1
|
Kaplińska-Kłosiewicz PM, Fura Ł, Kujawska T, Andrzejewski K, Kaczyńska K, Strzemecki D, Sulejczak M, Chrapusta SJ, Macias M, Sulejczak D. Study of Biological Effects Induced in Solid Tumors by Shortened-Duration Thermal Ablation Using High-Intensity Focused Ultrasound. Cancers (Basel) 2024; 16:2846. [PMID: 39199617 PMCID: PMC11352750 DOI: 10.3390/cancers16162846] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/21/2024] [Revised: 08/10/2024] [Accepted: 08/12/2024] [Indexed: 09/01/2024] Open
Abstract
The HIFU ablation technique is limited by the long duration of the procedure, which results from the large difference between the size of the HIFU beam's focus and the tumor size. Ablation of large tumors requires treating them with a sequence of single HIFU beams, with a specific time interval in-between. The aim of this study was to evaluate the biological effects induced in a malignant solid tumor of the rat mammary gland, implanted in adult Wistar rats, during HIFU treatment according to a new ablation plan which allowed researchers to significantly shorten the duration of the procedure. We used a custom, automated, ultrasound imaging-guided HIFU ablation device. Tumors with a 1 mm thickness margin of healthy tissue were subjected to HIFU. Three days later, the animals were sacrificed, and the HIFU-treated tissues were harvested. The biological effects were studied, employing morphological, histological, immunohistochemical, and ultrastructural techniques. Massive cell death, hemorrhages, tissue loss, influx of immune cells, and induction of pro-inflammatory cytokines were observed in the HIFU-treated tumors. No damage to healthy tissues was observed in the area surrounding the safety margin. These results confirmed the efficacy of the proposed shortened duration of the HIFU ablation procedure and its potential for the treatment of solid tumors.
Collapse
Affiliation(s)
- Patrycja Maria Kaplińska-Kłosiewicz
- Department of Experimental Pharmacology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (P.M.K.-K.); (S.J.C.)
| | - Łukasz Fura
- Department of Ultrasound, Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawinskiego 5b St., 02-106 Warsaw, Poland; (Ł.F.); (T.K.)
| | - Tamara Kujawska
- Department of Ultrasound, Institute of Fundamental Technological Research, Polish Academy of Sciences, Pawinskiego 5b St., 02-106 Warsaw, Poland; (Ł.F.); (T.K.)
| | - Kryspin Andrzejewski
- Department of Respiration Physiology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (K.A.); (K.K.)
| | - Katarzyna Kaczyńska
- Department of Respiration Physiology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (K.A.); (K.K.)
| | - Damian Strzemecki
- Department of Experimental Pharmacology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (P.M.K.-K.); (S.J.C.)
| | - Mikołaj Sulejczak
- Department of Animal Physiology, Faculty of Biology, University of Warsaw, I. Miecznikowa 1 St., 02-096 Warsaw, Poland;
| | - Stanisław J. Chrapusta
- Department of Experimental Pharmacology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (P.M.K.-K.); (S.J.C.)
| | - Matylda Macias
- Laboratory of Molecular and Cellular Neurobiology, International Institute of Molecular and Cell Biology, Ks. Trojdena 4 St., 02-109 Warsaw, Poland;
| | - Dorota Sulejczak
- Department of Experimental Pharmacology, Mossakowski Medical Research Institute, Polish Academy of Sciences, Pawinskiego 5 St., 02-106 Warsaw, Poland; (P.M.K.-K.); (S.J.C.)
| |
Collapse
|
2
|
Boregowda G, Mariappan P. Effect of high blood flow on heat distribution and ablation zone during microwave ablation-numerical approach. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING 2024; 40:e3835. [PMID: 38800993 DOI: 10.1002/cnm.3835] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/16/2024] [Revised: 04/29/2024] [Accepted: 05/04/2024] [Indexed: 05/29/2024]
Abstract
Microwave ablation has become a viable alternative for cancer treatment for patients who cannot undergo surgery. During this procedure, a single-slot coaxial antenna is employed to effectively deliver microwave energy to the targeted tissue. The success of the treatment was measured by the amount of ablation zone created during the ablation procedure. The significantly large blood vessel placed near the antenna causes heat dissipation by convection around the blood vessel. The heat sink effect could result in insufficient ablation, raising the risk of local tumor recurrence. In this study, we investigated the heat loss due to large blood vessels and the relationship between blood velocity and temperature distribution. The hepatic artery, with a diameter of 4 mm and a height of 50 mm and two branches, is considered in the computational domain. The temperature profile, localized tissue contraction, and ablation zones were simulated for initial blood velocities 0.05, 0.1, and 0.16 m/s using the 3D Pennes bio-heat equation, temperature-time dependent model, and cell death model, respectively. Temperature-dependent blood velocity is modeled using the Navier-Stokes equation, and the fluid-solid interaction boundary is treated as a convective boundary. For discretization, we utilized H curl Ω elements for the wave propagation model,H 1 Ω elements for the Pennes bio-heat model, andH 1 Ω 3 × L 0 2 Ω elements for the Navier-Stokes equation, where Ω represents the computational domain. The simulated results show that blood vessels and blood velocity have a significant impact on temperature distribution, tissue contraction, and the volume of the ablation zone.
Collapse
Affiliation(s)
- Gangadhara Boregowda
- Department of Mathematics and Statistics, Indian Institute of Technology Tirupati, Tirupati, India
| | - Panchatcharam Mariappan
- Department of Mathematics and Statistics, Indian Institute of Technology Tirupati, Tirupati, India
| |
Collapse
|
3
|
Heshmat A, O’Connor CS, Albuquerque Marques Silva J, Paolucci I, Jones AK, Odisio BC, Brock KK. Using Patient-Specific 3D Modeling and Simulations to Optimize Microwave Ablation Therapy for Liver Cancer. Cancers (Basel) 2024; 16:2095. [PMID: 38893214 PMCID: PMC11171243 DOI: 10.3390/cancers16112095] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/25/2024] [Revised: 05/25/2024] [Accepted: 05/29/2024] [Indexed: 06/21/2024] Open
Abstract
Microwave ablation (MWA) of liver tumors presents challenges like under- and over-ablation, potentially leading to inadequate tumor destruction and damage to healthy tissue. This study aims to develop personalized three-dimensional (3D) models to simulate MWA for liver tumors, incorporating patient-specific characteristics. The primary objective is to validate the predicted ablation zones compared to clinical outcomes, offering insights into MWA before therapy to facilitate accurate treatment planning. Contrast-enhanced CT images from three patients were used to create 3D models. The simulations used coupled electromagnetic wave propagation and bioheat transfer to estimate the temperature distribution, predicting tumor destruction and ablation margins. The findings indicate that prolonged ablation does not significantly improve tumor destruction once an adequate margin is achieved, although it increases tissue damage. There was a substantial overlap between the clinical ablation zones and the predicted ablation zones. For patient 1, the Dice score was 0.73, indicating high accuracy, with a sensitivity of 0.72 and a specificity of 0.76. For patient 2, the Dice score was 0.86, with a sensitivity of 0.79 and a specificity of 0.96. For patient 3, the Dice score was 0.8, with a sensitivity of 0.85 and a specificity of 0.74. Patient-specific 3D models demonstrate potential in accurately predicting ablation zones and optimizing MWA treatment strategies.
Collapse
Affiliation(s)
- Amirreza Heshmat
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (C.S.O.); (A.K.J.); (K.K.B.)
| | - Caleb S. O’Connor
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (C.S.O.); (A.K.J.); (K.K.B.)
| | - Jessica Albuquerque Marques Silva
- Department of Interventional Radiology, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (J.A.M.S.); (I.P.); (B.C.O.)
| | - Iwan Paolucci
- Department of Interventional Radiology, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (J.A.M.S.); (I.P.); (B.C.O.)
| | - Aaron Kyle Jones
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (C.S.O.); (A.K.J.); (K.K.B.)
| | - Bruno C. Odisio
- Department of Interventional Radiology, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (J.A.M.S.); (I.P.); (B.C.O.)
| | - Kristy K. Brock
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, USA; (C.S.O.); (A.K.J.); (K.K.B.)
| |
Collapse
|
4
|
Tehrani MHH, Moradi Kashkooli F, Soltani M. Spatiotemporal modeling of nano-delivered chemotherapeutics for synergistic microwave ablation cancer therapy. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE 2024; 247:108102. [PMID: 38447317 DOI: 10.1016/j.cmpb.2024.108102] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/16/2023] [Revised: 02/25/2024] [Accepted: 02/25/2024] [Indexed: 03/08/2024]
Abstract
BACKGROUND AND OBJECTIVE The effectiveness of current microwave ablation (MWA) therapies is limited. Administration of thermosensitive liposomes (TSLs) which release drugs in response to heat has presented a significant potential for enhancing the efficacy of thermal ablation treatment, and the benefits of targeted drug delivery. However, a complete knowledge of the mechanobiological processes underlying the drug release process, especially the intravascular drug release mechanism and its distribution in response to MWA needs to be improved. Multiscale computational-based modeling frameworks, integrating different biophysical phenomena, have recently emerged as promising tools to decipher the mechanobiological events in combo therapies. The present study aims to develop a novel multiscale computational model of TSLs delivery following MWA implantation. METHODS Due to the complex interplay between the heating procedure and the drug concentration maps, a computational model is developed to determine the intravascular release of doxorubicin from TSL, its transvascular transport into the interstitium, transport in the interstitium, and cell uptake. Computational models can estimate the interplays among liposome and drug properties, tumor perfusion, and heating regimen to examine the impact of essential parameters and to optimize a targeted drug delivery platform. RESULTS Results indicated that the synergy of TSLs with MWA allows more localized drug delivery with lower side effects. The drug release rate and tumor permeability play crucial roles in the efficacy of TSLs during MWA treatment. The computational model predicted an unencapsulated drug lime around the ablated zone, which can destroy more cancer cells compared to MWA alone by 40%. Administration of TSLs with a high release rate capacity can improve the percentage of killed cancer cells by 24%. Since the heating duration in MWA is less than 15 min, the presented combination therapy showed better performance for highly permeable tumors. CONCLUSION This study highlights the potential of the proposed computational framework to address complex and realistic scenarios in cancer treatment, which can serve as the future research foundation, including advancements in nanomedicine and optimizing the pair of TSL and MWA for both preclinical and clinical studies. The present model could be as a valuable tool for patient-specific calibration of essential parameters.
Collapse
Affiliation(s)
- Masoud H H Tehrani
- Department of Mechanical Engineering, K. N. Toosi University of Technology, Tehran Iran
| | | | - M Soltani
- Department of Mechanical Engineering, K. N. Toosi University of Technology, Tehran Iran; Department of Electrical and Computer Engineering, University of Waterloo, Waterloo, Canada; Centre for Biotechnology and Bioengineering, University of Waterloo, Waterloo, ON, Canada.
| |
Collapse
|
5
|
Mathy RM, Giannakis A, Franke M, Winiger A, Kauczor HU, Chang DH. Factors Impacting Microwave Ablation Zone Sizes: A Retrospective Analysis. Cancers (Basel) 2024; 16:1279. [PMID: 38610957 PMCID: PMC11011160 DOI: 10.3390/cancers16071279] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/02/2024] [Revised: 03/01/2024] [Accepted: 03/22/2024] [Indexed: 04/14/2024] Open
Abstract
PURPOSE Evaluation of the influence of intrinsic and extrinsic conditions on ablation zone volumes (AZV) after microwave ablation (MWA). METHODS Retrospective analysis of 38 MWAs of therapy-naïve liver tumours performed with the NeuWave PR probe. Ablations were performed either in the 'standard mode' (65 W, 10 min) or in the 'surgical mode' (95 W, 1 min, then 65 W, 10 min). AZV measurements were obtained from contrast-enhanced computed tomography immediately post-ablation. RESULTS AZVs in the 'standard mode' were smaller than predicted by the manufacturer (length 3.6 ± 0.6 cm, 23% below 4.7 cm; width 2.7 ± 0.6, 23% below 3.5 cm). Ablation zone past the tip was limited to 6 mm in 28/32 ablations. Differences in AZV between the 'surgical mode' and 'standard mode' were not significant (15.6 ± 7.8 mL vs. 13.9 ± 8.8 mL, p = 0.6). AZVs were significantly larger in case of hepatocellular carcinomas (HCCs) (n = 19) compared to metastasis (n = 19; 17.8 ± 9.9 mL vs. 10.1 ± 5.1 mL, p = 0.01) and in non-perivascular tumour location (n = 14) compared to perivascular location (n = 24, 18.7 ± 10.4 mL vs. 11.7 ± 6.1 mL, p = 0.012), with both factors remaining significant in two-way analysis of variance (HCC vs. metastasis: p = 0.02; perivascular vs. non-perivascular tumour location: p = 0.044). CONCLUSION Larger AZVs can be expected in cases of HCCs compared with metastases and in non-perivascular locations. Using the 'surgical mode' does not increase AZV significantly.
Collapse
Affiliation(s)
- René Michael Mathy
- Department of Diagnostic and Interventional Radiology, University Hospital of Heidelberg, 69120 Heidelberg, Germany; (A.G.)
| | - Athanasios Giannakis
- Department of Diagnostic and Interventional Radiology, University Hospital of Heidelberg, 69120 Heidelberg, Germany; (A.G.)
- 2nd Department of Radiology, University General Hospital, “ATTIKON” Medical School, National and Kapodistrian University of Athens, 12462 Athens, Greece
| | - Mareike Franke
- Department of Radiology, Cantonal Hospital of Lucerne, 6000 Lucerne, Switzerland
| | - Alain Winiger
- Department of Radiology, Cantonal Hospital of Lucerne, 6000 Lucerne, Switzerland
| | - Hans-Ulrich Kauczor
- Department of Diagnostic and Interventional Radiology, University Hospital of Heidelberg, 69120 Heidelberg, Germany; (A.G.)
| | - De-Hua Chang
- Department of Diagnostic and Interventional Radiology, University Hospital of Heidelberg, 69120 Heidelberg, Germany; (A.G.)
- Department of Radiology, Cantonal Hospital of Lucerne, 6000 Lucerne, Switzerland
| |
Collapse
|
6
|
Cafarchio A, Iasiello M, Vanoli GP, Andreozzi A. Microwave ablation modeling with AMICA antenna: Validation by means a numerical analysis. Comput Biol Med 2023; 167:107669. [PMID: 37948968 DOI: 10.1016/j.compbiomed.2023.107669] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/11/2023] [Revised: 10/16/2023] [Accepted: 10/31/2023] [Indexed: 11/12/2023]
Abstract
BACKGROUND AND OBJECTIVES Percutaneous microwave thermal ablation is based on electromagnetic waves that generate dielectric heating, and it is widely recognized as one of the mostly used techniques for tumor treatment. The aim of this work is to validate a predictive model capable of providing physicians with guidelines to be used during thermal ablation procedures avoiding collateral damage. METHODS A finite element commercial software, COMSOL Multiphysics, is employed to implement a tuning-parameter approach. Governing equations are written with reference to variable-porosity and Local Thermal Non-Equilibrium (LTNE) equations are employed. The simulations results are compared with available ex-vivo and in-vivo data with the help of regression analysis. For in-vivo data simulations, velocity vector modulus and direction are varied between 0.0007 and 0.0009 m/s and 90-270°, respectively, in order to use this parameter as a tuning one to simulate - and lately optimize with respect to the differences from experimental outcomes - all the possible directions of the blood flow with respect to the antenna, whose insertion angle is not registered in the dataset. RESULTS The model is validated using reference data provided by the manufacturer (AMICA), which is obtained from ex-vivo bovine liver. The model accurately predicts the size and shape of the ablated area, resulting in an overestimation lesser than 10 %. Additionally, predictive data are compared to an in-vivo dataset. The ablated volume is accurately predicted with a mean underestimation of 6 %. The sphericity index is calculated as 0.75 and 0.62 for the predictions and in-vivo data, respectively. CONCLUSION This study developed a predictive model for microwave ablation of liver tumors that showed good performance in predicting ablation dimensions and sphericity index for ex-vivo bovine liver and for in-vivo human liver data with the tuning technique. The study emphasizes the necessity for additional development and validation to enhance the accuracy and reliability of in-vivo application.
Collapse
Affiliation(s)
- A Cafarchio
- Dipartimento di Medicina e Scienze della Salute DIMES, Università degli Studi del Molise, Campobasso, Italy.
| | - M Iasiello
- Dipartimento di Ingegneria Industriale DII, Università degli Studi di Napoli "Federico II", Napoli, Italy
| | - G P Vanoli
- Dipartimento di Medicina e Scienze della Salute DIMES, Università degli Studi del Molise, Campobasso, Italy
| | - A Andreozzi
- Dipartimento di Ingegneria Industriale DII, Università degli Studi di Napoli "Federico II", Napoli, Italy
| |
Collapse
|
7
|
Cafarchio A, Iasiello M, Brunese MC, Francica G, Rocca A, Andreozzi A. Emprint Microwave Thermoablation System: Bridging Thermal Ablation Efficacy between Human Patients and Porcine Models through Mathematical Correlation. Bioengineering (Basel) 2023; 10:1057. [PMID: 37760159 PMCID: PMC10525213 DOI: 10.3390/bioengineering10091057] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/31/2023] [Revised: 08/30/2023] [Accepted: 09/06/2023] [Indexed: 09/29/2023] Open
Abstract
To investigate the in vivo ablation characteristics of a microwave ablation antenna in the livers of humans with tumors, a retrospective analysis of the ablation zones was conducted after applying Emprint microwave ablation systems for treatment. Percutaneous microwave ablations performed between January 2022 and September 2022 were included in this study. Subsequently, immediate post-ablation echography images were subjected to retrospective evaluation to state the long ablated diameter, short ablated diameter, and volume. The calculated ablation lengths and volume indices were then compared between in vivo and ex vivo results obtained from laboratory experiments conducted on porcine liver. The ex vivo data showed a good correlation between energy delivered and both increasing ablated dimensions (both p < 0.001) and volume (p < 0.001). The in vivo data showed a good correlation for dimensions (p = 0.037 and p = 0.019) and a worse correlation for volume (p = 0.142). When comparing ex vivo and in vivo data for higher energies, the ablated volumes grew much more rapidly in ex vivo cases compared to in vivo ones. Finally, a set of correlations to scale ex vivo results with in vivo ones is presented. This phenomenon was likely due to the absence of perfusion, which acts as a cooling system.
Collapse
Affiliation(s)
- Andrea Cafarchio
- Dipartimento di Medicina e Scienze della Salute DIMES, Università degli Studi del Molise, 86100 Campobasso, Italy; (M.C.B.); (A.R.)
| | - Marcello Iasiello
- Dipartimento di Ingegneria Industriale DII, Università degli Studi di Napoli “Federico II”, 80125 Napoli, Italy; (M.I.); (A.A.)
| | - Maria Chiara Brunese
- Dipartimento di Medicina e Scienze della Salute DIMES, Università degli Studi del Molise, 86100 Campobasso, Italy; (M.C.B.); (A.R.)
| | - Giampiero Francica
- Interventional Ultrasound Unit, Pineta Grande Hospital, 81030 Castel Volturno, Italy;
| | - Aldo Rocca
- Dipartimento di Medicina e Scienze della Salute DIMES, Università degli Studi del Molise, 86100 Campobasso, Italy; (M.C.B.); (A.R.)
| | - Assunta Andreozzi
- Dipartimento di Ingegneria Industriale DII, Università degli Studi di Napoli “Federico II”, 80125 Napoli, Italy; (M.I.); (A.A.)
| |
Collapse
|
8
|
Radmilović-Radjenović M, Bošković N, Sabo M, Radjenović B. An Analysis of Microwave Ablation Parameters for Treatment of Liver Tumors from the 3D-IRCADb-01 Database. Biomedicines 2022; 10:biomedicines10071569. [PMID: 35884874 PMCID: PMC9312906 DOI: 10.3390/biomedicines10071569] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/31/2022] [Revised: 06/13/2022] [Accepted: 06/17/2022] [Indexed: 11/28/2022] Open
Abstract
Simulation techniques are powerful tools for determining the optimal conditions necessary for microwave ablation to be efficient and safe for treating liver tumors. Owing to the complexity and computational resource consumption, most of the existing numerical models are two-dimensional axisymmetric models that emulate actual three-dimensional cancers and the surrounding tissue, which is often far from reality. Different tumor shapes and sizes require different input powers and ablation times to ensure the preservation of healthy tissues that can be determined only by the full three-dimensional simulations. This study aimed to tailor microwave ablation therapeutic conditions for complete tumor ablation with an adequate safety margin, while avoiding injury to the surrounding healthy tissue. Three-dimensional simulations were performed for a multi-slot microwave antenna immersed in two tumors obtained from the 3D-IRCADb-01 liver tumors database. The temperature dependence of the dielectric and thermal properties of healthy and tumoral liver tissues, blood perfusion, and water content are crucial for calculating the correct ablation time and, thereby, the correct ablation process. The developed three-dimensional simulation model may help practitioners in planning patient-individual procedures by determining the optimal input power and duration of the ablation process for the actual shape of the tumor. With proper input power, necrotic tissue is placed mainly in the tumor, and only a small amount of surrounding tissue is damaged.
Collapse
Affiliation(s)
- Marija Radmilović-Radjenović
- Institute of Physics, University of Belgrade, Pregrevica 118, 11080 Belgrade, Serbia; (N.B.); (B.R.)
- Correspondence:
| | - Nikola Bošković
- Institute of Physics, University of Belgrade, Pregrevica 118, 11080 Belgrade, Serbia; (N.B.); (B.R.)
| | - Martin Sabo
- Faculty of Informatics and Information Technologies, Slovak University of Technology in Bratislava, Ilkovicova 2, 84216 Bratislava, Slovakia;
| | - Branislav Radjenović
- Institute of Physics, University of Belgrade, Pregrevica 118, 11080 Belgrade, Serbia; (N.B.); (B.R.)
| |
Collapse
|
9
|
Das SS, Mahapatra SK. Study of heat sink effect of blood in a bifurcated vessel. Comput Methods Biomech Biomed Engin 2022; 26:721-733. [PMID: 35703320 DOI: 10.1080/10255842.2022.2085998] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/03/2022]
Abstract
Thermal ablation treatment uses elevated (hyperthermia) or depressed (hypothermia) tissue temperature to destroy tumor cells. The efficacy and effectiveness of thermal ablation therapy is dependent on the tissue temperature which is significantly affected due to heat sink effect of blood flow near the infected site. In this study, Euler-Euler multiphase model is used to analyze the effect of plasma and RBC concentration on the heat sink effect of blood in a bifurcated vessel. This study is divided into two separate cases. First case refers to the study of heat sink effect produced by a tumor patient suffering from HVS (hyperviscosity syndrome) and a normal (without blood disorder) tumor patient during hyperthermia treatment. The second case analyses the effect of RBCs on blood heat transfer. Temperature distribution and transient Nusselt number, which are used to represent heat sink effect, are calculated and compared for different cases of blood disorders. From the results, it is found that a patient with HVS blood disorder produces a smaller heat sink effect during hyperthermia treatment compared to a normal tumor patient. Also, the level of RBC concentration in the blood stream has a minimal effect on heat transfer.
Collapse
Affiliation(s)
- Sidharth Sankar Das
- Department of Mechanical Engineering, IIT Bhubaneswar, Khordha, Odisha, India
| | | |
Collapse
|
10
|
Servin F, Collins JA, Heiselman JS, Frederick-Dyer KC, Planz VB, Geevarghese SK, Brown DB, Miga MI. Fat Quantification Imaging and Biophysical Modeling for Patient-Specific Forecasting of Microwave Ablation Therapy. Front Physiol 2022; 12:820251. [PMID: 35185606 PMCID: PMC8850958 DOI: 10.3389/fphys.2021.820251] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/22/2021] [Accepted: 12/29/2021] [Indexed: 11/14/2022] Open
Abstract
Computational tools are beginning to enable patient-specific surgical planning to localize and prescribe thermal dosing for liver cancer ablation therapy. Tissue-specific factors (e.g., tissue perfusion, material properties, disease state, etc.) have been found to affect ablative therapies, but current thermal dosing guidance practices do not account for these differences. Computational modeling of ablation procedures can integrate these sources of patient specificity to guide therapy planning and delivery. This paper establishes an imaging-data-driven framework for patient-specific biophysical modeling to predict ablation extents in livers with varying fat content in the context of microwave ablation (MWA) therapy. Patient anatomic scans were segmented to develop customized three-dimensional computational biophysical models and mDIXON fat-quantification images were acquired and analyzed to establish fat content and determine biophysical properties. Simulated patient-specific microwave ablations of tumor and healthy tissue were performed at four levels of fatty liver disease. Ablation models with greater fat content demonstrated significantly larger treatment volumes compared to livers with less severe disease states. More specifically, the results indicated an eightfold larger difference in necrotic volumes with fatty livers vs. the effects from the presence of more conductive tumor tissue. Additionally, the evolution of necrotic volume formation as a function of the thermal dose was influenced by the presence of a tumor. Fat quantification imaging showed multi-valued spatially heterogeneous distributions of fat deposition, even within their respective disease classifications (e.g., low, mild, moderate, high-fat). Altogether, the results suggest that clinical fatty liver disease levels can affect MWA, and that fat-quantitative imaging data may improve patient specificity for this treatment modality.
Collapse
Affiliation(s)
- Frankangel Servin
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
- Vanderbilt Institute for Surgery and Engineering, Vanderbilt University, Nashville, TN, United States
| | - Jarrod A. Collins
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
| | - Jon S. Heiselman
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
- Vanderbilt Institute for Surgery and Engineering, Vanderbilt University, Nashville, TN, United States
| | - Katherine C. Frederick-Dyer
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States
| | - Virginia B. Planz
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States
| | - Sunil K. Geevarghese
- Department of Surgery, Vanderbilt University Medical Center, Nashville, TN, United States
| | - Daniel B. Brown
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States
| | - Michael I. Miga
- Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States
- Vanderbilt Institute for Surgery and Engineering, Vanderbilt University, Nashville, TN, United States
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, United States
- Department of Neurological Surgery, Vanderbilt University Medical Center, Nashville, TN, United States
- Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University Medical Center, Nashville, TN, United States
- *Correspondence: Michael I. Miga,
| |
Collapse
|
11
|
A New Thermal Damage-Controlled Protocol for Thermal Ablation Modeled with Modified Porous Media-Based Bioheat Equation with Variable Porosity. Processes (Basel) 2022. [DOI: 10.3390/pr10020236] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022] Open
Abstract
Thermal ablation of tumors is a minimally invasive technique more and more employed in cancer treatments. The main shortcomings of this technique are, on the one hand, the risk of an incomplete ablation, and on the other hand, the destruction of the surrounding healthy tissue. In this work, thermal ablation of a spherical hepatocellular carcinoma tumor (HCC) surrounded by healthy tissue is modeled. A modified porous media-based bioheat model is employed, including porosity variability from tumor core to healthy tissue, following experimental in vivo measures. Moreover, three different protocols are investigated: a constant heating protocol, a pulsating protocol, and a new developed damage-controlled protocol. The proposed damage-controlled protocol changes the heating source from constant to pulsating according to the thermal damage probability on the tumor rim. The equations are numerically solved by means of the commercial software COMSOL Multiphysics, and the outcomes show that the new proposed protocol is able to achieve the complete ablation in less time than the completely pulsating protocol, and to reach tissue temperature on the tumor rim 10 °C smaller than the constant protocol. These results are relevant to develop and improve different patient-based and automated protocols which can be embedded in medical devices’ software or in mobile applications, supporting medical staff with innovative technical solutions.
Collapse
|