van Rijn GA, Gaurisankar ZS, Saxena R, Gibbes D, Jongman HP, Haasnoot GW, Cheng YYY, Beenakker JWM, Luyten GPM. Implantation of an iris-fixated phakic intraocular lens for the correction of hyperopia: 15-year follow-up.
J Cataract Refract Surg 2021;
47:748-758. [PMID:
33298790 DOI:
10.1097/j.jcrs.0000000000000532]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/14/2020] [Accepted: 11/18/2020] [Indexed: 11/25/2022]
Abstract
PURPOSE
To assess the predictability, efficacy, stability, and safety of implantation of an Artisan iris-fixated phakic intraocular lens (IF-pIOL) for the correction of hyperopia with a follow-up of up to 15 years.
SETTING
Leiden University Medical Center, the Netherlands.
METHODS
Patients operated by a single surgeon up to 2007 were identified, and data on refraction, corrected distance visual acuity (CDVA), uncorrected distance visual acuity, endothelial cell (EC) density, and complications were collected.
RESULTS
A total of 61 eyes (32 patients) were analysed. The mean spherical equivalent decreased from +6.43 ± 1.78 diopters (D) preimplantation to -0.22 ± 0.57 D at 1 year postimplantation and remained stable throughout follow-up. A stable CDVA with safety indices ranging from 0.91 to 1.10 and efficacy indices between 0.43 and 0.86 were observed. Follow-up time had a significant effect on EC density with an estimated annual decline of 58 cells/mm2 after IF-pIOL implantation. IF-pIOL explantation was performed in a 10 eyes (16.4%) after 8.13 ± 5.11 years. The main reason for IF-pIOL explantation was EC loss (4 eyes [6.6%]). Pigment dispersion was the most encountered complication, observed in 9 eyes (14.8%).
CONCLUSIONS
Visual and refractive results after implantation of an IF-pIOL to correct hyperopia show favorable and stable results with long-term follow-up. Lifelong monitoring of EC counts is mandatory. Pigment dispersion might be a problem in hyperopic eyes implanted with an IF-pIOL; a shallower anterior chamber depth and a convex iris configuration might be predisposing factors.