Kasirajan K, Kwolek CJ, Gupta N, Fairman RM. Incidence of and outcomes after misaligned deployment of the talent thoracic stent graft system.
J Vasc Surg 2010;
51:1096-101. [PMID:
20299179 DOI:
10.1016/j.jvs.2009.11.071]
[Citation(s) in RCA: 10] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/01/2009] [Revised: 10/29/2009] [Accepted: 11/02/2009] [Indexed: 11/26/2022]
Abstract
OBJECTIVE
Various types of device-specific adverse events can occur during deployment of thoracic stent grafts due to the high flow rate and severe aortic angulation that is often encountered in the thoracic aorta. This study assessed the incidence, etiology, and overall effect of misaligned deployment of the Talent Thoracic Stent Graft (TSG) System. Techniques to predict and avoid this complication are discussed.
METHODS
Data collection included pivotal-trial follow-up, direct surveys of centers inside and outside the United States and principal investigators, a targeted literature search, and review of complaint files. Misaligned deployment was considered to occur when the proximal covered or uncovered stent apices of a thoracic stent graft folded back on itself and remained nonparallel to the wall of the aorta after deployment had been completed.
RESULTS
Of about 20,305 deployments to date of the Talent TSG, 24 misaligned deployments were identified for an incidence of 0.1%. Nineteen (79%) events occurred during treatment of degenerative aneurysms or penetrating ulcers, four (17%) during treatment of dissections, and the underlying pathology could not be determined for one patient. The misalignment was noted at the proximal end of the stent graft in 15 patients (63%), and the other 9 events (37%) occurred at the graft overlap junction. Two events were treated intraoperatively, with a second overlapping device placed in one patient and a snare used to reposition the proximal stent in another. Adverse clinical events occurred in three patients and included a persistent type I endoleak, continued false lumen perfusion in a patient with dissection, and delayed retrograde type A dissection in a patient undergoing total arch repair. No intraoperative contrast extravasation or computed tomography evidence of perforation was noted. There were no perioperative deaths or cerebrovascular events, with one report of paraplegia among the 24 patients in this series.
CONCLUSION
Misaligned deployment is an unusual phenomenon that tends to occur in the context of certain well-defined anatomic conditions in the thoracic aorta. To date, most of these events have not led to significant adverse sequelae. However, careful patient selection, periprocedural imaging, and case planning can help to identify anatomies in which misaligned opening is likely to occur, allowing physicians to avoid this complication.