Thibault B, Ducharme A, Baranchuk A, Dubuc M, Dyrda K, Guerra PG, Macle L, Mondésert B, Rivard L, Roy D, Talajic M, Andrade J, Nitzsché R, Khairy P. Very Low Ventricular Pacing Rates Can Be Achieved Safely in a Heterogeneous Pacemaker Population and Provide Clinical Benefits: The CANadian Multi-Centre Randomised Study-Spontaneous AtrioVEntricular Conduction pReservation (CAN-SAVE R) Trial.
J Am Heart Assoc 2015;
4:e001983. [PMID:
26206737 PMCID:
PMC4608083 DOI:
10.1161/jaha.115.001983]
[Citation(s) in RCA: 8] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/25/2015] [Accepted: 06/17/2015] [Indexed: 11/16/2022]
Abstract
BACKGROUND
It is well recognized that right ventricular apical pacing can have deleterious effects on ventricular function. We performed a head-to-head comparison of the SafeR pacing algorithm versus DDD pacing with a long atrioventricular delay in a heterogeneous population of patients with dual-chamber pacemakers.
METHODS AND RESULTS
In a multicenter prospective double-blinded randomized trial conducted at 10 centers in Canada, 373 patients, age 71±11 years, with indications for dual chamber DC pacemakers were randomized 1:1 to SafeR or DDD pacing with a long atrioventricular delay (250 ms). The primary objective was twofold: (1) reduction in the proportion of ventricular paced beats at 1 year; and (2) impact on atrial fibrillation burden at 3 years, defined as the ratio between cumulative duration of mode-switches divided by follow-up time. Statistical significance of both co-primary end points was required for the trial to be considered positive. At 1 year of follow-up, the median proportion of ventricular-paced beats was 4.0% with DDD versus 0% with SafeR (P<0.001). At 3 years of follow-up, the atrial fibrillation burden was not significantly reduced with SafeR versus DDD (median 0.00%, interquartile range [0.00% to 0.23%] versus median 0.01%, interquartile range [0.00% to 0.44%], respectively, P=0.178]), despite a persistent reduction in the median proportion of ventricular-paced beats (10% with DDD compared to 0% with SafeR).
CONCLUSIONS
A ventricular-paced rate <1% was safely achieved with SafeR in a population with a wide spectrum of indications for dual-chamber pacing. However, the lower percentage of ventricular pacing did not translate into a significant reduction in atrial fibrillation burden.
CLINICAL TRIAL REGISTRATION
URL: https://www.clinicaltrials.gov/ Unique identifier: NCT01219621.