Abstract
BACKGROUND
An absolute bioavailability study that utilized an intravenous [(14)C]microdose was conducted for saxagliptin (Onglyza(®)), a marketed drug product for the treatment of Type 2 diabetes mellitus. Concentrations of [(14)C]saxagliptin were determined by accelerator MS (AMS) after protein precipitation, chromatographic separation by UPLC and analyte fraction collection. A series of investigative experiments were conducted to maximize the release of the drug from high-affinity receptors and nonspecific adsorption, and to determine a suitable quantitation range.
RESULTS
A technique-appropriate validation demonstrated the accuracy, precision, specificity, stability and recovery of the AMS methodology across the concentration range of 0.025 to 15.0 dpm/ml (disintegration per minute per milliliter), the equivalent of 1.91-1144 pg/ml. Based on the study sample analysis, the mean absolute bioavailability of saxagliptin was 50% in the eight subjects with a CV of 6.6%. Incurred sample reanalysis data fell well within acceptable limits.
CONCLUSION
This study demonstrated that the optimized sample pretreatment and chromatographic separation procedures were critical for the successful implementation of an UPLC plus AMS method for [(14)C]saxagliptin. The use of multiple-point standards are useful, particularly during method development and validation, to evaluate and correct for concentration-dependent recovery, if observed, and to monitor and control process loss and operational variations.