Orbàn L, Sullivan JV, Chrambach A. A thin-layer multistrip agarose gel electrophoresis apparatus for Ferguson plot analysis at the sub-microgram load level.
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS 1989;
19:105-20. [PMID:
2809063 DOI:
10.1016/0165-022x(89)90055-9]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 01/02/2023]
Abstract
A method for the simultaneous horizontal agarose gel electrophoresis on thin-layer strips of different gel concentrations was developed for the purpose of generating Ferguson plots at the sub-microgram load level. Seven independent gel strips on a common GelBond support were formed by filling channels created by a comb-shaped spacer (polycarbonate) in a vertical multistrip cassette. Electrophoresis on a horizontal Peltier-cooled surface employed commercial apparatus (E-C Apparatus Corp.) with a modified cover which is airtight and holds anodic and cathodic voltage measurement probes for each strip. The application of the apparatus to Ferguson plot analysis in a single experiment was exemplified on the RNA-containing turnip crinkle virus (TCV) at a load of 50 ng/gel strip, using an optimized silver staining method (a modification of a procedure of FMC Corp. BioProducts) for detection. Within the range of 3.5 to 12.5 V/cm, the plot was found to be independent of field strength. Mobility is also independent of the concentration of detergent (CHAPS) up to 10 mM.
Collapse