1
|
Real evidence to assess clinical testing interference risk (REACTIR): A strategy using real world data to assess the prevalence of interfering substances in patients undergoing clinical laboratory testing. Clin Chim Acta 2021; 523:178-184. [PMID: 34499870 DOI: 10.1016/j.cca.2021.09.001] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/29/2021] [Revised: 08/31/2021] [Accepted: 09/03/2021] [Indexed: 11/20/2022]
Abstract
INTRODUCTION Laboratory test interferences can cause spurious test results and patient harm. Knowing the frequency of various interfering substances in patient populations likely to be tested with a particular laboratory assay may inform test development, test utilization and strategies to mitigate interference risk. METHODS We developed REACTIR (Real Evidence to Assess Clinical Testing Interference Risk), an approach using real world data to assess the prevalence of various interfering substances in patients tested with a particular type of assay. REACTIR uses administrative real world data to identify and subgroup patient cohorts tested with a particular laboratory test and evaluate interference risk. RESULTS We demonstrate the application REACTIR to point of care (POC) blood glucose testing. We found that exposure to several substances with the potential to interfere in POC blood glucose tests, including N-acetyl cysteine (NAC) and high dose vitamin C was uncommon in most patients undergoing POC glucose tests with several key exceptions, such as burn patients receiving high dose IV-vitamin C or acetaminophen overdose patients receiving NAC. CONCLUSIONS Findings from REACTIR may support risk mitigation strategies including targeted clinician education and clinical decision support. Likewise, adaptations of REACTIR to premarket assay development may inform optimal assay design and assessment.
Collapse
|
2
|
Karon BS. Clarity on the Use of Glucose Meters for Critically Ill Hospitalized Patients, But One Big Question Remains to be Answered. J Appl Lab Med 2021; 6:813-815. [PMID: 34089609 DOI: 10.1093/jalm/jfab050] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/19/2021] [Accepted: 04/21/2021] [Indexed: 11/12/2022]
Affiliation(s)
- Brad S Karon
- Department of Laboratory Medicine and Pathology, Mayo Clinic Rochester, Rochester, MN USA
| |
Collapse
|
3
|
Abstract
BACKGROUND To summarize new evidence regarding the methodological aspects of blood glucose control in the intensive care unit (ICU). METHODS We reviewed the literature on blood glucose control in the ICU up to August 2019 through Ovid Medline and Pubmed. RESULTS Since the publication of the Leuven studies, the benefits of glycemic control have been recognized. However, the methodology of blood glucose control, notably the blood glucose measurement accuracy and the insulin titration protocol, plays an important but underestimated role. This may partially explain the negative results of the large, pragmatic multicenter trials and made everyone realize that tight glycemic control with less-frequent glucose measurements on less accurate blood glucose meters is neither feasible nor advisable in daily practice. Blood gas analyzers remain the gold standard. New generation point-of-care blood glucose meters may be an alternative when using whole blood of critically ill patients in combination with a clinically validated insulin dosing algorithm. CONCLUSION When implementing blood glucose management in an ICU one needs to take into account the interaction between aimed glycemic target and blood glucose measurement methodology.
Collapse
Affiliation(s)
- Gert-Jan Eerdekens
- Department of Anesthesiology, University Hospitals Leuven, Belgium
- Department of Anesthesia and Intensive Care Medicine, ZOL-Genk, Belgium
- Gert-Jan Eerdekens, MD, Department of Anesthesia UZ Leuven, Herestraat 49, Leuven 3000, Belgium.
| | - Steffen Rex
- Department of Anesthesiology, University Hospitals Leuven, Belgium
- Department of Cardiovascular Sciences, KU Leuven, Belgium
| | - Dieter Mesotten
- Department of Anesthesia and Intensive Care Medicine, ZOL-Genk, Belgium
- Faculty of Medicine and Life Sciences, UHasselt, Belgium
| |
Collapse
|
4
|
Christodouleas DC, Kaur B, Chorti P. From Point-of-Care Testing to eHealth Diagnostic Devices (eDiagnostics). ACS CENTRAL SCIENCE 2018; 4:1600-1616. [PMID: 30648144 PMCID: PMC6311959 DOI: 10.1021/acscentsci.8b00625] [Citation(s) in RCA: 96] [Impact Index Per Article: 16.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/04/2018] [Indexed: 05/09/2023]
Abstract
Point-of-care devices were originally designed to allow medical testing at or near the point of care by health-care professionals. Some point-of-care devices allow medical self-testing at home but cannot fully cover the growing diagnostic needs of eHealth systems that are under development in many countries. A number of easy-to-use, network-connected diagnostic devices for self-testing are needed to allow remote monitoring of patients' health. This Outlook highlights the essential characteristics of diagnostic devices for eHealth settings and indicates point-of-care technologies that may lead to the development of new devices. It also describes the most representative examples of simple-to-use, point-of-care devices that have been used for analysis of untreated biological samples.
Collapse
Affiliation(s)
| | - Balwinder Kaur
- Department of Chemistry, University
of Massachusetts Lowell, Lowell, Massachusetts 01854, United States
| | - Parthena Chorti
- Department of Chemistry, University
of Massachusetts Lowell, Lowell, Massachusetts 01854, United States
| |
Collapse
|
6
|
Manthei DM, Wesener N, Twarkowski D, Giacherio DA, Schroeder LF. Retrospective Accuracy Study of Point-of-Care Hemoglobin A1c in Field Conditions. Clin Chem 2017; 63:780-783. [DOI: 10.1373/clinchem.2016.267765] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Affiliation(s)
- David M Manthei
- Department of Pathology University of Michigan 1301 Catherine St. Ann Arbor, MI
| | - Nicholas Wesener
- Department of Pathology University of Michigan 1301 Catherine St. Ann Arbor, MI
| | - Denise Twarkowski
- Department of Pathology University of Michigan 1301 Catherine St. Ann Arbor, MI
| | - Donald A Giacherio
- Department of Pathology University of Michigan 1301 Catherine St. Ann Arbor, MI
| | - Lee F Schroeder
- Department of Pathology University of Michigan 1301 Catherine St. Ann Arbor, MI
| |
Collapse
|
7
|
Leopold JH, van Hooijdonk RTM, Boshuizen M, Winters T, Bos LD, Abu-Hanna A, Hoek AMT, Fischer JC, van Dongen-Lases EC, Schultz MJ. Point and trend accuracy of a continuous intravenous microdialysis-based glucose-monitoring device in critically ill patients: a prospective study. Ann Intensive Care 2016; 6:68. [PMID: 27436191 PMCID: PMC4951389 DOI: 10.1186/s13613-016-0171-3] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/08/2016] [Accepted: 07/04/2016] [Indexed: 11/24/2022] Open
Abstract
Background Microdialysis is a well-established technology that can be used for continuous blood glucose monitoring. We determined point and trend accuracy, and reliability of a microdialysis-based continuous blood glucose-monitoring device (EIRUS®) in critically ill patients. Methods Prospective study involving patients with an expected intensive care unit stay of ≥48 h. Every 15 min, device readings were compared with blood glucose values measured in arterial blood during blocks of 8 h per day for a maximum of 3 days. The Clarke error grid, Bland–Altman plot, mean absolute relative difference and glucose prediction error analysis were used to express point accuracy and the rate error grid to express trend accuracy. Reliability testing included aspects of the device and the external sensor, and the special central venous catheter (CVC) with a semipermeable membrane for use with this device. Results We collected 594 paired values in 12 patients (65 [26–80; 8–97] (median [IQR; total range]) paired values per patient). Point accuracy: 93.6 % of paired values were in zone A of the Clarke error grid, 6.4 % were in zone B; bias was 4.1 mg/dL with an upper limit of agreement of 28.6 mg/dL and a lower level of agreement of −20.5 mg/dL in the Bland–Altman analysis; 93.6 % of the values ≥75 mg/dL were within 20 % of the reference values in the glucose prediction error analysis; the mean absolute relative difference was 7.5 %. Trend accuracy: 96.4 % of the paired values were in zone A, and 3.3 and 0.3 % were in zone B and zone C of the rate error grid. Reliability: out of 16 sensors, 4 had to be replaced prematurely; out of 12 CVCs, two malfunctioned (one after unintentional flushing by unsupervised nurses of the ports connected to the internal microdialysis chamber, causing rupture of the semipermeable membrane; one for an unknown reason). Device start-up time was 58 [56–67] min; availability of real-time data was 100 % of the connection time. Conclusions In this study in critically ill patients who had no hypoglycemic episodes and a limited number of hyperglycemic excursions, point accuracy of the device was moderate to good. Trend accuracy was very good. The device had no downtimes, but 4 out of 16 external sensors and 2 out of 12 CVCs had practical problems. Electronic supplementary material The online version of this article (doi:10.1186/s13613-016-0171-3) contains supplementary material, which is available to authorized users.
Collapse
Affiliation(s)
- J H Leopold
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands. .,Department of Medical Informatics, Academic Medical Center, Amsterdam, The Netherlands.
| | - R T M van Hooijdonk
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands
| | - M Boshuizen
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands
| | - T Winters
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands
| | - L D Bos
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands
| | - A Abu-Hanna
- Department of Medical Informatics, Academic Medical Center, Amsterdam, The Netherlands
| | - A M T Hoek
- Department of Clinical Chemistry, Academic Medical Center, Amsterdam, The Netherlands
| | - J C Fischer
- Department of Clinical Chemistry, Academic Medical Center, Amsterdam, The Netherlands
| | - E C van Dongen-Lases
- Department of Clinical Chemistry, Academic Medical Center, Amsterdam, The Netherlands
| | - M J Schultz
- Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands.,Laboratory of Experimental Intensive care and Anesthesiology (L.E.I.C.A), Academic Medical Center, Amsterdam, The Netherlands
| |
Collapse
|
8
|
Nydegger U, Lung T, Risch L, Risch M, Medina Escobar P, Bodmer T. Inflammation Thread Runs across Medical Laboratory Specialities. Mediators Inflamm 2016; 2016:4121837. [PMID: 27493451 PMCID: PMC4963559 DOI: 10.1155/2016/4121837] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/31/2016] [Accepted: 05/31/2016] [Indexed: 12/16/2022] Open
Abstract
We work on the assumption that four major specialities or sectors of medical laboratory assays, comprising clinical chemistry, haematology, immunology, and microbiology, embraced by genome sequencing techniques, are routinely in use. Medical laboratory markers for inflammation serve as model: they are allotted to most fields of medical lab assays including genomics. Incessant coding of assays aligns each of them in the long lists of big data. As exemplified with the complement gene family, containing C2, C3, C8A, C8B, CFH, CFI, and ITGB2, heritability patterns/risk factors associated with diseases with genetic glitch of complement components are unfolding. The C4 component serum levels depend on sufficient vitamin D whilst low vitamin D is inversely related to IgG1, IgA, and C3 linking vitamin sufficiency to innate immunity. Whole genome sequencing of microbial organisms may distinguish virulent from nonvirulent and antibiotic resistant from nonresistant varieties of the same species and thus can be listed in personal big data banks including microbiological pathology; the big data warehouse continues to grow.
Collapse
Affiliation(s)
- Urs Nydegger
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| | - Thomas Lung
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| | - Lorenz Risch
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| | - Martin Risch
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| | - Pedro Medina Escobar
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| | - Thomas Bodmer
- Labormedizinisches Zentrum Dr. Risch and Kantonsspital Graubünden, 7000 Chur, Switzerland
| |
Collapse
|
9
|
van Hooijdonk RTM, Krinsley JS, Schultz MJ. DETECT the Extremes That Usually Remain Undetected in Conventional Observational Studies. Clin Chem 2016; 62:668-70. [PMID: 26988583 DOI: 10.1373/clinchem.2016.254250] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/03/2016] [Accepted: 02/08/2016] [Indexed: 12/16/2022]
Affiliation(s)
| | - James S Krinsley
- Division of Critical Care, Stamford Hospital, Columbia University College of Physicians and Surgeons, Stamford, CT
| | - Marcus J Schultz
- Department of Intensive Care and Laboratory of Experimental Intensive Care and Anesthesiology, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands
| |
Collapse
|