Nicola M, Soliman YMA, Hussein R, Saeed H, Abdelrahim M. Comparison between traditional and nontraditional add-on devices used with pressurised metered-dose inhalers.
ERJ Open Res 2020;
6:00073-2020. [PMID:
33083443 PMCID:
PMC7553115 DOI:
10.1183/23120541.00073-2020]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/17/2020] [Accepted: 06/02/2020] [Indexed: 11/28/2022] Open
Abstract
Add-on devices that are attached to metered-dose inhalers (MDIs) were introduced to improve aerosol delivery. The objective of this study was to determine the efficacy of drug delivery from an MDI when attached to different add-on devices at different inhalation volumes.
The total emitted dose (TED) of salbutamol was estimated for the MDI alone and the MDI connected to five different add-on devices (Able valved holding chamber, Tips-haler valved holding chamber, Aerochamber plus flow Vu valved holding chamber, Dolphin chamber, and a handmade water bottle spacer), at inhalation flow of 28.3 L·min−1 with flow volume of 1, 2 and 4 L, assuming young child (aged <6 years), old child (>6 years) and adult inhalation volumes, respectively.
The TED% ranged between 84.1% and 87.2% at all inhalation volumes from the MDI alone, which was significantly greater than all MDI add-on device combinations (p<0.05). The TED% delivered to MDI sampling apparatus by a homemade water bottle spacer and Dolphin chamber, as non-antistatic add-on devices, ranged between 30.5% and 35.3%. However, washing these non-antistatic add-on devices with a light detergent before use improved their TED to range between 47.6% and 51.2%. Non-antistatic add-on devices had significantly lower TED (p<0.05) than that delivered by most antistatic add-on devices, which ranged from 51.3% to 71.6%.
This study suggests that antistatic add-on devices delivered much more aerosol than non-antistatic add-on devices. However, it may be advised to still use a non-antistatic add-on device, for the sake of solving the coordination problem, and wash it with light detergent before use to improve TED.
Antistatic add-on devices deliver much more aerosol from MDIs than non-antistatic add-on deviceshttps://bit.ly/3gSE5aR