Abstract
The auditory brain stem responses (ABRs) to 500- and 2,000-Hz bone-conducted (BC) tones were recorded from 48 infants with ears exhibiting various external and middle ear states (normal, otitis media, auditory meatal atresia). Amplitudes were greater, wave V latencies longer, and detectability better for responses to 500-Hz BC tones compared to 2,000-Hz BC tones. Overall, most (94% to 100%) infants with normal cochlear sensitivity demonstrate ABRs to 20-dB normal hearing level (nHL) 500-Hz BC tones and 30-dB nHL 2,000-Hz BC tones. In cases in which masking is difficult (eg, bilateral atresia), infant ipsilateral/contralateral ABR asymmetries may help determine from which cochlea a response to the BC tones originates. In conclusion, two-channel ABR recordings to BC tones appear to be feasible for demonstrating normal cochlear sensitivity in infants.