Sabbioni C, Saracino MA, Mandrioli R, Pinzauti S, Furlanetto S, Gerra G, Raggi MA. Simultaneous liquid chromatographic analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol in human plasma.
J Chromatogr A 2004;
1032:65-71. [PMID:
15065778 DOI:
10.1016/j.chroma.2004.01.008]
[Citation(s) in RCA: 96] [Impact Index Per Article: 4.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/22/2022]
Abstract
The comparison of two HPLC methods, one with electrochemical detection and the other with coulometric detection, for the simultaneous analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma is presented. The careful pre-treatment of plasma samples is based on an innovative two-step procedure by means of solid-phase extraction (SPE) which uses one single hydrophilic-lipophilic balance cartridge. The extraction yield values found were higher than 85% for epinephrine, norepinephrine and MHPG, and higher than 70% for dopamine. The assays carried out on real plasma samples with the coulometric system gave good results in terms of sensitivity (limits of quantitation: 0.10-0.15 ng ml(-1) for catecholamines, 0.6 ng ml(-1) for MHPG) and selectivity, while interference was sometimes found when using the amperometric system. Precision was also satisfactory, with relative standard deviation values for intermediate precision always lower than 6%. The HPLC method with coulometric detection coupled to a novel SPE procedure is thus suitable for the simultaneous determination of catecholamines and MHPG in plasma of volunteers subjected to experimental stress.
Collapse