Mai J, Erickson B, Rownd J, Gillin M. Comparision of four different dose specification methods for high-dose-rate intracavitary radiation for treatment of cervical cancer.
Int J Radiat Oncol Biol Phys 2001;
51:1131-41. [PMID:
11704338 DOI:
10.1016/s0360-3016(01)01771-0]
[Citation(s) in RCA: 26] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/18/2022]
Abstract
PURPOSE
To compare the dose delivered to target tissues and dose-limiting structures as defined by specific dose points with high-dose-rate intracavitary brachytherapy using tandem and ring or tandem and ovoids applicators, and to provide a reasonable approach to dose optimization.
METHODS AND MATERIALS
Dosimetry was obtained using four different dose specifications: (1) 100% of the dose prescribed in a tapered fashion along the tandem and 140% at the ovoid/ring surface, (2) 100% of the dose prescribed along the tandem and 100% at the ovoid/ring surface, (3) 100% of the dose prescribed to point A without any additional applicator specification points, and (4) nonoptimized plan using relative dwell weighting to simulate classic Fletcher low-dose-rate (LDR) loading with the dose specified at point A. Point doses were recorded at A, B, and T (cervical tumor point), ICRU rectum, and ovoid/ring surface.
RESULTS
For the tandem and ovoids applicators, significant differences were found among the four different dose specification methods for point T and vaginal mucosal doses. When the dose was optimized to point A alone, the ovoid dwell weights were reduced, resulting in higher point T doses and underdosing of the vaginal mucosa. Fixed weighting based on Fletcher LDR loading specifications resulted in higher vaginal mucosa doses. For the tandem and ring applicators, significant differences were observed for vaginal mucosal doses and the ICRU rectal dose. Optimization to point A alone resulted in widely varying dosimetric distributions and vaginal mucosa doses up to 632% of the prescription dose. With nonoptimized fixed weighting, the vaginal wall dose and ICRU rectal dose were increased.
CONCLUSION
Prescribing to dose optimization points in a tapered fashion along the tandem and at the ovoid/ring surface results in a pear-shaped dose distribution resembling classic LDR systems. The other dose specification methods may result in underdosing of important target tissues or overdosing of adjacent dose-limiting structures.