Comparison of Six Automated Treponema-Specific Antibody Assays.
J Clin Microbiol 2015;
54:163-7. [PMID:
26560543 DOI:
10.1128/jcm.02593-15]
[Citation(s) in RCA: 15] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/25/2015] [Accepted: 11/05/2015] [Indexed: 12/19/2022] Open
Abstract
Six different Treponema (TP)-specific immunoassays were compared to the fluorescent treponemal antibody absorption (FTA-ABS) test. A total of 615 samples were tested. The overall percent agreement, analytical sensitivity, and analytical specificity of each assay compared to the FTA-ABS test were as follows: Architect Syphilis TP, 99.2%, 96.8%, and 100%; Cobas Syphilis, 99.8%, 99.4%, and 100%; ADVIA Centaur Syphilis, 99.8%, 99.4%, and 100%; HISCL Anti-TP assay kit, 99.7%, 98.7%, and 100%; Immunoticles Auto3 TP, 99.0%, 97.5%, and 99.6%; Mediace TPLA, 98.0%, 98.1%, and 98.0%. All results that were discrepant between the TP-specific assays were associated with samples from noninfectious cases (11 immunoassay false positives and 7 from previous syphilis cases). Our study demonstrated that TP-specific immunoassays generally showed high sensitivities, specificities, and percentages of agreement compared to FTA-ABS, with rare cases of false-positive or false-negative results. Therefore, most TP-specific immunoassays are acceptable for use in screening for syphilis. However, it is important to perform a thorough review of a patient's clinical and treatment history for interpreting the results of syphilis serology.