Chung HW, Ko SM, Hwang HK, So Y, Yi JG, Lee EJ. Diagnostic Performance of Coronary CT Angiography, Stress Dual-Energy CT Perfusion, and Stress Perfusion Single-Photon Emission Computed Tomography for Coronary Artery Disease: Comparison with Combined Invasive Coronary Angiography and Stress Perfusion Cardiac MRI.
Korean J Radiol 2017;
18:476-486. [PMID:
28458600 PMCID:
PMC5390617 DOI:
10.3348/kjr.2017.18.3.476]
[Citation(s) in RCA: 10] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/23/2016] [Accepted: 09/07/2016] [Indexed: 12/24/2022] Open
Abstract
Objective
To investigate the diagnostic performance of coronary computed tomography angiography (CCTA), stress dual-energy computed tomography perfusion (DE-CTP), stress perfusion single-photon emission computed tomography (SPECT), and the combinations of CCTA with myocardial perfusion imaging (CCTA + DE-CTP and CCTA + SPECT) for identifying coronary artery stenosis that causes myocardial hypoperfusion. Combined invasive coronary angiography (ICA) and stress perfusion cardiac magnetic resonance (SP-CMR) imaging are used as the reference standard.
Materials and Methods
We retrospectively reviewed the records of 25 patients with suspected coronary artery disease, who underwent CCTA, DE-CTP, SPECT, SP-CMR, and ICA. The reference standard was defined as ≥ 50% stenosis by ICA, with a corresponding myocardial hypoperfusion on SP-CMR.
Results
For per-vascular territory analysis, the sensitivities of CCTA, DE-CTP, SPECT, CCTA + DE-CTP, and CCTA + SPECT were 96, 96, 68, 93, and 68%, respectively, and specificities were 72, 75, 89, 85, and 94%, respectively. The areas under the receiver operating characteristic curve (AUCs) were 0.84 ± 0.05, 0.85 ± 0.05, 0.79 ± 0.06, 0.89 ± 0.04, and 0.81 ± 0.06, respectively. For per-patient analysis, the sensitivities of CCTA, DE-CTP, SPECT, CCTA + DE-CTP, and CCTA + SPECT were 100, 100, 89, 100, and 83%, respectively; the specificities were 14, 43, 57, 43, and 57%, respectively; and the AUCs were 0.57 ± 0.13, 0.71 ± 0.11, 0.73 ± 0.11, 0.71 ± 0.11, and 0.70 ± 0.11, respectively.
Conclusion
The combination of CCTA and DE-CTP enhances specificity without a loss of sensitivity for detecting hemodynamically significant coronary artery stenosis, as defined by combined ICA and SP-CMR.
Collapse