Scott WJ, Tauber S, Eck CD, Scott RA, Ohly J. The Clinical Relationship of Anterior Capsular Tears and Vertical Spacing in the Femtosecond Laser Capsulotomy Procedure.
J Refract Surg 2019;
35:280-284. [PMID:
31059576 DOI:
10.3928/1081597x-20190410-04]
[Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/13/2018] [Accepted: 04/09/2019] [Indexed: 01/19/2023]
Abstract
PURPOSE
To evaluate the frequency of anterior capsule tears and capsule edge irregularities in femtosecond laser anterior capsulotomy, including the effect of optimized vertical spacing settings.
METHODS
Three thousand eight hundred thirty-four cases were examined and grouped according to the femtosecond laser capsulotomy settings using manufacturer default settings with a vertical spacing of 10, 15, and 20 µm. Capsulotomy edge irregularities and anterior capsule tears were recorded intraoperatively.
RESULTS
The anterior capsule tear rates for the 10, 15, and 20 µm groups were 0.79%, 0.35%, and 0.09%, respectively. The capsule edge irregularity rates for the 10, 15, and 20 µm groups were 6.25%, 1.13%, and 1.15%, respectively. The treatment times for the 10, 15, and 20 µm groups were 1.6, 0.7, and 0.8 second, respectively. The risk reduction of anterior capsule tears positively correlated with the 15 and 20 µm group settings. The anterior capsule tear rate difference between the 10 and 20 µm group settings was statistically significant. Capsule edge irregularity rates were reduced in the 15 and 20 µm group settings. Eye movement was monitored in the 20 µm group and was observed in 9 of 13 patients with capsule edge irregularities.
CONCLUSIONS
Anterior capsule tear rates are low with femtosecond laser cataract surgery. The rate can be further reduced with optimization of the settings, including an increase in vertical spacing. [J Refract Surg. 2019;35(5):280-284.].
Collapse