Popowski Y, Hiltbrand E, Joliat D, Rouzaud M. Open magnetic resonance imaging using titanium-zirconium needles: improved accuracy for interstitial brachytherapy implants?
Int J Radiat Oncol Biol Phys 2000;
47:759-65. [PMID:
10837962 DOI:
10.1016/s0360-3016(00)00488-0]
[Citation(s) in RCA: 41] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/17/2022]
Abstract
PURPOSE
To evaluate the benefit of using an open magnetic resonance (MR) machine and new MR-compatible needles to improve the accuracy of brachytherapy implants in pelvic tumors.
METHODS AND MATERIALS
The open MR machine, foreseen for interventional procedures, allows direct visualization of the pelvic structures that are to be implanted. For that purpose, we have developed MR- and CT-compatible titanium-zirconium (Ti-Zr) brachytherapy needles that allow implantations to be carried out under the magnetic field. In order to test the technical feasibility of this new approach, stainless steel (SS) and Ti-Zr needles were first compared in a tissue-equivalent phantom. In a second step, two patients implanted with Ti-Zr needles in the brachytherapy operating room were scanned in the open MR machine. In a third phase, four patients were implanted directly under open MR control.
RESULTS
The artifacts induced by both materials were significantly different, strongly favoring the Ti-Zr needles. The implantation in both first patients confirmed the excellent quality of the pictures obtained with the needles in vivo and showed suboptimal implant geometry in both patients. In the next 4 patients, the tumor could be punctured with excellent accuracy, and the adjacent structures could be easily avoided.
CONCLUSION
We conclude that open MR using MR-compatible needles is a very promising tool in brachytherapy, especially for pelvic tumors.
Collapse