Prospective validation of standardized, 3-dimensional, quantitative coronary computed tomographic plaque measurements using radiofrequency backscatter intravascular ultrasound as reference standard in intermediate coronary arterial lesions: results from the ATLANTA (assessment of tissue characteristics, lesion morphology, and hemodynamics by angiography with fractional flow reserve, intravascular ultrasound and virtual histology, and noninvasive computed tomography in atherosclerotic plaques) I study.
JACC Cardiovasc Interv 2011;
4:198-208. [PMID:
21349459 DOI:
10.1016/j.jcin.2010.10.008]
[Citation(s) in RCA: 148] [Impact Index Per Article: 11.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/21/2010] [Revised: 10/05/2010] [Accepted: 10/20/2010] [Indexed: 12/17/2022]
Abstract
OBJECTIVES
This study sought to determine the accuracy of 3-dimensional, quantitative measurements of coronary plaque by computed tomography angiography (CTA) against intravascular ultrasound with radiofrequency backscatter analysis (IVUS/VH).
BACKGROUND
Quantitative, 3-dimensional coronary CTA plaque measurements have not been validated against IVUS/VH.
METHODS
Sixty patients in a prospective study underwent coronary X-ray angiography, IVUS/VH, and coronary CTA. Plaque geometry and composition was quantified after spatial coregistration on segmental and slice-by-slice bases. Correlation, mean difference, and limits of agreement were determined.
RESULTS
There was significant correlation for all pre-specified parameters by segmental and slice-by-slice analyses (r = 0.41 to 0.84; all p < 0.001). On a segmental basis, CTA underestimated minimal lumen diameter by 21% and overestimated diameter stenosis by 39%. Minimal lumen area was overestimated on CTA by 27% but area stenosis was only underestimated by 5%. Mean difference in noncalcified plaque volume and percent and calcified plaque volume and percent were 38%, -22%, 104%, and 64%. On a slice-by-slice basis, lumen, vessel, noncalcified-, and calcified-plaque areas were overestimated on CTA by 22%, 19%, 44%, and 88%. There was significant correlation for percentage of atheroma volume (0.52 vs. 0.54; r = 0.51; p < 0.001). Compositional analysis suggested that high-density noncalcified plaque on CTA best correlated with fibrous tissue and low-density noncalcified plaque correlated with necrotic core plus fibrofatty tissue by IVUS/VH.
CONCLUSIONS
This is the first validation that standardized, 3-dimensional, quantitative measurements of coronary plaque correlate with IVUS/VH. Mean differences are small, whereas limits of agreement are wide. Low-density noncalcified plaque correlates with necrotic core plus fibrofatty tissue on IVUS/VH.