Knäusl B, Belotti G, Bertholet J, Daartz J, Flampouri S, Hoogeman M, Knopf AC, Lin H, Moerman A, Paganelli C, Rucinski A, Schulte R, Shimizu S, Stützer K, Zhang X, Zhang Y, Czerska K. A review of the clinical introduction of 4D particle therapy research concepts. Phys Imaging Radiat Oncol 2024; 29:100535. [PMID: 38298885 PMCID: PMC10828898 DOI: 10.1016/j.phro.2024.100535] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/11/2023] [Revised: 12/12/2023] [Accepted: 01/04/2024] [Indexed: 02/02/2024] Open
Abstract
Background and purpose Many 4D particle therapy research concepts have been recently translated into clinics, however, remaining substantial differences depend on the indication and institute-related aspects. This work aims to summarise current state-of-the-art 4D particle therapy technology and outline a roadmap for future research and developments. Material and methods This review focused on the clinical implementation of 4D approaches for imaging, treatment planning, delivery and evaluation based on the 2021 and 2022 4D Treatment Workshops for Particle Therapy as well as a review of the most recent surveys, guidelines and scientific papers dedicated to this topic. Results Available technological capabilities for motion surveillance and compensation determined the course of each 4D particle treatment. 4D motion management, delivery techniques and strategies including imaging were diverse and depended on many factors. These included aspects of motion amplitude, tumour location, as well as accelerator technology driving the necessity of centre-specific dosimetric validation. Novel methodologies for X-ray based image processing and MRI for real-time tumour tracking and motion management were shown to have a large potential for online and offline adaptation schemes compensating for potential anatomical changes over the treatment course. The latest research developments were dominated by particle imaging, artificial intelligence methods and FLASH adding another level of complexity but also opportunities in the context of 4D treatments. Conclusion This review showed that the rapid technological advances in radiation oncology together with the available intrafractional motion management and adaptive strategies paved the way towards clinical implementation.
Collapse
Affiliation(s)
- Barbara Knäusl
- Department of Radiation Oncology, Medical University of Vienna, Vienna, Austria
| | - Gabriele Belotti
- Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milano, Italy
| | - Jenny Bertholet
- Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern, Bern, Switzerland
| | - Juliane Daartz
- Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA
| | | | - Mischa Hoogeman
- Department of Medical Physics & Informatics, HollandPTC, Delft, The Netherlands
- Erasmus MC Cancer Institute, University Medical Center Rotterdam, Department of Radiotherapy, Rotterdam, The Netherlands
| | - Antje C Knopf
- Institut für Medizintechnik und Medizininformatik Hochschule für Life Sciences FHNW, Muttenz, Switzerland
| | - Haibo Lin
- New York Proton Center, New York, NY, USA
| | - Astrid Moerman
- Department of Medical Physics & Informatics, HollandPTC, Delft, The Netherlands
| | - Chiara Paganelli
- Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milano, Italy
| | - Antoni Rucinski
- Institute of Nuclear Physics Polish Academy of Sciences, PL-31342 Krakow, Poland
| | - Reinhard Schulte
- Division of Biomedical Engineering Sciences, School of Medicine, Loma Linda University
| | - Shing Shimizu
- Department of Carbon Ion Radiotherapy, Osaka University Graduate School of Medicine, Osaka, Japan
| | - Kristin Stützer
- OncoRay – National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany
- Helmholtz-Zentrum Dresden – Rossendorf, Institute of Radiooncology – OncoRay, Dresden, Germany
| | - Xiaodong Zhang
- Department of Radiation Physics, Division of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, TX, USA
| | - Ye Zhang
- Center for Proton Therapy, Paul Scherrer Institute, Villigen PSI, Switzerland
| | - Katarzyna Czerska
- Center for Proton Therapy, Paul Scherrer Institute, Villigen PSI, Switzerland
| |
Collapse
|
2
|
Fredriksson A, Glimelius L, Bokrantz R. The LET trilemma: Conflicts between robust target coverage, uniform dose, and dose-averaged LET in carbon therapy. Med Phys 2023; 50:7338-7348. [PMID: 37820319 DOI: 10.1002/mp.16771] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/16/2023] [Revised: 08/28/2023] [Accepted: 09/15/2023] [Indexed: 10/13/2023] Open
Abstract
BACKGROUND Linear energy transfer (LET) is closely related to the biological effect of ionizing radiation. Increasing the dose-averaged LET (LETd ) within the target volume has been proposed as a means to improve clinical outcome for hypoxic tumors. However, doing so can lead to reduced robustness to range uncertainty. PURPOSE To quantify the relationship between robust target coverage, target dose uniformity, and LETd , we employ robust optimization using dose-based and LETd -based functions and allow varying amounts of target non-uniformity. METHODS AND MATERIALS Robust carbon therapy optimization is used to create plans for phantom cases with increasing target sizes (radii 1, 3, and 5 cm). First, the influence of respectively range and setup uncertainty on the LETd in the target is studied. Second, we employ strategies allowing overdosage in the clinical target volume (CTV) or gross tumor volume (GTV), which enable increased LETd in the target. The relationship between robust target coverage and LETd in the target is illustrated by tradeoff curves generated by optimization using varying weights for the LETd -based functions. RESULTS As the range uncertainty used in the robust optimization increased from 0% to 5%, the near-minimum nominal LETd decreased by 17%-29% (9-21 keV/µm) for the different target sizes. The effect of increasing setup uncertainty was marginal. Allowing 10% overdosage in the CTV enabled 9%-29% (6-12 keV/µm) increased near-minimum worst case LETd for the different target sizes, compared to uniform dose plans. When 10% overdosage was allowed in the GTV only, the increase was 1%-20% (1-8 keV/µm). CONCLUSIONS There is an inherent conflict between range uncertainty robustness and high LETd in the target, which is aggravated with increasing target size. For large tumors, it is possible to simultaneously achieve two of the three qualities range robustness, uniform dose, and high LETd in the target.
Collapse
|
3
|
Ma Y, Mao J, Liu X, Dai Z, Zhang H, Li Y, Li Q. Selection of breathing phase number in 4D scanned proton treatment planning optimization for lung tumors. Phys Med 2023; 114:103152. [PMID: 37783030 DOI: 10.1016/j.ejmp.2023.103152] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/04/2023] [Revised: 09/14/2023] [Accepted: 09/22/2023] [Indexed: 10/04/2023] Open
Abstract
The standard four-dimensional (4D) treatment planning includes all breathing states in the optimization process, which is time-consuming. This work was aimed to optimize the number of intermediate phases needed for 4D proton treatment planning optimization to reduce the computational cost. Five 4D optimization strategies adopting different numbers of intermediate states and one three-dimensional (3D) optimization plan were studied for fifteen lung cancer patients treated with scanned protons, optimizing on all ten phases (4D_10), two extreme phases (4D_2), six phases during the exhalation stage (4D_6EX), six phases during the inhalation stage (4D_6IN), two extreme phases plus an intermediate state (4D_3) and average computed tomography image (3D), respectively. The 4D dose evaluation was conducted on all the ten phases, considering the interplay effect. The resulting doses accumulated on the reference phase were computed and compared. Compared to the 4D optimization plans, the 3D optimization plan performed inferiorly in target coverage, but superiorly in organ at risks (OARs) sparing. For the 4D optimization, all the five 4D plans showed similar performance in OARs protection. However, the 4D_6EX and 4D_6IN strategies out-performed the 4D_2 and 4D_3 plans in dose homogeneity. The computing times of the 4D_2, 4D_3, 4D_6EX and 4D_6IN approaches decreased to 32%, 41%, 66% and 67% of the 4D_10 method, respectively. Thus, our study suggested that the use of all phases during inhalation or exhalation stage might be a feasible approach substituting for the full phase strategy to reduce the calculation load while guaranteeing the plan quality for scanned proton therapy.
Collapse
Affiliation(s)
- Yuanyuan Ma
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China
| | - Jingfang Mao
- Department of Radiation Oncology, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai 201321, China; Shanghai Key Laboratory of Radiation Oncology, Shanghai 201321, China; Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai 201321, China
| | - Xinguo Liu
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China
| | - Zhongying Dai
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China
| | - Hui Zhang
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China
| | - Yazhou Li
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China; University of Chinese Academy of Sciences, Beijing 100049, China; Gansu Provincial Hospital, Lanzhou 730000, China
| | - Qiang Li
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China; University of Chinese Academy of Sciences, Beijing 100049, China.
| |
Collapse
|
4
|
Meschini G, Vai A, Barcellini A, Fontana G, Molinelli S, Mastella E, Pella A, Vitolo V, Imparato S, Orlandi E, Ciocca M, Baroni G, Paganelli C. Time-resolved MRI for off-line treatment robustness evaluation in carbon-ion radiotherapy of pancreatic cancer. Med Phys 2022; 49:2386-2395. [PMID: 35124811 PMCID: PMC9306947 DOI: 10.1002/mp.15510] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/15/2021] [Revised: 09/28/2021] [Accepted: 01/20/2022] [Indexed: 11/25/2022] Open
Abstract
PURPOSE In this study, we investigate the use of magnetic resonance imaging (MRI) for the clinical evaluation of gating treatment robustness in carbon-ion radiotherapy (CIRT) of pancreatic cancer. Indeed, MRI allows radiation-free repeated scans and fast dynamic sequences for time-resolved (TR) imaging (cine-MRI), providing information on inter- and intra-fraction cycle-to-cycle variations of respiratory motion. MRI can therefore support treatment planning and verification, overcoming the limitations of the current clinical standard, that is, four-dimensional computed tomography (4DCT), which describes an "average" breathing cycle neglecting breathing motion variability. METHODS We integrated a technique to generate a virtual CT (vCT) from 3D MRI with a method for 3D reconstruction from 2D cine-MRI, to produce TR vCTs for dose recalculations. For eight patients, the method allowed evaluating inter-fraction variations at end-exhale and intra-fraction cycle-to-cycle variability within the gating window in terms of tumor displacement and dose to the target and organs at risk. RESULTS The median inter-fraction tumor motion was in the range 3.33-12.16 mm, but the target coverage was robust (-0.4% median D95% variation). Concerning cycle-to-cycle variations, the gating technique was effective in limiting tumor displacement (1.35 mm median gating motion) and corresponding dose variations (-3.9% median D95% variation). The larger exposure of organs at risk (duodenum and stomach) was caused by inter-fraction motion, whereas intra-fraction cycle-to-cycle dose variations were limited. CONCLUSIONS This study proposed a method for the generation of TR vCTs from MRI, which enabled an off-line evaluation of gating treatment robustness and suggested its feasibility to support treatment planning of pancreatic tumors in CIRT.
Collapse
Affiliation(s)
- Giorgia Meschini
- Department of Electronics, Information and BioengineeringPolitecnico di MilanoMilanItaly
| | - Alessandro Vai
- Medical Physics UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Amelia Barcellini
- Clinical DepartmentNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Giulia Fontana
- Clinical Bioengineering UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Silvia Molinelli
- Medical Physics UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Edoardo Mastella
- Medical Physics UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Andrea Pella
- Clinical Bioengineering UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Viviana Vitolo
- Clinical DepartmentNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Sara Imparato
- Radiology UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Ester Orlandi
- Clinical DepartmentNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Mario Ciocca
- Medical Physics UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Guido Baroni
- Department of Electronics, Information and BioengineeringPolitecnico di MilanoMilanItaly
- Clinical Bioengineering UnitNational Center for Oncological Hadrontherapy (Fondazione CNAO)PaviaItaly
| | - Chiara Paganelli
- Department of Electronics, Information and BioengineeringPolitecnico di MilanoMilanItaly
| |
Collapse
|
5
|
Pakela JM, Knopf A, Dong L, Rucinski A, Zou W. Management of Motion and Anatomical Variations in Charged Particle Therapy: Past, Present, and Into the Future. Front Oncol 2022; 12:806153. [PMID: 35356213 PMCID: PMC8959592 DOI: 10.3389/fonc.2022.806153] [Citation(s) in RCA: 15] [Impact Index Per Article: 7.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/31/2021] [Accepted: 02/04/2022] [Indexed: 12/14/2022] Open
Abstract
The major aim of radiation therapy is to provide curative or palliative treatment to cancerous malignancies while minimizing damage to healthy tissues. Charged particle radiotherapy utilizing carbon ions or protons is uniquely suited for this task due to its ability to achieve highly conformal dose distributions around the tumor volume. For these treatment modalities, uncertainties in the localization of patient anatomy due to inter- and intra-fractional motion present a heightened risk of undesired dose delivery. A diverse range of mitigation strategies have been developed and clinically implemented in various disease sites to monitor and correct for patient motion, but much work remains. This review provides an overview of current clinical practices for inter and intra-fractional motion management in charged particle therapy, including motion control, current imaging and motion tracking modalities, as well as treatment planning and delivery techniques. We also cover progress to date on emerging technologies including particle-based radiography imaging, novel treatment delivery methods such as tumor tracking and FLASH, and artificial intelligence and discuss their potential impact towards improving or increasing the challenge of motion mitigation in charged particle therapy.
Collapse
Affiliation(s)
- Julia M Pakela
- Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, United States
| | - Antje Knopf
- Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, Netherlands.,Department I of Internal Medicine, Center for Integrated Oncology Cologne, University Hospital of Cologne, Cologne, Germany
| | - Lei Dong
- Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, United States
| | - Antoni Rucinski
- Institute of Nuclear Physics, Polish Academy of Sciences, Krakow, Poland
| | - Wei Zou
- Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, United States
| |
Collapse
|
6
|
Extension of RBE-weighted 4D particle dose calculation for non-periodic motion. Phys Med 2021; 91:62-72. [PMID: 34715550 DOI: 10.1016/j.ejmp.2021.10.009] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/08/2021] [Revised: 10/01/2021] [Accepted: 10/05/2021] [Indexed: 01/02/2023] Open
Abstract
PURPOSE Highly conformal scanned Carbon Ion Radiotherapy (CIRT) might permit dose escalation and improved local control in advanced stage thoracic tumors, but is challenged by target motion. Dose calculation algorithms typically assume a periodically repeating, regular motion. To assess the effect of realistic, irregular motion, new algorithms of validated accuracy are needed. METHODS We extended an in-house treatment planning system to calculate RBE-weighted dose distributions in CIRT on non-periodic CT image sequences. Dosimetric accuracy was validated experimentally on a moving, time-resolved ionization chamber array. Log-file based dose reconstructions were compared by gamma analysis and correlation to measurements at every intermediate detector frame during delivery. The impact of irregular motion on treatment quality was simulated on a virtual 4DCT thorax phantom. Periodic motion was compared to motion with varying amplitude and period ± baseline drift. Rescanning as a mitigation strategy was assessed on all scenarios. RESULTS In experimental validation, average gamma pass rates were 99.89+-0.30% for 3%/3 mm and 88.2+-2.2% for 2%/2 mm criteria. Average correlation for integral dose distributions was 0.990±0.002. Median correlation for single 200 ms frames was 0.947±0.006. In the simulations, irregular motion deteriorated V95 target coverage to 81.2%, 76.6% and 79.0% for regular, irregular motion and irregular motion with base-line drift, respectively. Rescanning restored V95 to >98% for both scenarios without baseline drift, but not with additional baseline drift at 83.7%. CONCLUSIONS The validated algorithm permits to study the effects of irregular motion and to develop and adapt appropriate motion mitigation techniques.
Collapse
|
7
|
He P, Li Q. Motion management with variable cycle-based respiratory guidance method for carbon-ion pencil beam scanning treatment. Phys Med 2021; 87:99-105. [PMID: 34134014 DOI: 10.1016/j.ejmp.2021.06.005] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/09/2021] [Revised: 05/09/2021] [Accepted: 06/04/2021] [Indexed: 02/08/2023] Open
Abstract
PURPOSE A novel variable cycle-based respiratory guidance method was proposed to synchronize the patterns between patients' breathing and the magnetic excitation of synchrotron under the mode of full-energy depth scanning beam delivery, in order to improve the treatment precision and efficiency for carbon ion therapy. METHODS Audio-visual biofeedback system with variable cycle-based respiratory guidance method was developed. We enrolled 6 healthy volunteers and a simulation study of the fixed cycle-based and variable cycle-based respiratory guidance with three treatment fractions was performed. A total of 72 breathing curves were collected for 4D dose calculations with three 4DCT datasets of lung tumor cases. Target dose coverage (D95: the percent dose covering 95% of the target), dose homogeneity (D5-D95), and treatment time were analyzed. The Wilcoxon signed-rank test was used for statistical difference analysis, and p < 0.05 was considered significant. RESULTS With the variable cycle-based respiratory guidance method, the breath hold phase of breathing curve could be synchronized with the synchrotron flat-top phase over time. The dose homogeneity was improved by factors of 1.94-2.92 compared to the fixed cycle-based respiratory guidance maneuvers alone or in combination with gating technique. Moreover, the treatment efficiency increased by 11-23%, depending on the duty cycle settings of the gating window. CONCLUSIONS The proposed variable cycle-based respiratory guidance method could improve both the treatment efficiency and precision under the mode of the full-energy depth scanning beam delivery for synchrotron-based carbon ion therapy.
Collapse
Affiliation(s)
- Pengbo He
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China; University of Chinese Academy of Sciences, Beijing 100049, China
| | - Qiang Li
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Heavy Ion Radiation Biology and Medicine of Chinese Academy of Sciences, Lanzhou 730000, China; Key Laboratory of Basic Research on Heavy Ion Radiation Application in Medicine, Gansu Province, Lanzhou 730000, China; University of Chinese Academy of Sciences, Beijing 100049, China.
| |
Collapse
|