Frizzarin RM, Rocha FRP. A multi-pumping flow-based procedure with improved sensitivity for the spectrophotometric determination of acid-dissociable cyanide in natural waters.
Anal Chim Acta 2012;
758:108-13. [PMID:
23245902 DOI:
10.1016/j.aca.2012.10.059]
[Citation(s) in RCA: 15] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/24/2012] [Accepted: 10/30/2012] [Indexed: 12/11/2022]
Abstract
An analytical procedure with improved sensitivity was developed for cyanide determination in natural waters, exploiting the reaction with the complex of Cu(I) with 2,2'-biquinoline 4,4'-dicarboxylic acid (BCA). The flow system was based on the multi-pumping approach and long pathlength spectrophotometry with a flow cell based on a Teflon AF 2400(®) liquid core waveguide was exploited to increase sensitivity. A linear response was achieved from 5 to 200μg L(-1), with coefficient of variation of 1.5% (n=10). The detection limit and the sampling rate were 2μg L(-1) (99.7% confidence level), and 22h(-1), respectively. Per determination, 48ng of Cu(II), 5μg of ascorbic acid and 0.9μg of BCA were consumed. As high as 100mg L(-1) thiocyanate, nitrite or sulfite did not affect cyanide determination. Sulfide did not interfere at concentrations lower than 40 and 200μg L(-1) before or after sample pretreatment with hydrogen peroxide. The results for natural waters samples agreed with those obtained by a fluorimetric flow-based procedure at the 95% confidence level. The proposed procedure is then a reliable, fast and environmentally friendly alternative for cyanide determination in natural waters.
Collapse