1
|
Iliaskou C, Rossi G, Sachpazidis I, Boronikolas V, Gainey M, Baltas D. Evaluation of RADIANCE Monte Carlo algorithm for treatment planning in electron based Intraoperative Radiotherapy (IOERT). Z Med Phys 2024:S0939-3889(23)00149-6. [PMID: 38182457 DOI: 10.1016/j.zemedi.2023.12.002] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/22/2023] [Revised: 12/01/2023] [Accepted: 12/04/2023] [Indexed: 01/07/2024]
Abstract
PURPOSE To perform experimental as well as independent Monte Carlo (MC) evaluation of the MC algorithm implemented in RADIANCE version 4.0.8, a dedicated treatment planning system (TPS) for 3D electron dose calculations in intraoperative radiation therapy (IOERT). METHODS AND MATERIALS The MOBETRON 2000 (IntraOp Medical Corporation, Sunnyvale, CA) IOERT accelerator was employed. PDD and profiles for five cylindrical plastic applicators with 50-90 mm diameter and 0°, 30° beveling were measured in a water phantom, at nominal energies of 6, 9 and 12 MeV. Additional PDD measurements were performed for all the energies without applicator. MC modeling of the MOBETRON was performed with the user code BEAMnrc and egs_chamber of the MC simulation toolkit EGSnrc. The generated phase space files of the two 0°-bevel applicators (50 mm, 80 mm) and three energies in both RADIANCE and BEAMnrc, were used to determine PDD and profiles in various set-ups of virtual water phantoms with air and bone inhomogeneities. 3D dose distributions were also calculated in image data sets of an anthropomorphic tissue-equivalent pelvis phantom. Image acquisitions were realized with a CT scanner (Philips Big Bore CT, Netherlands). Gamma analysis was applied to quantify the deviations of the RADIANCE calculations to the measurements and EGSnrc calculations. Gamma criteria normalized to the global maximum were investigated between 2%, 2 mm and 3%, 3 mm. RESULTS RADIANCE MC calculations satisfied the gamma criteria of 3%, 3 mm with a tolerance limit of 85% passing rate compared to in- water phantom measurements, except for the dose profiles of the 30° beveled applicators. Mismatches lay in surface doses, in umbra regions and in the beveled end of the 30° applicators. A very good agreement to the EGSnrc calculations in heterogeneous media was observed. Deviations were more pronounced for the larger applicator diameter and higher electron energy. In 3D dose comparisons in the anthropomorphic phantom, gamma passing rates were higher than 96 % for both simulated applicators. CONCLUSIONS RADIANCE MC algorithm agrees within 3%, 3 mm criteria with in-water phantom measurements and EGSnrc MC dose distributions in heterogeneous media for 0°-bevel applicators. The user should be aware of missing scattering components and the 30° beveled applicators should be used with attention.
Collapse
Affiliation(s)
- Charoula Iliaskou
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany.
| | - Giulio Rossi
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany
| | - Ilias Sachpazidis
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany
| | - Vasilios Boronikolas
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany
| | - Mark Gainey
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany
| | - Dimos Baltas
- Division of Medical Physics, Department of Radiation Oncology, University Medical Center, Freiburg 79106, Germany; German Cancer Consortium (DKTK), Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg 69120, Germany
| |
Collapse
|