Electrochemical analysis of 8-hydroxy-2'-deoxyguanosine with enhanced sensitivity based on exonuclease-mediated functional nucleic acid.
Talanta 2019;
199:324-328. [PMID:
30952266 DOI:
10.1016/j.talanta.2019.02.080]
[Citation(s) in RCA: 9] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/22/2018] [Revised: 02/16/2019] [Accepted: 02/21/2019] [Indexed: 01/27/2023]
Abstract
In this work, an electrochemical method for sensitive analysis of 8-hydroxy-2'-deoxyguanosine, a key biomarker that is widely used to study oxidative injury-related diseases, is proposed based on exonuclease-mediated functional nucleic acid. In the design, exonuclease can not only distinguish the existence of target, but also suppress the background noise, thus the sensitivity can be enhanced. Moreover, DNAzyme designed in the functional nucleic acid can further improve the sensitivity of the analysis during signal generation process. Therefore, exonuclease-mediated functional nucleic acid may ensure high sensitivity of the assay. Further studies reveal that the detection of 8-hydroxy-2'-deoxyguanosine can be achieved with a linearity from 0.01 nM to 7.0 μM and a detection limit of 6.82 pM. The new method has also been successfully applied to the determination of 8-OHdG in urine with good results, indicating its great potential for practical use.