Iacovelli F, Burattini O, Sturdà F, Branca M, Stabile E, Fimiani L, Salemme L, Cioppa A, Pucciarelli A, Cafaro A, Bortone AS, Contegiacomo G, Pignatelli A, Tesorio T. Single suture-mediated closure system after transfemoral transcatheter aortic valve implantation: A single-center real-world experience.
Catheter Cardiovasc Interv 2024;
103:1125-1137. [PMID:
38639140 DOI:
10.1002/ccd.31054]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/05/2024] [Revised: 03/27/2024] [Accepted: 04/09/2024] [Indexed: 04/20/2024]
Abstract
BACKGROUND
Despite the use of two crossed Perclose ProGlide™ (Abbott Vascular Devices) is the most widespread technique to close the main arterial access in transfemoral transcatheter aortic valve implantation (TF-TAVI), the safest and most effective strategy still remains much debated.
AIMS
The aim of the present study was to evaluate the performance of a single Perclose ProGlide suture-mediated closure device to obtain femoral hemostasis after sheathless implantation of self-expanding transcatheter heart valves through their 14 F-equivalent fix delivery systems.
METHODS
This prospective observational study included 439 patients undergoing TF-TAVI at the "Montevergine" Clinic of Mercogliano, Italy. All patients underwent hemostasis of the large-bore access using a single Perclose ProGlide with preclose technique, after sheathless implantation of self-expanding transcatheter heart valves through 14 F-equivalent fix delivery systems. A multidetector computed tomography analysis of size, tortuosity, atherosclerotic, and calcification burdens of the ilio-femoral access route was made by a dedicated corelab. Vascular complications (VCs), percutaneous closure device (PCD) failure, and bleedings were adjudicated by a clinical events committee.
RESULTS
A total of 81 different VCs were observed in 60 patients (13.7%); among these, 41 (5% of patients) were categorized as major. PCD failure occurred in 14 patients (3.2%). At the logistic regression analysis, no predictors of PCD failure have been identified.
CONCLUSION
This registry suggests that the use of a single suture-mediated closure device could be considered a safe and efficient technique to achieve access site hemostasis in patients undergoing TF-TAVI through 14 F-equivalent fix delivery systems.
Collapse