Ito S, Kinoshita K, Endo A, Kami R, Kotake Y, Nakamura M. Inter- and intra-core laboratory variability in the quantitative coronary angiography analysis for drug-eluting stent treatment and follow up.
Ther Adv Cardiovasc Dis 2020;
14:1753944720958982. [PMID:
32993464 PMCID:
PMC7534069 DOI:
10.1177/1753944720958982]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/06/2019] [Accepted: 08/19/2020] [Indexed: 11/16/2022] Open
Abstract
AIM
To evaluate inter-core laboratory variability of quantitative coronary angiography (QCA) parameters in comparison with intra-core laboratory variability in a randomized controlled trial evaluating drug-eluting stents.
METHODS
A total of 50 patients with 62 coronary lesions were analyzed by four analysis experts belonging to an Angiographic Core Laboratory (ACL: 1 expert) and a Cardiovascular Imaging Core Laboratory (CICL: 3 experts). QCA was based on the same standard operating procedure, but selections of projection and cine frames were at the discretion of each analyst. Inter- and intra-core laboratory variabilities were evaluated by accuracy, precision, Bland Altman analysis, and coefficient of variation.
RESULTS
Pre-MLD (minimal lumen diameter) was significantly smaller in results from ACL than those from all CICL experts. Number of analyzed projections did not affect pre-MLD results. Acute gain was larger in ACL than in CICL2. No significant difference was observed in late loss and loss index between inter-core laboratories. Agreement between core labs in the Bland-Altman analysis for each QCA parameter was as follows (mean difference, 95% limits of agreement): pre-MLD (-0.32, -0.74 to 0.10), stent MLD (0.08, -0.28 to 0.44), acute gain (0.22, -0.44 to 0.88), and late loss (-0.07, -0.69 to 0.55). Agreement between analysts in CICL (mean difference, 95% limits of agreement) was: pre MLD (-0.03, -0.37 to 0.31), stent MLD (0.15, -0.15 to 0.45), acute gain (0.05, -0.45 to 0.55), and late loss (0.04, -0.52 to 0.60). The widest limits of agreement among three analyses were shown in both analyses. Width of limited agreement in the intra-core laboratory analysis tended to be smaller than the inter-core laboratory analysis with these parameters. Coefficient of variation tended to be larger in lesion length (LL), acute gain, late loss, and loss index in inter- and in intra- core laboratory comparisons.
CONCLUSION
Inter-core laboratory QCA variability in late loss and loss index analysis could be similar to intra-core laboratory variability, but more strict alignment between core laboratories would be necessary for initial procedural data analysis.