Kwun JS, Lee JH, Park BE, Park JS, Kim HJ, Kim SH, Jeon KH, Cho HW, Kang SH, Lee W, Youn TJ, Chae IH, Yoon CH. Diagnostic Value of a Wearable Continuous Electrocardiogram Monitoring Device (AT-Patch) for New-Onset Atrial Fibrillation in High-Risk Patients: Prospective Cohort Study.
J Med Internet Res 2023;
25:e45760. [PMID:
37721791 PMCID:
PMC10546264 DOI:
10.2196/45760]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/16/2023] [Revised: 03/07/2023] [Accepted: 08/10/2023] [Indexed: 09/19/2023] Open
Abstract
BACKGROUND
While conventional electrocardiogram monitoring devices are useful for detecting atrial fibrillation, they have considerable drawbacks, including a short monitoring duration and invasive device implantation. The use of patch-type devices circumvents these drawbacks and has shown comparable diagnostic capability for the early detection of atrial fibrillation.
OBJECTIVE
We aimed to determine whether a patch-type device (AT-Patch) applied to patients with a high risk of new-onset atrial fibrillation defined by the congestive heart failure, hypertension, age ≥75 years, diabetes mellitus, stroke, vascular disease, age 65-74 years, sex scale (CHA2DS2-VASc) score had increased detection rates.
METHODS
In this nonrandomized multicenter prospective cohort study, we enrolled 320 adults aged ≥19 years who had never experienced atrial fibrillation and whose CHA2DS2-VASc score was ≥2. The AT-Patch was attached to each individual for 11 days, and the data were analyzed for arrhythmic events by 2 independent cardiologists.
RESULTS
Atrial fibrillation was detected by the AT-Patch in 3.4% (11/320) of patients, as diagnosed by both cardiologists. Interestingly, when participants with or without atrial fibrillation were compared, a previous history of heart failure was significantly more common in the atrial fibrillation group (n=4/11, 36.4% vs n=16/309, 5.2%, respectively; P=.003). When a CHA2DS2-VASc score ≥4 was combined with previous heart failure, the detection rate was significantly increased to 24.4%. Comparison of the recorded electrocardiogram data revealed that supraventricular and ventricular ectopic rhythms were significantly more frequent in the new-onset atrial fibrillation group compared with nonatrial fibrillation group (3.4% vs 0.4%; P=.001 and 5.2% vs 1.2%; P<.001), respectively.
CONCLUSIONS
This study detected a moderate number of new-onset atrial fibrillations in high-risk patients using the AT-Patch device. Further studies will aim to investigate the value of early detection of atrial fibrillation, particularly in patients with heart failure as a means of reducing adverse clinical outcomes of atrial fibrillation.
TRIAL REGISTRATION
ClinicalTrials.gov NCT04857268; https://classic.clinicaltrials.gov/ct2/show/NCT04857268.
Collapse