Richards CE, Obaid DR. Low-Dose Radiation Advances in Coronary Computed Tomography Angiography in the Diagnosis of Coronary Artery Disease. Curr Cardiol Rev 2019; 15:304-315. [PMID: 30806322 PMCID: PMC8142354 DOI: 10.2174/1573403x15666190222163737] [Citation(s) in RCA: 20] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 11/28/2018] [Revised: 12/24/2018] [Accepted: 12/26/2018] [Indexed: 01/16/2023] Open
Abstract
Background
Coronary computed tomography angiography (CCTA) is now widely used in the diagnosis of coronary artery disease since it is a rapid, minimally invasive test with a diagnostic accuracy comparable to coronary angiography. However, to meet demands for increasing spatial and temporal resolution, higher x-ray radiation doses are required to circumvent the resulting increase in image noise. Exposure to high doses of ionizing radiation with CT imaging is a major health concern due to the potential risk of radiation-associated malignancy. Given its increasing use, a number of dose saving algorithms have been implemented to CCTA to minimize radiation exposure to “as low as reasonably achievable (ALARA)” without compromising diagnostic image quality. Objective
The purpose of this review is to outline the most recent advances and current status of dose saving techniques in CCTA. Method
PubMed, Medline, EMBASE and Scholar databases were searched to identify feasibility studies, clinical trials, and technology guidelines on the technical advances in CT scanner hardware and reconstruction software. Results
Sub-millisievert (mSv) radiation doses have been reported for CCTA due to a combination of strategies such as prospective electrocardiogram-gating, high-pitch helical acquisition, tube current modulation, tube voltage reduction, heart rate reduction, and the most recent novel adaptive iterative reconstruction algorithms. Conclusion
Advances in radiation dose reduction without loss of image quality justify the use of CCTA as a non-invasive alternative to coronary catheterization in the diagnosis of coronary artery disease.
Collapse
Affiliation(s)
- Caryl E Richards
- Department of Cardiology, Morriston Hospital, Heol Maes Eglwys, Morriston, Swansea, SA6 6NL, United Kingdom
| | - Daniel R Obaid
- Department of Cardiology, Morriston Hospital, Heol Maes Eglwys, Morriston, Swansea, SA6 6NL, United Kingdom.,Swansea University Medical School, Swansea University, Grove Building, Singleton Park, Sketty, Swansea SA2 8PP, United Kingdom
| |
Collapse
|
2
|
Sabarudin A, Siong TW, Chin AW, Hoong NK, Karim MKA. A comparison study of radiation effective dose in ECG-Gated Coronary CT Angiography and calcium scoring examinations performed with a dual-source CT scanner. Sci Rep 2019; 9:4374. [PMID: 30867480 PMCID: PMC6416329 DOI: 10.1038/s41598-019-40758-5] [Citation(s) in RCA: 13] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/23/2018] [Accepted: 02/21/2019] [Indexed: 11/08/2022] Open
Abstract
In this report we have evaluated radiation effective dose received by patients during ECG-gated CCTA examinations based on gender, heart rate, tube voltage protocol and body mass index (BMI). A total of 1,824 patients were retrospectively recruited (1,139 men and 685 women) and they were divided into Group 1 (CCTA with calcium scoring), Group 2 (CCTA without calcium scoring) and Group 3 (only calcium scoring), where the association between gender, heart rate, tube voltage protocol and body mass index (BMI) were analysed. Examinations were performed using a retrospective ECG-gated CCTA protocol and the effective doses were calculated from the dose length product with a conversion coefficient of 0.026 mSv.mGy-1cm-1. No significant differences were observed in the mean effective dose between gender in all groups. The mean estimated dose was significantly higher when the heart rate was lower in Group 1 (p < 0.001) and Group 2 (p = 0.002). There were also significant differences between the mean effective dose in tube voltage protocol and BMI among the three groups. The mean effective dose was positively correlated with BMI (p < 0.001), but inversely related to the heart rate. This study supported the theory that a high heart rate, low tube voltage and low BMI could significantly reduce radiation dose exposure.
Collapse
Affiliation(s)
- Akmal Sabarudin
- Diagnostic Imaging & Radiotherapy Program, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300, Kuala Lumpur, Malaysia
| | - Tiong Wei Siong
- Diagnostic Imaging & Radiotherapy Program, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300, Kuala Lumpur, Malaysia
| | - Ang Wee Chin
- Department of Physics, Faculty of Science, Universiti Teknologi Malaysia, 81300, Johor Bharu, Johor, Malaysia
| | - Ng Kwan Hoong
- Department of Biomedical Imaging, Universiti of Malaya Medical Centre, 50603, Kuala Lumpur, Malaysia
| | - Muhammad Khalis Abdul Karim
- Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400, Serdang, Selangor, Malaysia.
- Center for Diagnostic Nuclear Imaging, Faculty of Medicine Universiti Putra Malaysia, 43400, Serdang, Selangor, Malaysia.
| |
Collapse
|
3
|
Chian TC, Nassir NM, Ibrahim MI, Yusof AKM, Sabarudin A. Quantitative assessment on coronary computed tomography angiography (CCTA) image quality: comparisons between genders and different tube voltage settings. Quant Imaging Med Surg 2017; 7:48-58. [PMID: 28275559 DOI: 10.21037/qims.2017.02.02] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
BACKGROUND This study was carried out to quantify and compare the quantitative image quality of coronary computed tomography angiography (CCTA) between genders as well as between different tube voltages scan protocols. METHODS Fifty-five cases of CCTA were collected retrospectively and all images including reformatted axial images at systolic and diastolic phases as well as images with curved multi planar reformation (cMPR) were obtained. Quantitative image quality including signal intensity, image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of right coronary artery (RCA), left anterior descending artery (LAD), left circumflex artery (LCx) and left main artery (LM) were quantified using Analyze 12.0 software. RESULTS Six hundred and fifty-seven coronary arteries were evaluated. There were no significant differences in any quantitative image quality parameters between genders. 100 kilovoltage peak (kVp) scanning protocol produced images with significantly higher signal intensity compared to 120 kVp scanning protocol (P<0.001) in all coronary arteries in all types of images. Higher SNR was also observed in 100 kVp scan protocol in all coronary arteries except in LCx where 120 kVp showed better SNR than 100 kVp. CONCLUSIONS There were no significant differences in image quality of CCTA between genders and different tube voltages. Lower tube voltage (100 kVp) scanning protocol is recommended in clinical practice to reduce the radiation dose to patient.
Collapse
Affiliation(s)
- Teo Chee Chian
- Diagnostic Imaging & Radiotherapy Program, School of Diagnostic and Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300 Kuala Lumpur, Malaysia
| | - Norziana Mat Nassir
- Diagnostic Imaging & Radiotherapy Program, School of Diagnostic and Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300 Kuala Lumpur, Malaysia
| | - Mohd Izuan Ibrahim
- Diagnostic Imaging & Radiotherapy Program, School of Diagnostic and Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300 Kuala Lumpur, Malaysia
| | | | - Akmal Sabarudin
- Diagnostic Imaging & Radiotherapy Program, School of Diagnostic and Applied Health Sciences, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, 50300 Kuala Lumpur, Malaysia
| |
Collapse
|
4
|
Li X, Chen GZ, Zhao YE, Schoepf UJ, Albrecht MH, Bickford MW, Gu HF, Zhang LJ, Lu GM. Radiation Optimized Dual-source Dual-energy Computed Tomography Pulmonary Angiography: Intra-individual and Inter-individual Comparison. Acad Radiol 2017; 24:13-21. [PMID: 27789203 DOI: 10.1016/j.acra.2016.09.009] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.9] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/06/2016] [Revised: 09/11/2016] [Accepted: 09/14/2016] [Indexed: 11/28/2022]
Abstract
OBJECTIVES This study aimed to intra-individually and inter-individually compare image quality, radiation dose, and diagnostic accuracy of dual-source dual-energy computed tomography pulmonary angiography (CTPA) protocols in patients with suspected pulmonary embolism (PE). METHODS Thirty-three patients with suspected PE underwent initial and follow-up dual-energy CTPA at 80/Sn140 kVp (group A) or 100/Sn140 kVp (group B), which were assigned based on tube voltages. Subjective and objective CTPA image quality and lung perfusion map image quality were evaluated. Diagnostic accuracies of CTPA and perfusion maps were assessed by two radiologists independently. Effective dose (ED) was calculated and compared. RESULTS Mean computed tomography (CT) values of pulmonary arteries were higher in group A than group B (P = .006). There was no difference in signal-to-noise ratio and contrast-to-noise ratio between the two groups (both P > .05). Interobserver agreement for evaluating subjective image quality of CTPA and color-coded perfusion images was either good (κ = 0.784) or excellent (κ = 0.887). Perfusion defect scores and diagnostic accuracy of CTPA showed no difference between both groups (both P > .05). Effective dose of group A was reduced by 45.8% compared to group B (P < .001). CONCLUSIONS Second-generation dual-source dual-energy CTPA with 80/Sn140 kVp allows for sufficient image quality and diagnostic accuracy for detecting PE while substantially reducing radiation dose.
Collapse
Affiliation(s)
- Xie Li
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China
| | - Guo Zhong Chen
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China
| | - Yan E Zhao
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China
| | - U Joseph Schoepf
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
| | - Moritz H Albrecht
- Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
| | - Matthew W Bickford
- Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
| | - Hai Feng Gu
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China
| | - Long Jiang Zhang
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China.
| | - Guang Ming Lu
- Department of Medical Imaging, Jinling Hospital, Medical School of Nanjing University, No. 305, Zhongshan East Road, Nanjing, Jiangsu, 210002, China.
| |
Collapse
|
5
|
Roobottom CA, Loader R. Virtual Special Issue Radiation dose reduction in CT: dose optimisation gains both increasing importance and complexity! Clin Radiol 2016; 71:438-41. [PMID: 26983651 DOI: 10.1016/j.crad.2016.02.004] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/25/2016] [Revised: 02/03/2016] [Accepted: 02/08/2016] [Indexed: 12/30/2022]
Affiliation(s)
- C A Roobottom
- Department of Radiology, Derriford Hospital, Plymouth PL6 8DH, UK.
| | - R Loader
- Department of Physics, Derriford Hospital, Plymouth PL6 8DH, UK
| |
Collapse
|
6
|
Automated tube voltage selection in thoracoabdominal computed tomography at high pitch using a third-generation dual-source scanner: image quality and radiation dose performance. Invest Radiol 2015; 50:352-60. [PMID: 25591129 DOI: 10.1097/rli.0000000000000133] [Citation(s) in RCA: 39] [Impact Index Per Article: 4.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/15/2022]
Abstract
OBJECTIVES The objective of this study was to evaluate the radiation dose and image quality performance of thoracoabdominal examinations with an automated tube voltage selection (tube voltage adaptation), tube current modulation, and high pitch using a third-generation dual-source computed tomography (CT) compared intraindividually with 120-kV examinations with tube current modulation with special attention on clinically relevant lesions in the liver, the lungs, and extrahepatic soft tissues. MATERIALS AND METHODS This study was approved by the institutional review board. Computed tomography of the body was performed using a third-generation dual-source system in 95 patients (mean body mass index, 25 kg/m²; range, 18-35 kg/m²). For 49 of these patients, all calculated tube settings and resulting dose values were recorded for each of the 12 gradual contrast weightings of the tube voltage adaptation algorithm. Spiral CT was performed for all patients with an intermediate weighting (grade 7) in a portal venous phase at 120 reference kV, 180 reference mAs, and pitch of 1.55. Objective image quality was assessed on the basis of contrast-to-noise ratio. Subjective image quality was assessed on the basis of clarity and sharpness of anatomical and pathological structures as well as interfering beam hardening and spiral and motion artifacts (heart, lungs, diaphragm). Previous examinations on a 64-slice scanner served as reference. RESULTS All examinations were rated good or excellent for clinical diagnosis. Automated tube voltage selection resulted in significantly lower effective radiation dose (9.5 mSv) compared with the reference (12.0 mSv; P < 0.01). Contrast-to-noise ratio and image quality of soft tissue lesions were significantly increased (P < 0.01). Motion artifacts were significantly reduced (P < 0.01). CONCLUSIONS Automated tube voltage adaptation combined with high-pitch protocols allows for a substantial radiation dose reduction while substantially increasing the image quality, even at large-volume exposure.
Collapse
|