Lomasney AR, Guillo C, Sidebottom AM, Roper MG. Optimization of capillary electrophoresis conditions for a glucagon competitive immunoassay using response surface methodology.
Anal Bioanal Chem 2009;
394:313-9. [PMID:
19189083 DOI:
10.1007/s00216-009-2622-2]
[Citation(s) in RCA: 22] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/03/2008] [Accepted: 01/14/2009] [Indexed: 11/30/2022]
Abstract
The capillary electrophoresis (CE) conditions for a competitive immunoassay of glucagon were optimized for highest sensitivity of the immunoassay and resolution of the electrophoretic peaks using a Box-Behnken design. Injection time, voltage ramp time, and separation voltage were varied between three levels and two responses, bound-to-free (B/F) ratio of the immunoassay peaks and resolution between the peaks, were measured. Analysis of variance was applied to fit a predictive model, and a desirability function was used to simultaneously optimize both responses. A 10-s injection, 1.6-min ramp time, and a 22-kV separation voltage were the conditions found when high B/F was given more emphasis than high resolution. To test the model, calibration curves of a glucagon immunoassay were measured at the optimum and least optimum CE conditions. Optimal conditions increased the sensitivity of the immunoassay by 388% compared to the least optimum conditions while maintaining adequate resolution.
Collapse