Rona Z, Klebermass K, Cardona F, Czaba CD, Brugger PC, Weninger M, Pollak A, Prayer D. Comparison of neonatal MRI examinations with and without an MR-compatible incubator: advantages in examination feasibility and clinical decision-making.
Eur J Paediatr Neurol 2010;
14:410-7. [PMID:
20471292 DOI:
10.1016/j.ejpn.2010.03.005]
[Citation(s) in RCA: 18] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/02/2008] [Revised: 03/02/2010] [Accepted: 03/19/2010] [Indexed: 11/19/2022]
Abstract
PURPOSE
To assess the utility of an MRI-compatible incubator (INC) by comparing.
METHODS
In a retrospective study, the clinical and radiological aspects of 129 neonatal MRI examinations during a 3 year period were analyzed. Routine protocols including fast spin-echo T2-weighted (w) sequences, axial T1w, Gradient-echo, diffusion sequences, and 3D T1 gradient-echo sequences were performed routinely, angiography and spectroscopy were added in some cases. Diffusion-tensor imaging was done in 50% of the babies examined in the INC and 26% without INC. Sequences, adapted from fetal MR-protocols were done in infants younger than 32 gestational weeks. Benefit from MR-information with respect to further management was evaluated.
RESULTS
The number of the examinations increased (30-99), while the mean age (43-38, 8 weeks of gestational age) and weight (3308-2766 g) decreased significantly with the use of the MR-compatible incubator. The mean imaging time (34, 43-30, 29 min) decreased, with a mean of one additionally performed sequence in the INC group. All infants received sedatives according to our anaesthetic protocol preceding imaging, but a repeated dose was never necessary (10% without INC) using the INC. Regarding all cases, MR-based changes in clinical management were initiated in 58%, while in 57% of cases the initial ultrasound diagnosis was changed or further specified.
CONCLUSIONS
The use of the INC enables the MR access of unstable infants with suspect CNS problems to the management, of whom is improved by MR information to significantly higher percentage, than without INC.