detection array pixel design. Such improvements have a particularly beneficial effect on performance in applications involving low exposures and/or high spatial frequencies, where detective quantum efficiency is strongly reduced due to the relatively high level of additive electronic noise compared to signal levels of AMFPI devices. In this article, an examination of various signal properties, as determined through measurements and calculations related to novel array designs, is reported in the context of the evolution of AMFPI pixel design. For these studies, dark, optical, and radiation signal measurements were performed on prototype imagers incorporating a variety of increasingly sophisticated array designs, with pixel pitches ranging from 75 to 127 microm. For each design, detailed measurements of fundamental pixel-level properties conducted under radiographic and fluoroscopic operating conditions are reported and the results are compared. A series of 127 microm pitch arrays employing discrete photodiodes culminated in a novel design providing an optical fill factor of approximately 80% (thereby assuring improved x-ray sensitivity), and demonstrating low dark current, very low charge trapping and charge release, and a large range of linear signal response. In two of the designs having 75 and 90 microm pitches, a novel continuous photodiode structure was found to provide fill factors that approach the theoretical maximum of 100%. Both sets of novel designs achieved large fill factors by employing architectures in which some, or all of the photodiode structure was elevated above the plane of the pixel addressing transistor. Generally, enhancement of the fill factor in either discrete or continuous photodiode arrays was observed to result in no degradation in MTF due to charge sharing between pixels. While the continuous designs exhibited relatively high levels of charge trapping and release, as well as shorter ranges of linearity, it is possible that these behaviors can be addressed through further refinements to pixel design. Both the continuous and the most recent discrete photodiode designs accommodate more sophisticated pixel circuitry than is present on conventional AMFPIs--such as a pixel clamp circuit, which is demonstrated to limit signal saturation under conditions corresponding to high exposures. It is anticipated that photodiode structures such as the ones reported in this study will enable the development of even more complex pixel circuitry, such as pixel-level amplifiers, that will lead to further significant improvements in imager performance.
Collapse
Affiliation(s)
- Larry E Antonuk
- Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan 48109, USA.
| | | | | | | | | | | | | | | | | |
Collapse
|
14
|
Abstract
Flat detectors are the heart of high-performance imaging systems that provide new capabilities as well as new hazards. The superior image quality enables operators to work with heavier patients and steeper projections. Under these conditions, exposure control computers automatically increase the production of x-rays to compensate for absorption by body tissues. Image quality is preserved, and operators may not be aware of the very high skin doses delivered during prolonged procedures. Although it is assumed that flat detector systems are safe, the potential for radiation overexposure and skin injury is real. This article examines the unique radiation hazards of flat detector fluoroscopy and suggests practical steps that clinicians can take to protect themselves and their patients from radiation injury.
Collapse
Affiliation(s)
- John C Gurley
- The Linda and Jack Gill Heart Institute, University of Kentucky Medical Center, 760 South Limestone Street, Lexington, KY 40513, USA.
| |
Collapse
|
15
|
El-Mohri Y, Antonuk LE, Koniczek M, Zhao Q, Li Y, Street RA, Lu JP. Active pixel imagers incorporating pixel-level amplifiers based on polycrystalline-silicon thin-film transistors. Med Phys 2009; 36:3340-55. [PMID: 19673229 PMCID: PMC2805355 DOI: 10.1118/1.3116364] [Citation(s) in RCA: 12] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/16/2008] [Revised: 03/17/2009] [Accepted: 03/18/2009] [Indexed: 12/31/2022] Open
Abstract
Active matrix, flat-panel imagers (AMFPIs) employing a 2D matrix of a-Si addressing TFTs have become ubiquitous in many x-ray imaging applications due to their numerous advantages. However, under conditions of low exposures and/or high spatial resolution, their signal-to-noise performance is constrained by the modest system gain relative to the electronic additive noise. In this article, a strategy for overcoming this limitation through the incorporation of in-pixel amplification circuits, referred to as active pixel (AP) architectures, using polycrystalline-silicon (poly-Si) TFTs is reported. Compared to a-Si, poly-Si offers substantially higher mobilities, enabling higher TFT currents and the possibility of sophisticated AP designs based on both n- and p-channel TFTs. Three prototype indirect detection arrays employing poly-Si TFTs and a continuous a-Si photodiode structure were characterized. The prototypes consist of an array (PSI-1) that employs a pixel architecture with a single TFT, as well as two arrays (PSI-2 and PSI-3) that employ AP architectures based on three and five TFTs, respectively. While PSI-1 serves as a reference with a design similar to that of conventional AMFPI arrays, PSI-2 and PSI-3 incorporate additional in-pixel amplification circuitry. Compared to PSI-1, results of x-ray sensitivity demonstrate signal gains of approximately 10.7 and 20.9 for PSI-2 and PSI-3, respectively. These values are in reasonable agreement with design expectations, demonstrating that poly-Si AP circuits can be tailored to provide a desired level of signal gain. PSI-2 exhibits the same high levels of charge trapping as those observed for PSI-1 and other conventional arrays employing a continuous photodiode structure. For PSI-3, charge trapping was found to be significantly lower and largely independent of the bias voltage applied across the photodiode. MTF results indicate that the use of a continuous photodiode structure in PSI-1, PSI-2, and PSI-3 results in optical fill factors that are close to unity. In addition, the greater complexity of PSI-2 and PSI-3 pixel circuits, compared to that of PSI-1, has no observable effect on spatial resolution. Both PSI-2 and PSI-3 exhibit high levels of additive noise, resulting in no net improvement in the signal-to-noise performance of these early prototypes compared to conventional AMFPIs. However, faster readout rates, coupled with implementation of multiple sampling protocols allowed by the nondestructive nature of pixel readout, resulted in a significantly lower noise level of approximately 560 e (rms) for PSI-3.
Collapse
Affiliation(s)
- Youcef El-Mohri
- Department of Radiation Oncology, University of Michigan Medical Center, Ann Arbor, Michigan 48109 USA.
| | | | | | | | | | | | | |
Collapse
|
16
|
Method for reducing noise in X-ray images by averaging pixels based on the normalized difference with the relevant pixel. Radiol Phys Technol 2008; 1:188-95. [DOI: 10.1007/s12194-008-0028-z] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/26/2008] [Revised: 05/26/2008] [Accepted: 05/26/2008] [Indexed: 10/21/2022]
|
17
|
Kuhls-Gilcrist A, Yadava G, Patel V, Jain A, Bednarek DR, Rudin S. The Solid-State X-Ray Image Intensifier (SSXII): An EMCCD-Based X-Ray Detector. PROCEEDINGS OF SPIE--THE INTERNATIONAL SOCIETY FOR OPTICAL ENGINEERING 2008; 6913:69130K. [PMID: 18836568 DOI: 10.1117/12.772724] [Citation(s) in RCA: 16] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/14/2022]
Abstract
The solid-state x-ray image intensifier (SSXII) is an EMCCD-based x-ray detector designed to satisfy an increasing need for high-resolution real-time images, while offering significant improvements over current flat panel detectors (FPDs) and x-ray image intensifiers (XIIs). FPDs are replacing XIIs because they reduce/eliminate veiling glare, pincushion or s-shaped distortions and are physically flat. However, FPDs suffer from excessive lag and ghosting and their performance has been disappointing for low-exposure-per-frame procedures due to excessive instrumentation-noise. XIIs and FPDs both have limited resolution capabilities of ~3 cycles/mm. To overcome these limitations a prototype SSXII module has been developed, consisting of a 1k x 1k, 8 mum pixel EMCCD with a fiber-optic input window, which views a 350 mum thick CsI(Tl) phosphor via a 4:1 magnifying fiber-optic-taper (FOT). Arrays of such modules will provide a larger field-of-view. Detector MTF, DQE, and instrumentation-noise equivalent exposure (INEE) were measured to evaluate the SSXIIs performance using a standard x-ray spectrum (IEC RQA5), allowing for comparison with current state-of-the-art detectors. The MTF was 0.20 at 3 cycles/mm, comparable to standard detectors, and better than 0.05 up to 7 cycles/mm, well beyond current capabilities. DQE curves indicate no degradation from high-angiographic to low-fluoroscopic exposures (< 2% deviation in overall DQE from 1.3 mR to 2.7 muR), demonstrating negligible instrumentation-noise, even with low input signal intensities. An INEE of < 0.2 muR was measured for the highest-resolution mode (32 mum effective pixel size). Comparison images between detector technologies qualitatively demonstrate these improved imaging capabilities provided by the SSXII.
Collapse
Affiliation(s)
- Andrew Kuhls-Gilcrist
- University at Buffalo (State University of New York), Toshiba Stroke Research Center, 3435 Main St., Buffalo, NY USA 14214
| | | | | | | | | | | |
Collapse
|
18
|
Du H, Antonuk LE, El-Mohri Y, Zhao Q, Su Z, Yamamoto J, Wang Y. Investigation of the signal behavior at diagnostic energies of prototype, direct detection, active matrix, flat-panel imagers incorporating polycrystalline HgI2. Phys Med Biol 2008; 53:1325-51. [PMID: 18296765 DOI: 10.1088/0031-9155/53/5/011] [Citation(s) in RCA: 18] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
Active matrix, flat-panel x-ray imagers based on a-Si:H thin-film transistors offer many advantages and are widely utilized in medical imaging applications. Unfortunately, the detective quantum efficiency (DQE) of conventional flat-panel imagers incorporating scintillators or a-Se photoconductors is significantly limited by their relatively modest signal-to-noise ratio, particularly in applications involving low x-ray exposures or high spatial resolution. For this reason, polycrystalline HgI2 is of considerable interest by virtue of its low effective work function, high atomic number and the possibility of large-area deposition. In this study, a detailed investigation of the properties of prototype, flat-panel arrays coated with two forms of this high-gain photoconductor are reported. Encouragingly, high x-ray sensitivity, low dark current and spatial resolution close to the theoretical limits were observed from a number of prototypes. In addition, input-quantum-limited DQE performance was measured from one of the prototypes at relatively low exposures. However, high levels of charge trapping, lag and polarization, as well as pixel-to-pixel variations in x-ray sensitivity are of concern. While the results of the current study are promising, further development will be required to realize prototypes exhibiting the characteristics necessary to allow practical implementation of this approach.
Collapse
Affiliation(s)
- Hong Du
- Department of Radiation Oncology, University of Michigan Medical Center, Ann Arbor, Michigan 48109, USA
| | | | | | | | | | | | | |
Collapse
|
19
|
Blakesley JC, Speller R. Modeling the imaging performance of prototype organic x-ray imagers. Med Phys 2007; 35:225-39. [DOI: 10.1118/1.2805479] [Citation(s) in RCA: 26] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
|
20
|
El-Mohri Y, Antonuk LE, Zhao Q, Wang Y, Li Y, Du H, Sawant A. Performance of a high fill factor, indirect detection prototype flat-panel imager for mammography. Med Phys 2006; 34:315-27. [PMID: 17278517 DOI: 10.1118/1.2403967] [Citation(s) in RCA: 35] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
Empirical and theoretical investigations of the performance of a small-area, high-spatial-resolution, active matrix flat-panel imager, operated under mammographic conditions, is reported. The imager is based on an indirect detection array incorporating a continuous photodiode design, as opposed to the discrete photodiode design employed in conventional flat-panel imagers. Continuous photodiodes offer the prospect of higher fill factors, particularly for arrays with pixel pitches below approximately 100 microm. The array has a pixel-to-pixel pitch of 75 microm and a pixel format of 512 x 512, resulting in an active area of approximately 3.8 x 3.8 cm2. The array was coupled to two commercially available, structured CsI: Tl scintillators of approximately 150 microm thickness: one optimized for high light output (FOS-HL) and the other for high spatial resolution (FOS-HR), resulting in a pair of imager configurations. Measurements of sensitivity, modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE) were performed with a 26 kVp mammography beam at exposures ranging from approximately 0.5 to approximately 19 mR. MTF results from both CsI:Tl scintillators show that the array demonstrates good spatial resolution, indicating effective isolation between adjacent pixels. The effect of additive noise of the system on DQE was observed to be significantly higher for the FOS-HR scintillator compared to the FOS-HL scintillator due to lower sensitivity of the former. For the FOS-HL scintillator, DQE performance was generally high at high exposures, limited by the x-ray quantum efficiency, Swank factor and the MTF of the scintillators. For both scintillators, the DQE performance degrades at lower exposures due to the relatively large contribution of additive noise. Theoretical calculations based on a cascaded systems model were found to be in general agreement with the empirically determined NPS and DQE values. Finally, such calculations were used to predict potential DQE performance for hypothetical 50 microm pixel pitch imagers, employing similar continuous photodiode design and realistic inputs derived from the empirical measurements.
Collapse
Affiliation(s)
- Youcef El-Mohri
- Department of Radiation Oncology, University of Michigan Medical Center, Ann Arbor Michigan 48109, USA.
| | | | | | | | | | | | | |
Collapse
|
21
|
Jiang Y, Wilson DL. Optimization of detector pixel size for stent visualization in x-ray fluoroscopy. Med Phys 2006; 33:668-78. [PMID: 16878570 DOI: 10.1118/1.2169907] [Citation(s) in RCA: 14] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
Pixel size is of great interest in the flat-panel detector design because of its potential impact on image quality. In the particular case of angiographic x-ray fluoroscopy, small pixels are required in order to adequately visualize interventional devices such as guidewires and stents which have wire diameters as small as 200 and 50 microm, respectively. We used quantitative experimental and modeling techniques to investigate the optimal pixel size for imaging stents. Image quality was evaluated by the ability of subjects to perform two tasks: detect the presence of a stent and discriminate a partially deployed stent from a fully deployed one in synthetic images. With measurements at 50, 100, 200, and 300 microm, the 100 microm pixel size gave the maximum contrast sensitivity for the detection experiment with the idealized direct detector. For an idealized indirect detector with a scintillating layer, an optimal pixel size was obtained at 200 microm pixel size. A channelized human observer model predicted a peak at 150 and 170 microm, for the idealized direct and indirect detectors, respectively. With regard to the stent deployment task for both detector types, smaller pixel sizes are favored and there is a steep drop in performance with larger pixels. In general, with the increasing exposures, the model and measurements give the enhanced contrast sensitivities and a smaller optimal pixel size. The effects of electronic noise and fill factor were investigated using the model. We believe that the experimental results and human observer model predications can help guide the flat-panel detector design. In addition, the human observer model should work on the similar images and be applicable to the future model and actual flat-panel implementations.
Collapse
Affiliation(s)
- Yuhao Jiang
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106, USA
| | | |
Collapse
|
22
|
Abstract
We developed a computer noise simulation model for cone beam computed tomography imaging using a general purpose PC cluster. This model uses a mono-energetic x-ray approximation and allows us to investigate three primary performance components, specifically quantum noise, detector blurring and additive system noise. A parallel random number generator based on the Weyl sequence was implemented in the noise simulation and a visualization technique was accordingly developed to validate the quality of the parallel random number generator. In our computer simulation model, three-dimensional (3D) phantoms were mathematically modelled and used to create 450 analytical projections, which were then sampled into digital image data. Quantum noise was simulated and added to the analytical projection image data, which were then filtered to incorporate flat panel detector blurring. Additive system noise was generated and added to form the final projection images. The Feldkamp algorithm was implemented and used to reconstruct the 3D images of the phantoms. A 24 dual-Xeon PC cluster was used to compute the projections and reconstructed images in parallel with each CPU processing 10 projection views for a total of 450 views. Based on this computer simulation system, simulated cone beam CT images were generated for various phantoms and technique settings. Noise power spectra for the flat panel x-ray detector and reconstructed images were then computed to characterize the noise properties. As an example among the potential applications of our noise simulation model, we showed that images of low contrast objects can be produced and used for image quality evaluation.
Collapse
Affiliation(s)
- Shu-Ju Tu
- Department of Imaging Physics, The University of Texas M D Anderson Cancer Center, Houston, TX 77030-4009, USA.
| | | | | |
Collapse
|
23
|
Su Z, Antonuk LE, El-Mohri Y, Hu L, Du H, Sawant A, Li Y, Wang Y, Yamamoto J, Zhao Q. Systematic investigation of the signal properties of polycrystalline HgI2 detectors under mammographic, radiographic, fluoroscopic and radiotherapy irradiation conditions. Phys Med Biol 2005; 50:2907-28. [PMID: 15930610 DOI: 10.1088/0031-9155/50/12/012] [Citation(s) in RCA: 32] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
The signal properties of polycrystalline mercuric iodide (HgI2) film detectors, under irradiation conditions relevant to mammographic, radiographic, fluoroscopic and radiotherapy x-ray imaging, are reported. Each film detector consists of an approximately 230 to approximately 460 microm thick layer of HgI2 (fabricated through physical vapour deposition or a screen-print process) and a thin barrier layer, sandwiched between a pair of opposing electrode plates. The high atomic number, high density and low effective ionization energy, W(EFF), of HgI2 make it an attractive candidate for significantly improving the performance of active matrix, flat-panel imagers (AMFPIs) for several x-ray imaging applications. The temporal behaviour of current from the film detectors in the presence and in the absence of radiation was used to examine dark current levels, the lag and reciprocity of the signal response, x-ray sensitivity and W(EFF). The results are discussed in the context of present AMFPI performance. This study provides performance data for a wide range of potential medical x-ray imaging applications from a single set of detectors and represents the first investigation of the signal properties of polycrystalline mercuric iodide for the radiotherapy application.
Collapse
Affiliation(s)
- Zhong Su
- Department of Radiation Oncology, University of Michigan, 519 W. William St., Argus Building 1, Ann Arbor, MI 48103-4943, USA
| | | | | | | | | | | | | | | | | | | |
Collapse
|
24
|
Sawant A, Antonuk LE, El-Mohri Y, Li Y, Su Z, Wang Y, Yamamoto J, Zhao Q, Du H, Daniel J, Street R. Segmented phosphors: MEMS-based high quantum efficiency detectors for megavoltage x-ray imaging. Med Phys 2005; 32:553-65. [PMID: 15789602 DOI: 10.1118/1.1854774] [Citation(s) in RCA: 35] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
Current electronic portal imaging devices (EPIDs) based on active matrix flat panel imager (AMFPI) technology use a metal plate+phosphor screen combination for x-ray conversion. As a result, these devices face a severe trade-off between x-ray quantum efficiency (QE) and spatial resolution, thus, significantly limiting their imaging performance. In this work, we present a novel detector design for indirect detection-based AMFPI EPIDs that aims to circumvent this trade-off. The detectors were developed using micro-electro-mechanical system (MEMS)-based fabrication techniques and consist of a grid of up to approximately 2 mm tall, optically isolated cells of a photoresist material, SU-8. The cells are dimensionally matched to the pixels of the AMFPI array, and packed with a scintillating phosphor. In this paper, various design considerations for such detectors are examined. An empirical evaluation of three small-area (approximately 7 x 7 cm2) prototype detectors is performed in order to study the effects of two design parameters--cell height and phosphor packing density, both of which are important determinants of the imaging performance. Measurements of the x-ray sensitivity, modulation transfer function (MTF) and noise power spectrum (NPS) were performed under radiotherapy conditions (6 MV), and the detective quantum efficiency (DQE) was determined for each prototype SU-8 detector. In addition, theoretical calculations using Monte Carlo simulations were performed to determine the QE of each detector, as well as the inherent spatial resolution due to the spread of absorbed energy. The results of the present studies were compared with corresponding measurements published in an earlier study using a Lanex Fast-B phosphor screen coupled to an indirect detection array of the same design. The SU-8 detectors exhibit up to 3 times higher QE, while achieving spatial resolution comparable or superior to Lanex Fast-B. However, the DQE performance of these early prototypes is significantly lower than expected due to high levels of optical Swank noise. Consequently, the SU-8 detectors presently exhibit DQE values comparable to Lanex Fast-B at zero spatial frequency and significantly lower than Fast-B at higher frequencies. Finally, strategies for reducing Swank noise are discussed and theoretical calculations, based on the cascaded systems model, are presented in order to estimate the performance improvement that can be achieved through such noise reduction.
Collapse
Affiliation(s)
- Amit Sawant
- Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan 48103, USA
| | | | | | | | | | | | | | | | | | | | | |
Collapse
|
25
|
Tsapaki V, Kottou S, Kollaros N, Kyriakidis Z, Neofotistou V. Comparison of a CCD and a flat-panel digital system in an Interventional Cardiology Laboratory. RADIATION PROTECTION DOSIMETRY 2005; 117:93-6. [PMID: 16461500 DOI: 10.1093/rpd/nci746] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/06/2023]
Abstract
We evaluated the performances of angiographic units equipped with a flat-panel (FP) detector and image intensifier (II) charge-coupled device (CCD) in the Interventional Cardiology (IC) Department. Entrance dose rate and dose per image, along with the dose at the II level were measured using 2 mm copper sheets to simulate a patient. Image quality (IQ) was evaluated using a phantom. Doses increased with fluoroscopy level changing from low to high. FP presented higher doses than CCD. Periodic measurements showed differences of up to 35%. Low mode IQ did not significantly differ from normal and high mode for both systems. Low fluoroscopy mode was decided to be used routinely. Both X-ray systems performed within international recommendations for conventional systems with the exception of higher cine radiation doses and II dose rates, stressing the fact that more studies are required to investigate whether dose levels should be adjusted.
Collapse
Affiliation(s)
- V Tsapaki
- Medical Physics Department, Konstantopoulio Agia Olga Hospital, 3-5 Agias Olgas, Nea Ionia, 14233 Athens, Greece
| | | | | | | | | |
Collapse
|
26
|
Tsapaki V, Kottou S, Kollaros N, Dafnomili P, Koutelou M, Vano E, Neofotistou V. Comparison of a conventional and a flat-panel digital system in interventional cardiology procedures. Br J Radiol 2004; 77:562-7. [PMID: 15238402 DOI: 10.1259/bjr/83257582] [Citation(s) in RCA: 59] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022] Open
Abstract
The purpose of the study was to analyse the technical characteristics of a newly installed flat-panel fluoroscopy (FPF) system in an interventional cardiology (IC) department and compare it with an older conventional system. A patient survey was performed to investigate the radiation doses delivered by the X-ray systems. Finally, methods of technique optimization regarding the new digital system were investigated. Dose rates in all fluoroscopic and cine modes were measured and image quality assessed using a dedicated test tool. 200 patients were investigated, half using the conventional and half using the digital FPF system. Patient data collected were: sex, age, weight, height, dose-area product (DAP), fluoroscopy time (T) and total number of frames (F). Our results are: (1) Digital FPF system: high contrast resolution (HCR) is not affected by fluoroscopic mode, whereas low contrast resolution (LCR) is slightly decreased in the low mode. (2) The digital FPF system has 2.5 times better HCR than the conventional system, with 5 times lower dose in the fluoroscopy mode. (3) Median values of DAP, T and F, respectively, in coronary angiography (CA) are: 27.7 Gycm(2), 4.1 min and 876 for the digital and 39.3 Gycm(2), 5.3 min and 1600 for the conventional system. Median values for percutaneous transluminal coronary angioplasty (PTCA) are: 51.1 Gycm(2), 12.7 min and 1184 for the digital and 44.3 Gycm(2), 7.4 min and 1936 for the conventional system. Digital DAP in CA is reduced by 30%, suggesting that a dose reduction in the FPF system is possible. The results of the study concerning the FPF system lead to the conclusion that the lowest fluoroscopic mode and the lowest frame rate should be used in routine practice.
Collapse
Affiliation(s)
- V Tsapaki
- Medical Physics Department, Konstantopoulio Agia Olga Hospital, Medical Physics Department, Medical School, Athens University, Athens, Greece
| | | | | | | | | | | | | |
Collapse
|
27
|
Srinivas Y, Wilson DL. Quantitative image quality evaluation of pixel-binning in a flat-panel detector for x-ray fluoroscopy. Med Phys 2004; 31:131-41. [PMID: 14761029 DOI: 10.1118/1.1628278] [Citation(s) in RCA: 13] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
X-ray fluoroscopy places stringent design requirements on new flat-panel (FP) detectors, requiring both low-noise electronics and high data transfer rates. Pixel-binning, wherein data from more that one detector pixel are collected simultaneously, not only lowers the data transfer rate but also increases x-ray counts and pixel signal-to-noise ratio (SNR). In this study, we quantitatively assessed image quality of image sequences from four acquisition methods; no-binning and three types of binning; in synthetic images using a clinically relevant task of detecting an extended guidewire in a four-alternative forced-choice paradigm. Binning methods were conventional data-line (D) and gate-line (G) binning, and a novel method in which alternate frames in an image sequence used D and G binning. Two detector orientations placed the data lines either parallel or perpendicular to the guide wire. At a low exposure of 0.6 microR (1.548 x 10(-10) C/kg) per frame, irrespective of detector orientation, D binning with its reduced electronic noise was significantly (p<0.1) better than the other acquisition methods. On average, alternate binning performed better than G binning. At a higher exposure of 4.0 microR (10.32 x 10(-10) C/kg) per frame, with data lines parallel to the guidewire, detection with D binning was significantly (p<0.1) better than G binning. However, with data lines perpendicular to the guidewire, G binning was significantly (p<0.1) better than D binning because the partial area effect was reduced. Alternate binning was the best binning method when results were averaged over both orientations, and it was as good as the best binning method at either orientation. In addition, at low and high exposures, alternate binning gave a temporally fused image with a smooth guidewire, an important image quality feature not assessed in a detection experiment. While at high exposure, detection with no binning was as good, or better, than the best binning method, it might be impractical at fluoroscopy imaging rates. A computational observer model based on signal detection theory successfully fit data and was used to predict effects of similar acquisition methods. Results from this study suggest the use of exposure-dependent detector binning in fluoroscopy that switches between D binning and alternate binning at low and high exposures, respectively.
Collapse
Affiliation(s)
- Yogesh Srinivas
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106, USA
| | | |
Collapse
|
28
|
Bissonnette JP, Munro P, Cunningham IA. Monte Carlo simulation of the image formation process in portal imaging. Med Phys 2003; 30:3243-50. [PMID: 14713091 DOI: 10.1118/1.1625511] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
We have written Monte Carlo programs to simulate the formation of radiological images. Our code is used to propagate a simulated x-ray fluence through each component of an existing video-based portal imaging system. This simulated fluence consists of a 512 x 512 pixel image containing both contrast-detail patterns as well as checker patterns to assess spatial resolution of the simulated portal imager. All of the components of the portal imaging system were modeled as a cascade of eight linear stages. Using this code, one can assess the visual impact of changing components in the imaging chain by changing the appropriate probability density function. Virtual experiments were performed to assess the visual impact of replacing the lens and TV camera by an amorphous silicon array, and the effect of scattered radiation on portal images.
Collapse
Affiliation(s)
- Jean-Pierre Bissonnette
- Radiation Therapy Physics, Princess Margaret Hospital, 610 University Avenue, Toronto, Ontario M5G 2M9, Canada.
| | | | | |
Collapse
|
29
|
Jee KW, Antonuk LE, El-Mohri Y, Zhao Q. System performance of a prototype flat-panel imager operated under mammographic conditions. Med Phys 2003; 30:1874-90. [PMID: 12906206 DOI: 10.1118/1.1585051] [Citation(s) in RCA: 28] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
The results of an empirical and theoretical investigation of the performance of a high-resolution, active matrix flat-panel imager performed under mammographic conditions are reported. The imager is based upon a prototype, indirect detection active matrix array incorporating a discrete photodiode in each pixel and a pixel-to-pixel pitch of 97 microm. The investigation involved three imager configurations corresponding to the use of three different x-ray converters with the array. The converters were a conventional Gd2O2S-based mammographic phosphor screen (Min-R) and two structured CsI:Tl scintillators: one optimized for high spatial resolution (FOS-HR) and the other for high light output (FOS-HL). Detective quantum efficiency for mammographic exposures ranging from approximately 2 to approximately 40 mR at 26 kVp were determined for each imager configuration through measurements of x-ray sensitivity, modulation transfer function (MTF), and noise power spectrum (NPS). All configurations were found to provide significant presampling MTF at frequencies beyond the Nyquist frequency of the array, approximately 5.2 mm(-1) , consistent with the high spatial resolution of the converters. In addition, the effect of additive electronic noise on the NPS was found to be significantly larger for the configuration with lower system gain (FOS-HR) than for the configurations with higher gain (Min-R, FOS-HL). The maximum DQE values obtained with the CsI:Tl scintillators were considerably greater than those obtained with the Min-R screen due to the significantly lower Swank noise of the scintillators. Moreover, DQE performance was found to degrade with decreasing exposure, although this exposure-dependence was considerably reduced for the higher gain configurations. Theoretical calculations based on the cascaded systems model were found to be in generally good agreement with these empirically determined NPS and DQE values. In this study, we provide an example of how cascaded systems modeling can be used to identify factors limiting system performance and to examine trade-offs between factors toward the goal of maximizing performance.
Collapse
Affiliation(s)
- Kyung-Wook Jee
- Department of Radiation Oncology, University of Michigan Medical Center, Ann Arbor, Michigan 48109, USA.
| | | | | | | |
Collapse
|
30
|
Karim K, Nathan A, Rowlands J, Kasap S. X-ray detector with on-pixel amplification for large area diagnostic medical imaging. ACTA ACUST UNITED AC 2003. [DOI: 10.1049/ip-cds:20030602] [Citation(s) in RCA: 26] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/19/2022]
|