Chen X, Xing S, Zhu Z, Wang H, Yu Z, Bai X, Li X. Accuracy of the Horizontal Calibrator in Correcting Leg Length and Restoring Femoral Offset in Total Hip Arthroplasty.
Front Surg 2022;
9:845364. [PMID:
35310443 PMCID:
PMC8927056 DOI:
10.3389/fsurg.2022.845364]
[Citation(s) in RCA: 2] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/29/2021] [Accepted: 01/28/2022] [Indexed: 11/23/2022] Open
Abstract
Background
Limb length discrepancy (LLD) is one of the most common postoperative complications and can cause serious consequences. Poor recovery of femoral offset (OD) will result in weakness of the patient's external rotator muscles and affect the patient's postoperative function. The study is aimed to present a simple approach that compensates for the shortcomings of previous measuring devices and combines the advantages of different measuring devices to provide more accurate limb length and femoral offset restoration in total hip arthroplasty (THA).
Methods
This study was a prospective controlled trial involving 89 patients with THA. Group I (n = 44) was used for intraoperative measurement of THA with our self-designed horizontal calibrator. Group II (n = 45) was measured by a traditional freehand technique. The main outcome indicators were measured on the Neusoft PACS, including LLD, femoral offset deviation, and operative time. IBM SPSS 23.0 was used for data analysis.
Results
The independent sample t-test was performed for all the data. The operative time, preoperative radiographic LLD, and OD of Group I and Group II had no statistical significance. Postoperative LLD of Group I and Group II were 2.5 ± 2.1 mm (range −5.7 to 8.3 mm) and 6.2 ± 4.3 mm (range −18.0 to 15.2 mm), and the independent sample t-test data of both (P < 0.001; 95% CI = −5.1, −2.2) showed statistical significance. In Group I, there were 38 THAs with LLD <5 mm, accounting for 86% and there were 44 THAs with LLD <10 mm, accounting for 100%. In Group II, there were 20 THAs with LLD <5 mm, accounting for 44%. There were 36 THAs with LLD <10 mm, covering for 80%. There was no significant difference in postoperative femoral offset and OD.
Conclusion
The horizontal calibrator can provide more accurate limb length and femoral offset recovery in THA. It is a simple surgical technique that does not add additionally surgical costs and does not significantly increase operative time, providing a new solution for surgeons to resolve postoperative LLD and restore femoral offset.