Survival and performance of a dual tapered-wedge fully HA-coated press fit femoral stem.
Arch Orthop Trauma Surg 2023;
143:1651-1661. [PMID:
35239021 PMCID:
PMC8892108 DOI:
10.1007/s00402-022-04393-6]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 08/02/2021] [Accepted: 02/10/2022] [Indexed: 11/30/2022]
Abstract
INTRODUCTION
Fully hydroxyapatite-coated titanium alloy double-tapered stems have been successful implant designs for THA. The Novation Element Stem (Exactech, Gainesville, FL) is one of several of these implants following a modified Corail design philosophy. Although a great deal is known about the results of the Corail stem, little is known about the more recent Corail-inspired implants. The authors evaluated the clinical outcomes, radiographic findings and survival statistics of the collarless version of the Element Stem when used routinely in a diverse patient population.
METHODS
A retrospective review was performed for all primary THR cases from 2010 to 2018. Patient demographics and HHS/OHS/patient satisfaction scores were obtained. Radiographs were evaluated for stem subsidence, radiolucencies, and bone responses to the implant.
RESULTS
One hundred and fifty seven Element Stems were implanted with greater than 2 years of radiographic follow-up, while 105 Element Stems implanted during the study period had a survival of 5 years or more. Average HHS was 91.7 and average OHS was 44.0. 54 of 157 hips had radiolucencies, all of which included a proximal zonal radiolucency. Average subsidence was 3.33 mm; 18 hips had subsidence > 4 mm. There were 8 revisions for: aseptic loosening (4), trauma-related peri-prosthetic fractures, and early proximal-medial fractures. At 5 years, the all-cause survival rate is 92.4%, 96.2% based only on aseptic loosening and 98.1% based only on proximal medial fractures.
CONCLUSION
Clinical outcomes using the collarless Novation Element Stem are good, but early proximal medial fracture is still a factor in patients with poor-quality bone. Proximal radiolucencies progressing to aseptic loosening post-operatively are also a concern. The 11.5% rate of subsidence is comparable to other fully HA-coated collarless stems. This study provides a thorough critical analysis of outcomes and midterm survival data of this dual tapered-wedge fully HA-coated collarless stem used routinely in a diverse patient population.