Bisbee MG, Oksuz I, VanZile MP, Cherepy NJ, Cao LR. An automated fast neutron computed tomography instrument with on-line focusing for non-destructive evaluation.
THE REVIEW OF SCIENTIFIC INSTRUMENTS 2022;
93:113702. [PMID:
36461467 DOI:
10.1063/5.0091032]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/10/2022] [Accepted: 09/30/2022] [Indexed: 06/17/2023]
Abstract
A fast neutron tomography imaging instrument has been designed, built, and tested at The Ohio State University 500 kW Research Reactor on a fast neutron beamline with a peak neutron flux ≈5.4 × 107 n·cm-2·s-1 at 1.6 MeV median neutron energy. The instrument and beamline are also configurable for thermal neutron imaging. The imaging apparatus is composed of a lens coupled, water-cooled Electron Multiplying Charge Coupled Device camera, a front-surface mirror, and a high light yield plastic Polyvinyl toluene scintillator. The instrument sits on a mobile cart. A total of 5 motion-control stages are built into the system for XYZ and rotational degrees of freedom for sample positioning; the fifth stage fine tunes the focal distance between the camera and the scintillator to achieve on-line focusing. A Python code with a user-friendly graphical user interface controls the fully automated image acquisition, not requiring user interaction, yet facilitating tracking of the image acquisition. A complete fast neutron computed tomography dataset with 360 projections requires less than 3 h, with 30 s per projection. On-line focusing is accomplished with a commercial, off-the-shelf, dielectrically actuated liquid lens. Finally, tomographic reconstructions are visualized using the Livermore Tomography Tools software package. The effective pixel size (width and height) is ≈0.1058 mm, yielding a minimum voxel size of 0.1058 × 0.1058 × 0.1058 mm3, and produces a spatial resolution of 231 μm when calculated from knife-edge measurements.
Collapse