Sakamoto T, Matsumoto K, Otsuka M, Nanbu H, Okumura M. [Gantry QA Using Three-dimensional Diode Array Detector on O-ring Linear Accelerator].
Nihon Hoshasen Gijutsu Gakkai Zasshi 2022;
78:263-269. [PMID:
35314535 DOI:
10.6009/jjrt.780303]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 06/14/2023]
Abstract
PURPOSE
It is necessary to perform gantry quality assurance (QA) in high precision radiotherapy. However, the O-ring type linear accelerator (Halcyon) does not have a light field and laser as a reference of isocenter point. The aim of this study is to investigate the usefulness of a three-dimensional diode array detector for gantry angle QA, and an O-ring type linear accelerator.
METHOD
The gantry angle and rotational center were verified using the ArcCHECK 3D diode array on the general linear accelerator (TrueBeam) as a reference and Halcyon. The gantry angles were measured at 0, 90, 180, and 270°. The accuracy of the gantry rotational center was evaluated using rotational irradiation in the clockwise and counterclockwise directions between 181° and 179°.
RESULTS
The QA system with ArcCHECK was able to apply on the TrueBeam and Halcyon. As a result of the accuracy of the gantry angle, the maximum error of value calculated from ArcCHECK was 0.1° compared with the nominal gantry angle of Halcyon. As a result of the accuracy of the gantry rotation isocenter of Halcyon, the distance between the isocenter and the gantry rotation center was 0.45 mm and 0.41 mm in the clockwise and counterclockwise directions, respectively.
CONCLUSION
The QA system with ArcCHECK was useful for the gantry angle and the rotation center accuracy on the Halcyon.