Loyprasert S, Hedström M, Thavarungkul P, Kanatharana P, Mattiasson B. Sub-attomolar detection of cholera toxin using a label-free capacitive immunosensor.
Biosens Bioelectron 2010;
25:1977-83. [PMID:
20167466 DOI:
10.1016/j.bios.2010.01.020]
[Citation(s) in RCA: 71] [Impact Index Per Article: 5.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/12/2009] [Revised: 12/28/2009] [Accepted: 01/15/2010] [Indexed: 11/26/2022]
Abstract
A label-free immunosensor for the direct detection of cholera toxin (CT) at sub-attomolar level has been developed based on potential-step capacitance measurements. Anti-CT antibody was adsorbed on gold nanoparticles (AuNPs) incorporated on a polytyramine-modified gold electrode. The concentration of CT was determined by detecting the change of capacitance caused by the formation of antibody-antigen complexes. By using AuNPs adsorbed to the sensing surface, the signal was dramatically increased leading to a significantly more sensitive assay. In fact, under optimum conditions the immunosensor could detect CT concentration with a limit of detection of 9 x 10(-20)M or 0.09 aM, with a dynamic range between 0.1 aM and 10 pM. Good analytical reproducibility could be obtained by injecting CT up to 36 times with an RSD of 2.5%. In addition, good performance of the developed immunosensor was achieved when applied to turbid water samples collected from a local stream that were spiked with CT.
Collapse