Demir B, Ahmed AS, Babalik E, Demir M, Gürmen T. Verification and uniformity control of doses for Sr/Y intravascular brachytherapy sources using radiochromic film dosimetry.
J Med Phys 2011;
33:54-9. [PMID:
19893691 PMCID:
PMC2772026 DOI:
10.4103/0971-6203.41193]
[Citation(s) in RCA: 3] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/01/2008] [Accepted: 03/26/2008] [Indexed: 11/04/2022] Open
Abstract
Intravascular brachytherapy (IVBT) is a useful treatment modality for the recurrence of in-stent restenosis following drug-eluting stents (DES) or IVBT failure. The objective of this study was to measure the dose rate of (90)Sr/(90)Y IVBT sources for comparison with that given by the manufacturer and to control the dose uniformities of these sources along the source axis. The dose rates of (90)Sr/(90)Y beta sources were measured with a radiochromic film in a custom-made phantom. The films for calibration were irradiated using (60)Co photon beams. The results for the three sources were 4.5%, 2.3%, and 3.5% higher than the corresponding certificate values. Maximum and minimum of the dose rates varied within +/-10% of those at source center; and maximum dose discrepancy for the first (90)Sr/(90)Y source train was 8.2%; for the second source train, 7.1%; and for the third source train, 5.1%. Our study showed that the dose rates given by the manufacturer for the three (90)Sr/(90)Y IVBT sources were reliable and dose uniformities were within +/-10% along two thirds of the treatment length.