Abstract
BACKGROUND
Liver shear stiffness measurement using magnetic resonance elastography (MRE) aids in the noninvasive diagnosis and staging of liver fibrosis. Inadequate breath-holds can lead to inaccurate stiffness estimation and/or failed MRE exams.
PURPOSE
To prospectively evaluate the performance of compressed sensitivity encoding (C-SENSE) accelerated rapid MRE measurement of liver shear stiffness using displacement wave polarity-inversion motion encoding.
STUDY TYPE
Retrospective.
SUBJECTS
Eleven with liver disease and 10 asymptomatic subjects.
FIELD STRENGTH/SEQUENCE
1.5 T; gradient-recalled-echo (GRE) MRE.
ASSESSMENT
All participants underwent: 1) two-dimensional (2D) GRE MRE with inflow saturation using SENSE acceleration factor (R) of 2 (standard of care [SC]); 2) 2D rapid MRE with (RwS); and 3) without (RnS) inflow saturation using C-SENSE R = 3; and 4) spatial three-dimensional (3D) rapid MRE with inflow saturation (R3D) using C-SENSE R = 4; with nominally identical spatial resolution and coverage. Image analyst (D.G., 2 years of experience) drew identical and maximal regions of interest (ROIs) in right hepatic lobe.
STATISTICAL TESTS
Linear regression, intra-class correlation coefficients (ICC), Bland-Altman analyses, and the Wilcoxon signed-rank test were used to assess consistency and agreement of liver stiffness measurements for manually drawn identical and maximal ROIs.
RESULTS
In 21 participants (37 ± 14 years) with liver stiffness (2.3 ± 0.7 kPa), body mass index (BMI 27 ± 7 kg/m2 ), proton density fat fraction (PDFF 9 ± 9%), and T2 * (27 ± 4 msec); rapid MRE sequences showed excellent agreement (ICC > 0.95) with SC MRE and no correlation (r2 < 0.1) of the differences (mean difference <0.2 kPa, <6%; limits of agreement <0.4 kPa, <16%) with BMI, PDFF, and T2 *. Breath-hold times were: 14 seconds (SC), 5 seconds (RnS), 7 seconds (RwS) per slice, and 16 seconds for the R3D acquisition.
DATA CONCLUSIONS
C-SENSE accelerated GRE MRE sequences, using displacement wave polarity-inversion motion encoding, produce equivalent measurements of liver stiffness and have potential clinical benefit in patients with limited breath-holding capacity.
LEVEL OF EVIDENCE
1 TECHNICAL EFFICACY: Stage 1.