Lee CH, Tan HC, Wong HB, Zhang XL, Fun S, Gay M, Qu L, Lim J, Low A, Lim YT. Incidence, predictors, and outcomes of device failure of X-sizer thrombectomy: real-world experience of 200 cases in 5 years.
Am Heart J 2007;
153:14.e13-9. [PMID:
17174629 DOI:
10.1016/j.ahj.2006.10.007]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/02/2006] [Accepted: 10/08/2006] [Indexed: 12/22/2022]
Abstract
BACKGROUND
Clinical studies have evaluated the safety and efficacy of adjunctive X-sizer (EndiCOR Inc, San Clemente, CA) in percutaneous coronary intervention (PCI). However, patient and lesion subsets were highly selected, and extrapolation of the results to daily practice is problematic.
METHODS
X-sizer thrombectomy was performed in 200 procedures from August 2000 to July 2005. The indications for the procedures were primary PCI in 71%. Device failure was defined as the occurrence of 1 or more of the following conditions: (a) failure of the X-sizer to reach the target segment, (b) failure to achieve final thrombolysis in myocardial infarction 3 flow, (c) slow flow or no-reflow, (d) distal embolization, and (e) coronary perforation.
RESULTS
Device failure occurred in 48 procedures, giving a device failure rate of 24%. Logistic regression analysis showed that ostial lesion was the only independent predictor of device failure (OR 4.89, 95% CI 1.23-19.51, P = .024). A total of 19 patients had developed 22 adverse events, giving a 30-day adverse event rate of 9.5%. Among these, there were 13 deaths, 4 strokes, 2 reinfarctions, 2 repeat PCIs, and 1 coronary artery bypass grafting. Logistic regression analysis showed that X-sizer device failure was independently associated with 30-day adverse events (OR 3.42, 95% CI 1.04-11.25, P = .043).
CONCLUSIONS
The incidence of device failure of X-sizer thrombectomy was 24%. Ostial lesion was an independent predictor of device failure. Notably, X-sizer device failure was independently associated with 30-day adverse event. These highlight the importance of case selection and the problem with its use in ostial lesions.