Karpinski K, Akguen D, Gebauer H, Paksoy A, Lupetti M, Markova V, Zettinig O, Moroder P. Reliability of Manual Measurements Versus Semiautomated Software for Glenoid Bone Loss Quantification in Patients With Anterior Shoulder Instability.
Orthop J Sports Med 2024;
12:23259671231222938. [PMID:
38352173 PMCID:
PMC10863483 DOI:
10.1177/23259671231222938]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/30/2023] [Accepted: 08/10/2023] [Indexed: 02/16/2024] Open
Abstract
Background
The presence of glenoid bone defects is indicative in the choice of treatment for patients with anterior shoulder instability. In contrast to traditional linear- and area-based measurements, techniques such as the consideration of glenoid concavity have been proposed and validated.
Purpose
To compare the reliability of linear (1-dimensional [1D]), area (2-dimensional [2D]), and concavity (3-dimensional [3D]) measurements to quantify glenoid bone loss performed manually and to analyze how automated measurements affect reliability.
Study Design
Cohort study (diagnosis); Level of evidence, 3.
Methods
Computed tomography images of 100 patients treated for anterior shoulder instability with differently sized glenoid defects were evaluated independently by 2 orthopaedic surgeons manually using conventional software (OsiriX; Pixmeo) as well as automatically with a dedicated prototype software program (ImFusion Suite; ImFusion). Parameters obtained included 1D (defect diameter, best-fit circle diameter), 2D (defect area, best-fit circle area), and 3D (bony shoulder stability ratio) measurements. Mean values and reliability as expressed by the intraclass correlation coefficient [ICC]) were compared between the manual and automated measurements.
Results
When manually obtained, the measurements showed almost perfect agreement for 1D parameters (ICC = 0.83), substantial agreement for 2D parameters (ICC = 0.79), and moderate agreement for the 3D parameter (ICC = 0.48). When measurements were aided by automated software, the agreement between raters was almost perfect for all parameters (ICC = 0.90 for 1D, 2D, and 3D). There was a significant difference in mean values between manually versus automatically obtained measurements for 1D, 2D, and 3D parameters (P < .001 for all).
Conclusion
While more advanced measurement techniques that take glenoid concavity into account are more accurate in determining the biomechanical relevance of glenoid bone loss, our study showed that the reliability of manually performed, more complex measurements was moderate.
Collapse