Heller H, Gäbler R, Schuster KD. Real‐Time Detection of Nitric Oxide Isotopes in Lung Function Tests.
Methods Enzymol 2005;
396:592-7. [PMID:
16291266 DOI:
10.1016/s0076-6879(05)96051-5]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
In lung function tests, the determination of the pulmonary diffusing capacity (D) using the single-breath method is a commonly applied technique. The calculation of D is performed on the basis of accurate measurements of indicator gas concentrations. In this chapter, we demonstrate the appropriateness of the stable nitric oxide (NO) isotopes 14NO and 15NO in revealing reliable data of D. We performed studies on animals (14NO) by using respiratory mass spectrometry (M3) and on humans (15NO) by applying laser magnetic resonance spectroscopy (LMRS). The equipment was characterized by sufficient detection limits of 70 parts/billion at [14NO] = 0.001% (M3) and 40 parts/billion at [15NO] = 0.002 % (LMRS), respectively. Lastly, we were able to show that D-values for 14NO indeed reveal the entire diffusive properties of the alveolar-capillary membrane and that 15NO is a useful indicator gas for reflecting disturbances of pulmonary gas exchange.