Gim Y, Jun RM, Han KE. Agreement between Scheimpflug Camera and the Swept-source Optical Coherence Tomography Measurements in Keratometry and Higher-order Aberrations.
KOREAN JOURNAL OF OPHTHALMOLOGY 2021;
35:337-348. [PMID:
34237207 PMCID:
PMC8521334 DOI:
10.3341/kjo.2021.0076]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/14/2021] [Accepted: 06/30/2021] [Indexed: 11/23/2022] Open
Abstract
PURPOSE
To evaluate the compatibility of corneal curvature and astigmatism, and higher-order aberrations (HOAs) measured by the Scheimpflug camera Pentacam HR and the swept-source optical coherence tomography ANTERION.
METHODS
This prospective study included normal subjects with no ophthalmic history. Steep keratometry (K), flat K, astigmatism and its axis of the anterior and posterior surfaces, total corneal power, and HOAs using the two instruments were compared. To compare the mean values of the measurements, a paired t-test was used. Bland-Altman analysis was applied to assess the agreement between the two devices.
RESULTS
Fifty-three eyes of 53 subjects were evaluated. There were statistically significant differences for steep K, astigmatism, and vector J0, J45 in the anterior surface and total corneal power between the two devices (p < 0.05). There were also significant differences in the most of the keratometric values of the posterior corneal surface (p < 0.05) except J0 (p = 0.410). Both devices showed strong positive correlations in steep K, flat K, astigmatism (r > 0.81, p < 0.001) with wide ranges of a 95% limit of agreement. Vectoral components were significantly correlated (r > 0.78, p < 0.001) with narrow 95% limit of agreement, except J45 of the posterior surface (r = 0.39, p = 0.004). In the corneal HOAs, there were statistically significant differences in the vertical coma, horizontal trefoil, spherical aberration, and root mean square of each fifth- and sixth-order Zernike coefficient (p = 0.043, p = 0.041, p < 0.001, p < 0.001, and p < 0.001, respectively). Other HOAs showed moderate to strong positive correlations (r > 0.37, p < 0.05). Most HOAs, except for the horizontal trefoil, showed clinically acceptable agreements. The total root mean square of HOAs was not significantly different between the two devices (p = 0.122).
CONCLUSIONS
Most of the keratometric values cannot be used interchangeably. However, the vectoral component of astigmatism showed clinically good agreement. Several HOAs have statistically significant differences; however, almost all HOAs showed acceptable agreements, except for the horizontal trefoil.
Collapse