O'Connor BK, Case CL, Wienecke MM, Gillette PC. Late endocardial pacing lead body migration compromising Fontan physiology.
Pacing Clin Electrophysiol 1996;
19:2147-9. [PMID:
8994956 DOI:
10.1111/j.1540-8159.1996.tb03291.x]
[Citation(s) in RCA: 1] [Impact Index Per Article: 0.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 02/03/2023]
Abstract
Migration of intracardiac transvenous pacing leads may occur. There is a known risk of intrapulmonary ventricular pacing lead migration in patients with endocardial lead systems. In the current report we present the late intrapulmonary migration of an endocardial atrial pacing lead body. The patient had undergone antitachycardia pacemaker placement to control recurrent atrial tachyarrhythmias following the Fontan procedure. Although the lead electrode remained in place and continued to pace, the lead body migrated, causing severe obstruction to blood flow. This resulted in severe cardiac decompensation, which was ultimately ameliorated by lead repositioning.
Collapse